GE M
EDICAL
S
YSTEMS
PROPRIETARY TO GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
8-68
Section 8-13 - Loading Application Software
12.)During the reboot process, the system may ask for the option string (valid license) to be
entered. When this happens, enter the option string provided in the installation paperwork.
NOTE:
Starting with R3.x Software, all systems must have at least the basic option string in order for
the software to run. Any single option string may enable multiple options (i.e. Basic & B-Flow).
Demo option strings turn on all options for a limited period of time. Do Not leave a demo option
string in a unit when it is sold to a customer. The current R3.x Demo Option String expires 31
January, 2005.
Figure 8-75 Enter Option String