MAINTAINING FUNCTIONALITy
13 Maintaining Functionality
The CoboSafe-CBSF force measuring devices should be inspected before and after measure-
ment to ensure proper functioning. Look for apparent damage and excessive pollution. Never
use a damaged device.
If there is any uncertainty regarding proper functioning or precision, e.g., because the device
has been dropped or shows damage, the CoboSafe-CBSF should be sent to the manufacturer
to be inspected.
13.1 Maintenance Schedule
Inspection Interval
Maintenance Tasks
Personnel
Before each measurement
Inspection and cleaning
Laboratory scientist,
System integrator
After each measurement
Inspection and cleaning
Laboratory scientist,
System integrator
After approx. 20 operating
hours or when indicated
Charging batteries
Laboratory scientist,
System integrator
Annually
Calibration of
CoboSafe-CBSF
Manufacturer or certified ser-
vice personnel
Replacement of dam-
ping elements K1
Laboratory scientist,
System integrator
Every 2 years
Replacing batteries
Manufacturer
After updating standards or
in case of wear
If necessary, replace dam-
ping elements K1, firmware
update
Manufacturer
13.2
Checking and Cleaning
Instructions for checking and cleaning CoboSafe-CBSF force sensors and K1 damping ele-
ments.
Personnel
Personal Protective Equipment
Materials
■
Laboratory scientist
■
System integrator
■
Protective clothing
■
Protective gloves
■
Safety footwear
■
Soft, lint-free cloth