Chapter 8 Replacement Procedures
8-11
8-2-7
Front Bumper Replacement Procedure
8-2-7-1
Manpower
One person, 30 minutes
8-2-7-2
Tools
Phillips size 2
Torx screwdriver T-20
8-2-7-3
Preparations
1.) Shut down the system as described in
.
2.) Remove Front Cover as described in
8-2-7-4
Front Bumper Removal Procedure
1.) Unscrew 4 Torx screws under the bumper.
2.) Lift the bumper off.
8-2-7-5
Front Bumper Installation Procedure
1.) Put the new bumper in position.
2.) Fasten the 4 screws firmly.
NOTE:
Do NOT overtighten these screws.
3.) Install Front Cover as described in
. (click here to view the cover
Figure 8-7 System Front Bumper
Screw holes
Front Bumper