8-164
Section 8-8 - Peripherals Replacement/Installation Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
Section 8-8
Peripherals Replacement/Installation Procedures
8-8-1
Printer Replacement/Installation Procedure
8-8-1-1
Tools
Appropriate Phillips screwdriver
8-8-1-2
Time Required
10 min
8-8-1-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-8-1-4
Printer Removal Procedure
NOTE:
Perform these steps only for Venue™ systems with printer already installed on them. For
systems with no previously installed printer, skip this section and proceed to section
Installation Procedure (For Systems with No Existing Printer)
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, Right Side
eTower, Upper eTower Front Cover.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Upper eTower Front Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Summary of Contents for H45051VN
Page 477: ......