GE M
EDICAL
S
YSTEMS
PROPRIETARY
TO
GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
Chapter 8 Replacement Procedures
8-159
Section 8-41
Frogleg Replacement Procedure
8-41-1
Manpower
2 persons, 3 hours
8-41-2
Tools
•
Hexagon key, 10 mm, with torque/momentum controller
•
Hexagon keys, 4, 5 and 6 mm
•
Phillips screwdriver, size 2 and 3
8-41-3
Preparations
1.) Shut down the system as described in
Power Off / Shutdown on page 4-3
2.) Move Top Console to down position.
3.) Remove monitor as described in
Monitor Assembly Replacement Procedure on page 8-92
4.) Unscrew two (2) hexagon screws to loosen the Console Cable Neck from the Frogleg.
5.) Unscrew four (4) screws on the Upper Column Cover.