8-20
Section 8-3 - Covers - Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-3-11
Riser Cover Replacement Procedure
8-3-11-1
Tools
Appropriate Phillips screwdriver
8-3-11-2
Time Required
5 min
8-3-11-3
Preparations
Shut down the Venue™ ultrasound unit, as described in
Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-3-11-4
Riser Cover Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, RS Probe Cover. Right Side and Left Side
eTower covers, Upper Front eTower, and Riser Thermal Cover.
3.) Remove cockpit (monitor).
4.) Remove Scanner Arm.
5.) Loosen and remove four screws securing the riser cover to the riser (access screws from
underneath).
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
Lower Front eTower Cover Removal Procedure
•
RS Probe Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Upper eTower Front Cover Removal Procedure
•
Riser Thermal Cover Removal Procedure
•
Cockpit (Monitor) Removal Procedure
•
Summary of Contents for H45051VN
Page 477: ......