23
CoboSafe-CBSF
325-2810-012US15
PREPARING FOR MEASUREMENT
10 Preparing for Measurement
The force sensors CoboSafe-CBSF are the fundamental device for force and pressure measure-
ments. The measuring device is positioned at the collision points determined for the application.
The additional damping element K1 is placed on the measuring surface, when this is defined
in the test specification and included in the configuration of the measurement plan.
Depending on the measuring plan, a pressure measuring film from the CoboSafe-Scan or
CoboSafe-Tek system is placed on the damping element K1 or directly on the measuring plate
(Fig. 9). The collision measurement is carried out with this setup.
Keep in mind that CoboSafe-CBSF has to acclimate to the surroundings for at least
two hours before beginning the operation.
10.1 Safety When Preparing a Measurement
CAUTION
Sharp edges
Risk of cuts due to sharp edges!
■
Only establish the measuring setup with completely debur-
red components.
■
Wear personal protective equipment.
There is a risk of cutting injuries on edges that have not been
deburred.
Fig. 12: Measurement setup without and with K1 damping element