2-10
Section 2-3 - Facility Needs
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
2-3-2-2
Recommended Floor Plan
Figure 2-2 Recommended floor plan, 4.27 x 5.18 m (14 x 17 foot)
1.
Secretaries or Doctors Desk
10.
Dedicated Power Outlet - Circuit Breaker
protected and easily accessible
2.
File Cabinet
11.
Network Interface
3.
Film Viewer
12.
457 mm (18 inches) distance of Ultrasound
system from wall or objects
4.
Counter Top
13.
Stool
5.
Counter Top and Sink with hot and cold water
14.
Footswitch
6.
Overhead Lights Dimmer - Dual Level Lighting
(bright and dim)
15.
Storage for Linens and Equipment
7.
Emergency Oxygen
16.
Examination Table – 1930 x 610 mm
(76 x 24 inches)
8.
Suction Line
17.
Lavatory and Dressing Room
9.
Ultrasound system
18.
Door – at least 762 mm (30 inches)
Scale:
Each square equals one square foot (app. 31 x 31 cm)