8-36
Section 8-4 - System Modules- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-5
Cockpit (Monitor) Replacement Procedure
8-4-5-1
Tools
Appropriate Phillips screwdriver
Flat-head screwdriver
8-4-5-2
Time Required
15 min
8-4-5-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-5-4
Cockpit (Monitor) Removal Procedure
1.) On the back of the cockpit, release four screws and remove the transparent Wi-Fi dongle cover.
•
Disconnect the Wi-Fi dongle.
2.) Remove four Phillips screws securing the cockpit (monitor) to the arm.
FRU Part #
Refer to
.
Figure 8-22 Wi-Fi Cover and Dongle Removal
Summary of Contents for H45051VN
Page 477: ......