D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
4-12
Section 4-2 - General Procedures
P R E L I M I N A R Y
4-2-8-2
To Prepare the Venue™ to be Moved
1) Make sure the cockpit (monitor) is centered and in its down position.
2) Remove the plug from the wall.
3) Disconnect all cables linking the unit to any off-board peripheral devices and network.
4) Secure the unit’s power cable.
5) Place all probes in the probe holders. Ensure that the probe cables do not protrude from the unit or
interfere with the wheels.
6) Ensure that no loose items are left on the unit.
7) Unlock the casters.
Related information:
•
Cockpit (Monitor) Position Adjustment
•
CAUTION
Unlock the brakes on the front and back wheels when preparing to
transport the unit.
Lock the brakes on the front and back wheels, if parking the system on an
incline. Avoid ramp steeper than 10 degrees.