Chapter 8 Replacement Procedures
8-69
Section 8-14
Upgrading Application Software
8-14-1
Applications Software Upgrade Procedure
NOTE:
When installing software you may see error messages about not being able to replace the
following *.dll files:
ginastub.dll
and
msvcr70.dll
. These are known, non-critical issues that are
being tracked and will be fixed in the next revision of software.
1.) Place the “Application Software” CD-ROM into the CD-ROM drive.
2.) Power Down the scanner.
3.) Power up the scanner. Wait for a dialog box like the one in
.
WARNING
WARNING
While the software install procedure is designed to preserve data, you should save any
patient data, images, system setups to a MOD or hardcopy.
CAUTION
Remember to remove the MOD before continuing with the software install.
Figure 8-76 Start Application Window