8-206
Section 8-9 - Operating System and Application Software Loading Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
2.) Perform a full back-up for Patient Archive and System Configuration
(for instructions, refer to the Venue™ User Manual on your Document Media).
3.) Write down the following settings as it will be necessary to restore them at the end of the process:
-
TCP/IP address
-
Network printer (if present)
-
Wireless network settings (if present)
-
InSite ExC settings
8-9-4-2
Operating System and/or Application Software Installation Procedure
NOTE:
A minimum of
1 hour
is required to install both the Windows operating system and the Venue™
application software. This excludes preparation time, backup etc. and performing functional checks
following the installation procedures.
Throughout the installation processes, the Venue™ system will be unavailable for scanning.
on page 8-184 provides step-by-step instructions for installation of Windows
operating system software and Venue™ ultrasound scanner software application. However, these
procedures may be performed separately, as required.
Follow the installation instructions below, as appropriate:
•
Windows operating system and Venue™ s/w application:
•
Windows operating system ONLY:
•
Venue™ software application ONLY:
NOTE:
It is necessary to perform the recommended Functionality Checks after completion of the installation
procedures, when installing the following:
•
Windows operating system and Venue™ s/w application
•
Venue™ software application ONLY