8-14
Section 8-3 - Covers - Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-3-7
Printer Insert Cover Replacement Procedure
8-3-7-1
Tools
Appropriate Phillips screwdriver
8-3-7-2
Time Required
5 min
8-3-7-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-3-7-4
Printer Insert Cover Removal Procedure
1) Remove all accessories.
2) Remove the Upper eTower Front Cover.:.
3.) Release the two Phillips screws and remove the Printer Insert Cover.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
Upper eTower Front Cover Removal Procedure
Figure 8-7 Removing Printer Insert Cover
Printer Insert Cover
Summary of Contents for H45051VN
Page 477: ......