Chapter 10 Periodic Maintenance
10-9
10-5-2-2
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted. Refer
to the User Manual for a list of approved peripherals/options.
10-5-3
Input Power
10-5-3-1
Mains Cable Inspection
10-5-4
Cleaning
10-5-4-1
General Cleaning
Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks
Step
Item
Description
1
VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
2
B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3
Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
5
InSite/iLinq
Verify that InSite is functioning properly. Ensure two-way remote communications.
(Warranty & Contract Customers only)
6
Camera
Verify hardcopy output of the film camera. Clean as necessary.
7
Footswitch Verify that the footswitch is functioning as programed. Clean as necessary.
8
ECG
Verify basic operation with customer
Table 10-9 Mains Cable Inspection
Step
Item
Description
1
Unplug Cord Disconnect the mains cable from the wall and system.
2
Inspect
Inspect it and its connectors for damage of any kind.
3
Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no
strands may cause a short circuit.
4
Verify
Inlet connector retainer is functional.
Table 10-10 General Cleaning
Step
Item
Description
1
Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
2
Probe Holder Clean probe holders (they may need to be soaked to remove excess gel).
3
Monitor and
Touch Panel waiting for procedure from engineering