LV Contrast overview
Figure 6-1: The LV Contrast acquisition screen
LV Contrast controls
LV Contrast assignable controls
Width
Controls the size and angular width of the image sector. A
smaller angle generally produces a scan with a higher frame
rate.
Tilt
Enables the axis of the 2D image to be tilted to the left or to the
right. By default the axis of the 2D image is vertical.
Frequency
Enables the adjustment of the probe's operating frequency. A
higher frequency gives better resolution. Frequency is also
used to switch between Octave (single-pulse) and CPI (Coded
Phase inversion - multi-pulse).