Measurement and Analysis
Vivid S5/Vivid S6 User Manual
269
R2424458-100 Rev. 2
Figure 7-7: Segment Time to peak measurements screen
TSI trace
The TSI Time to peak measurement can be verified and
eventually manually changed from the TSI trace.
1.
Double click on the measurement point.
The ROI and the corresponding TSI curve are displayed
(see Figure 7-8).
2.
Press
SET
to anchor the ROI and trace.
3.
If required, select a new peak location in the trace.
4.
Click in the acquisition window to exit the TSI trace.