Chapter 4 - General Procedures and Functional Checks
4-11
P R E L I M I N A R Y
4-2-6
Deleting Patient Information
Ensure that
All Patient Information
has been deleted before:
•
shipping/returning the ultrasound system
•
you dispose of the hard drive
Wipe the SSD partitions as described here:
1) Reload the system software from the software media.
2) Select:
Format SSD and Install
.
4-2-7
Cockpit (Monitor) Position Adjustment
The Cockpit (monitor) position can be adjusted for easy viewing.
NOTE:
Before adjusting the cockpit (monitor) position, lock the front swivel and full lock caster located
on the system front left side (user right side).
4-2-7-1
To Adjust the Cockpit (Monitor) Position
By holding the cockpit (monitor) handle, the up/down and left/right position of the cockpit can be
adjusted.
To tilt the cockpit (monitor), use the lever handle and adjust the cockpit (monitor) to the desired angle.
4-2-8
Moving and Transporting the Venue™ Ultrasound Scanner
4-2-8-1
The Casters (Wheels) Control
The pedals located between the left front and right back wheels of the Venue™ control the swivel and
brake on the wheels.
Examine the wheels frequently for defects to avoid breaking or jamming.
WARNING
WARNING
Before you dispose of the hard drive, make sure you remove ALL
PATIENT DATA from the hard drive, given that the hard drive is still
functional. In some countries, you may be required to delete all software
from the disk before returning the hard drive to the parts warehouse.
Follow your local policies.
WARNING
WARNING
All patient information (images, reports and data base) must be backed
up before deleting it from the internal hard drive.
CAUTION
To avoid injury or damage, make sure nothing is within the
range of motion before moving the cockpit and monitor
arm. This includes both objects and people.