6
%
The light between the photoelectric receiving tube and the light-emitting tube of the device
must pass through the subject
’
s arteriole. Make sure the optical path is free from any
optical obstacles like rubberized fabric, to avoid inaccurate results.
%
Excessive ambient light may affect the measured results, such as surgical light (especially
xenon light sources), bilirubin lamp, fluorescent lamp, infrared heater and direct sunlight,
etc. In order to prevent interference from ambient light, make sure to place the sensor
properly and cover the sensor with opaque material.
%
requent movement (active or passive) of the subject or severe activity can affect the
measured accuracy.
%
The SpO2 probe should not be placed on a limb with the blood pressure cuff, arterial
ductus or intraluminal tube.
%
The measured value may be inaccurate during defibrillation and in a short period after
defibrillation, as it has not defibrillation function.
%
The device has been calibrated before leaving factory.
%
The device is calibrated to display functional oxygen saturation.
%
The equipment connected with the Oximeter interface should comply with the
requirements of IEC 60601-1.
%
Please select medical power adapter to charge it, when connecting the special adapter with
the socket, make sure there is no shelter near the socket and it is easy to plug and unplug,
Summary of Contents for RS01
Page 22: ...11 4 2 Introduzione all interfaccia Figura4 1 Interfaccia di misurazione ...
Page 41: ...30 Figura 5 11 Comeindossare la cannula di ossigeno nasale ...
Page 51: ...40 Figura 5 22 ...
Page 59: ...48 Figura 5 29 Figura 5 30 ...
Page 88: ...77 Fluttuazioni di tensione sfarfallamento IEC 61000 3 3 Applicabile ...
Page 115: ...12 4 2 Interface introduction Figure4 1 Measurement interface ...
Page 128: ...25 Figure 5 6 Record setting menu Figure 5 7 Recorded time interface ...
Page 133: ...30 Figure 5 11 Wearing mode of nasal oxygen cannula ...
Page 142: ...39 the case information is stored in case library Figure 5 21 ...
Page 143: ...40 Figure 5 22 ...
Page 151: ...48 Figure 5 30 ...
Page 178: ...75 Voltage fluctuations flicker emission IEC 61000 3 3 Applicable ...
Page 206: ...11 4 2 Introducción a la interfaz Figura Interfaz de medición ...
Page 220: ...25 registrará Figura Menú de configuración de registro ...
Page 225: ...30 Figura Modo de uso de la cánula de oxígeno nasal ...
Page 235: ...40 Figura ...
Page 243: ...48 Figura ...
Page 257: ...62 Figura 7 Haga clic en Ayuda aparecerá el documento de ayuda de este software ...
Page 271: ...76 Fluctuaciones de voltaje emisiones de parpadeo IEC 61000 3 3 Aplicable ...
Page 299: ...11 4 2 Introduction d interface Figure4 1 Interface mesure ...
Page 313: ...25 Figure 5 6 Menu de réglage enregistrement Figure 5 7 Interface temps enregistré ...
Page 318: ...30 Figure 5 11 Mode de port de la canule nasale à oxygène ...
Page 328: ...40 Figure 5 21 ...
Page 329: ...41 Figure 5 22 ...
Page 337: ...49 Figure 5 30 ...
Page 351: ...63 Figure 5 46 7 Cliquez sur Help le document d aide de ce logiciel apparaîtra ...
Page 366: ...78 Fluctuations de tension Scintillements CEI 61000 3 3 Applicable ...