GE M
EDICAL
S
YSTEMS
PROPRIETARY TO GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
8-46
Section 8-8 - BEP Patient I/O Replacement Procedure (BEP 2.x and above only)
Section 8-8
BEP Patient I/O Replacement Procedure (BEP 2.x and above only)
8-8-1
Manpower
One person, 1 hour + travel
8-8-2
Tools
Phillips screwdriver
Preparations (click here to view the cover removal video)
1.) Power Down/Shutdown the system as described in
2.) Remove the left- and/or right-side covers as shown in
3.) Remove the front cover as shown in
8-8-4
Patient I/O Removal Procedure
NOTE:
Mounting hardware may vary in size from M3 to 4-40 threads. Hardware length may seem the
same but in fact may be different and very critical. ALWAYS keep track of EXACTLY where
each screw was removed so that it is returned to the same place on re-assembly.
1.) Remove the side cover from the BEP. Carefully disconnect the cover fan power cable and set the
cover aside.
2.) Disconnect the cable from the back of the patient I/O. It may be easier to disconnect this cable at
the I/O VIC Tray because access is limited due to space behind the Patient I/O.
3.) Remove the one screw attaching the side of the Patient I/O as shown in
Figure 8-49 Patient I/O Connection