10-4
Section 10-3 - Why Do Maintenance?
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
Table 10-2 Customer Care & Maintenance Task Schedule
Service at Indicated Time
Daily
Weekly
Monthly
Per
Facility’s
QA Program
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Inspect AC Mains Cable
•
Mobile Ultrasound system: Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean Cockpit
•
Inspect Wheels, Casters, Brakes and
Swivel Locks
•
Mobile Unit: Check Daily
Check Articulated Arm Movement
•
Mobile Unit: Check Daily
Functional Checks
•
Also after corrective maintenance.
Safety Test
•*
* twice a year