The device may not be suitable for all users, if you can't get a satisfactory result, please
stop using it.
%
Data averaging and signal processing have a delay in the upgrade of SpO2 data values.
When the data update period is less than 30 seconds, the time for obtaining dynamic
average values will increase, which is arisen from signal degradation, low perfusion or
other interference, it depends on the PR value.
%
The device has 5-year service life, date of manufacture: see the label.
%
The device does not provide over-limit alarm function for SpO2 and PR, so it is
inapplicable for using in the place where need such function.
%
The maximum temperature at the SpO2 probe -tissue interface should be less than 41
℃
which is measured by the temperature tester.
%
During measuring, when abnormal conditions appear on the screen, please pull out your
finger and reinsert it to measure again.
%
If some unknown error appears during measuring,the product can be connected to the
adapter through the date line,long press
“
”
button about 10 seconds for reset and then
shut down to terminate the operation of the product.
%
Do not contort or drag the wire of the device.
%
The plethysmographic waveform is not normalized, as a signal inadequacy indicator, when
Summary of Contents for RS01
Page 22: ...11 4 2 Introduzione all interfaccia Figura4 1 Interfaccia di misurazione ...
Page 41: ...30 Figura 5 11 Comeindossare la cannula di ossigeno nasale ...
Page 51: ...40 Figura 5 22 ...
Page 59: ...48 Figura 5 29 Figura 5 30 ...
Page 88: ...77 Fluttuazioni di tensione sfarfallamento IEC 61000 3 3 Applicabile ...
Page 115: ...12 4 2 Interface introduction Figure4 1 Measurement interface ...
Page 128: ...25 Figure 5 6 Record setting menu Figure 5 7 Recorded time interface ...
Page 133: ...30 Figure 5 11 Wearing mode of nasal oxygen cannula ...
Page 142: ...39 the case information is stored in case library Figure 5 21 ...
Page 143: ...40 Figure 5 22 ...
Page 151: ...48 Figure 5 30 ...
Page 178: ...75 Voltage fluctuations flicker emission IEC 61000 3 3 Applicable ...
Page 206: ...11 4 2 Introducción a la interfaz Figura Interfaz de medición ...
Page 220: ...25 registrará Figura Menú de configuración de registro ...
Page 225: ...30 Figura Modo de uso de la cánula de oxígeno nasal ...
Page 235: ...40 Figura ...
Page 243: ...48 Figura ...
Page 257: ...62 Figura 7 Haga clic en Ayuda aparecerá el documento de ayuda de este software ...
Page 271: ...76 Fluctuaciones de voltaje emisiones de parpadeo IEC 61000 3 3 Aplicable ...
Page 299: ...11 4 2 Introduction d interface Figure4 1 Interface mesure ...
Page 313: ...25 Figure 5 6 Menu de réglage enregistrement Figure 5 7 Interface temps enregistré ...
Page 318: ...30 Figure 5 11 Mode de port de la canule nasale à oxygène ...
Page 328: ...40 Figure 5 21 ...
Page 329: ...41 Figure 5 22 ...
Page 337: ...49 Figure 5 30 ...
Page 351: ...63 Figure 5 46 7 Cliquez sur Help le document d aide de ce logiciel apparaîtra ...
Page 366: ...78 Fluctuations de tension Scintillements CEI 61000 3 3 Applicable ...