Measurement and Analysis
282
Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Figure 7-16: Tracking Validation screen
Softmenu
Speed
- Allows to slow down the loop play-back speed.
Syst YOYO
-
When turned ON the image will loop over the
systolic portion of the heart cycle.
Recalc
- Allows to go back to previous screen in order to
manually correct the ROI.
Approve
- Press
Approve
once the tracking quality has been
validated.
1. Display Quick Tips on tracking quality assessment
2. The ROI divided in segments
3. The Scoring table
•
: acceptable tracking
•
: unacceptable tracking
4. Bulls-Eye icon:
• Segments with yellow border: segments being analyzed.
• Green segments: segments already analyzed.
• Black segments: segments not analyzed.