Chapter 8 Replacement Procedures
8-137
8-35-3
How to Update Vital Product Data on LOGIQ
™
9
8-35-3-1
Initial Steps
The System should be in the scan mode. If not:
1.) Properly shut down the system.
2.) Plug in the service dongle to the USB Port on the External I/O.
3.) Power up system
8-35-3-2
Start Service Platform
1.) Right Click on the
Wrench
icon to start the service platform.
2.) A dialog box appears "Service Platform is starting ..."
3.) It will take a few seconds for the Service Login to appear. Select login as
GE Service
4.) Enter current Service Password for Service Diagnostics.
NOTE:
Remember that the Caps Lock must be OFF.
Figure 8-144 Service Login Screen