D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
Chapter 3 - System Setup
3-25
P R E L I M I N A R Y
3-5-5-2
Connect a Probe
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe.
There are 4 probe connectors on the rear panel of the system, numbered 1 through 4. Each connector
is comprised of a probe-socket (6) and a locking latch (5).
1) Before connecting the probe:
a.) Do a visual check of the probe pins and system sockets.
b.) Remove any dust or foam rests from the probe pins.
c.) Verify the probe and the probe cable for any visual damage.
CAUTION
Do not allow the probe head to hang freely. Excessive impact
to the probe will result in irreparable damage.
CAUTION
To prevent probe connector pins damage, or PCB board
damage, do not use excessive force when connecting the
probes.
CAUTION
Keep the probe cables away from the wheels.
Do not bend the probe cables.
Do not cross cables between probes.
Figure 3-26 Venue™ Probe Connectors
3