Measurement and Analysis
Vivid S5/Vivid S6 User Manual
263
R2424458-100 Rev. 2
7.
Trackball the cursor to the measurement end point.
The current velocity and pressure values are displayed in
the
Measurement result table
and are instantaneously
updated when moving the cursor.
The measurement
display color on the
spectrum changes
from green to red af-
ter completion of
the measurement.
8.
Press
SET
to anchor the end point of the measurement.
The following measurement results are displayed in the
Measurement result table
:
• Velocity and pressure at anchor point positions
• Velocity (V3) and pressure (p3) differences between
anchor point position
• Time difference (dT) between anchored points position
The measurement
display on the spec-
trum and the corre-
sponding result are
numbered.
9.
To assign a label to the measurement, see page 252.
10. Repeat steps 5 through 8 to make additional
measurements.
Manual Doppler trace measurements
Adjust Compress
and reject controls
to optimize the
Doppler signal.
1.
Generate the spectrum to be measured.
2.
Press
FREEZE
to stop the cineloop.
3.
Press
MEASURE
on the Control Panel.
Alternative: Press
CALIPER
on the con-
trol panel and press
the
MANUAL TRACE
assignable
4.
Select
Trace
in the
Measurement Menu
.
A vertical green cursor is displayed on the spectrum.
5.
Trackball the cursor to the start point on the left side of the
trace.
6.
Press
SET
to anchor the start point of the measurement.
7.
With the
trackball
, trace the Doppler envelope.
The trace can be adjusted, while tracing, by moving the
cursor backward to erase portion of the trace (or the entire
trace) and then create the trace again.
The measurement
display color on
spectrum changes
from green to red af-
ter completion of
the measurement.
8.
Press
SET
to complete the trace.
The following measurement results are displayed in the
Measurement result table
:
• Maximum and mean Velocities
• Maximum and mean pressures
• Env.
Ti
• Velocity time integral (VTI)
9.
Trackball the cursor to the start point of the next heart beat.
10. Press the
SET
to anchor the next heart beat starting point.
The heart rate (BPM) is displayed in the
Measurement
result table
.
Summary of Contents for Vivid S5
Page 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Page 689: ......
Page 690: ......