Scanning Modes
Vivid S5/Vivid S6 User Manual
151
R2424458-100 Rev. 2
•
Anatomical M-Mode (AMM): displays a distance/time plot
from a cursor line, which is independent from the axial
plane. AMM is available in greyscale, color, or in the
optional modes: TVI, Tissue Tracking, Strain rate, and
Strain modes.
•
Curved Anatomical M-Mode (CAMM): displays a
distance/time plot from a free-drawn cursor line. CAMM is
available in greyscale, color, and optional TVI, Tissue
Tracking, Strain rate and Strain modes.
M-Mode and 2D Mode display areas can be side by side or
top/bottom. Conventional M-Mode can be combined with Color
(see "Color M-Mode overview" on page 158).
Note
: The sweep speed information displayed in the bottom
right corner of the image represents the user selected sweep
speed and should be used only as a reference to confirm that
the image was acquired at the selected sweep speed. It is not
to be used for measurements or analysis. This is not an
absolute value, but simply a reference number. Users
performing studies using standardized protocols may find this
sweep speed information useful for reading studies from other
institutions.
M-Mode controls
M-Mode assignable controls
Horizontal sweep
Adjusts the horizontal refresh rate of the M-Mode area of the
display. Horizontal sweep does not change the acquisition
resolution, so that user can change the horizontal sweep in
replay (with no loss of quality).
On some low fre-
quencies, the sys-
tem switches
automatically to
second-harmonic
mode. The word
“Octave” appears
in the status win-
dow.
Frequency
Enables the adjustment of the probe's operating frequency.
The selected frequency is displayed in the status window. For
some probes/applications the lowest frequency settings will be
Octave imaging settings.
Note
: While changing the Frequency value on display, the
operator actually selects different transmit patterns associated
with that value, which includes transmit pulse shape, frequency
and transmit sequence.
Summary of Contents for Vivid S5
Page 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 689: ......
Page 690: ......