Summary of Contents for H45051VN

Page 1: ...Operating Documentation 2019 General Electric Company 16 JUN 2019 GEHC_FRNT_CVR FM P R E L I M I N A R Y Venue Service Manual Direction Number 5771498 100 Revision 6 Technical Publication ...

Page 2: ... 2019 General Electric Company Manufacturer GE Medical Systems Ultrasound Primary Care Diagnostic LLC 9900 Innovation Drive Wauwatosa WI 53226 USA 16 JUN 2019 GEHC_FRNT_CVR FM P R E L I M I N A R Y ...

Page 3: ...Standard Hazard Icons 1 7 Safety Considerations 1 8 Introduction 1 8 Human Safety 1 8 Mechanical Safety 1 11 Electrical Safety 1 13 Venue Battery Safety 1 17 Patient Data Safety 1 17 Dangerous Procedure Warnings 1 18 Lockout Tagout LOTO Requirements 1 19 Product Labels and Icons 1 20 Universal Product Labels 1 20 Label Descriptions 1 23 Venue External Labels Location 1 27 Returning Shipping Probes...

Page 4: ...ements 2 6 Facility Needs 2 7 Purchaser Responsibilities 2 7 Required Facility Needs 2 8 Networking Pre Installation Requirements 2 12 Connectivity Installation Worksheet 2 13 Overview 3 1 Purpose of Chapter 3 3 1 Setup Reminders 3 2 Average Setup Time 3 2 Setup Warnings 3 2 Safety Reminders 3 3 Receiving and Unpacking the Equipment 3 4 Warnings for Receiving and Unpacking the Equipment 3 4 Overvi...

Page 5: ... Software Options Configuration 3 42 Connectivity Overview 3 43 Physical Connection 3 43 Stand alone Venue 3 43 Sneaker Net Environment 3 43 Wireless Connection from Venue to DICOM Server 3 43 Connectivity Setup 3 44 Introduction 3 44 Select TCP IP Screen 3 45 Setting Up Non Broadcasting Hidden Wireless Network Connection 3 46 Changing the AE Title and or Port Number Port No 3 50 Wireless Network ...

Page 6: ...ng and Transporting the Venue Ultrasound Scanner 4 11 Functional Checks 4 14 Overview 4 14 Performance Checks 4 14 2D Mode B Mode Checks 4 14 M Mode Checks 4 15 PW CW Doppler Mode Checks 4 17 Probe Connectors Check 4 20 Cineloop Check 4 21 Audio Check 4 22 Peripheral Checks 4 23 Mechanical Functions Checks 4 26 Electrical Tests 4 29 Overview 5 1 Purpose of Chapter 5 5 1 General Information 5 2 Int...

Page 7: ...5 19 Signal Flow 5 19 Front End Power Supply T FEPS 5 20 Front End T CFE Board 5 20 Probe Selection Board T PSB 5 21 Back End Processor 5 22 Introduction 5 22 Back End Processing BEP Module 5 22 BEP Block Diagram 5 24 Back End Interface BIB Board 5 25 Solid State Hard Drive 5 25 System Power Distribution 5 26 Introduction 5 26 AC Distribution 5 26 DC Power Distribution 5 26 System Power Management...

Page 8: ...Swivel Resistance Adjustments 6 4 Arm Vertical Movement Adjustments 6 4 Arm Rotate Resistance Adjustments 6 5 Overview 7 1 Purpose of Chapter 7 7 1 Service Safety Considerations 7 2 Service Tools 7 3 Visual Guide 7 3 System Diagnostics 7 3 Gathering Troubleshooting Data 7 5 Purpose of this Section 7 5 Contents in this Section 7 5 Collect Vital System Information 7 5 Collect a Trouble Image with Lo...

Page 9: ...ooting 7 25 Purpose of this Section 7 25 MPB Diagnostics Procedure Troubleshooting 7 25 Overview 8 1 Purpose of Chapter 8 8 1 Visual Guide 8 1 Accessories Replacement Procedures 8 2 Basic Storage Basket Large Storage Basket Replacement Procedure 8 2 Power Cable Holder Replacement Procedure 8 3 Covers Replacement Procedures 8 4 Overview of Covers 8 4 Lower Front eTower Cover Replacement Procedure 8...

Page 10: ...rocedure 8 71 T TRx Box Replacement Procedure 8 73 T TRx Module Replacement Procedure 8 75 Mechanical Parts Replacement Procedures 8 78 MPB Front Metal Door Replacement Procedure 8 78 MPB Guide L and Guide R Replacement Procedure 8 80 Halo Handle Replacement Procedure 8 83 Articulated Arm Replacement Procedure 8 86 Riser Replacement Procedure 8 91 Casters Replacement Procedure 8 95 Plastic Cable G...

Page 11: ...g Procedures 8 184 Burning Disk on Key Media with SW Downloaded from GE Portal 8 184 Software Update Procedure 8 188 Software Upgrade Procedure from Venue R1 to Venue R2 8 193 Software Installation Procedure General Overview 8 205 Software Installation Procedure 8 207 8 211 Software Recovery Procedure 8 212 Functional Checks to be Performed after Replacement Procedures 8 216 General Overview 8 216...

Page 12: ...k Schedule 10 3 Tools Required 10 5 Tools Required for Servicing the Venue 10 5 System Maintenance 10 6 Preliminary Checks 10 6 Functional Checks 10 7 Physical Inspection 10 9 Cleaning 10 10 Probe Maintenance 10 12 Probe Related Checks 10 12 Probe Handling 10 12 Basic Probe Care 10 13 Probe Cleaning 10 13 Returning and Shipping of Defective Probes 10 14 Electrical Safety Tests 10 15 Overview 10 15...

Page 13: ...DIRECTION5771498 100 REVISION 6 VENUE SERVICE MANUAL Table of Contents 13 P R E L I M I N A R Y This page was intentionally left blank ...

Page 14: ...NE AUTRE LANGUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVEN TION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS ÉTÉ CONSULTÉ ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN L OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES MÉCANIQUES OU AUTRES DIESES KUNDENDIENST HANDBUCH EXISTIERT NUR IN EN...

Page 15: ... DO TÉCNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS MECÂNICOS OU OUTROS ESTE MANUAL DE ASSISTÊNCIA ESTÁ DISPONÍVEL APENAS EM INGLÊS SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA QUE NÃO A GEHC SOLICITAR ESTES MANUAIS NOUTRO IDIOMA É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO NÃO TENTE EFECTUAR REPARAÇÕES NO EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO PREVIAMENTE E...

Page 16: ...N LAITTEISTON KÄYTTÄJÄN TAI POTILAAN VAHINGOITTUMINEN SÄHKÖISKUN MEKAANISEN VIAN TAI MUUN VAARATILANTEEN VUOKSI ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ ΔΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗ ΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ...

Page 17: ...RICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO JINÉMU RIZIKU DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE DENNE SERVICEMANUAL ER ...

Page 18: ...TEIKĖJAS OPERATORIUS AR PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO MECHANINIŲ AR KITŲ PAVOJŲ DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØ...

Page 19: ...НИЙ МОЖЕТ ПРИВЕСТИ К ТОМУ ЧТО СПЕЦИАЛИСТ ПО ТЕХОБСЛУЖИВАНИЮ ОПЕРАТОР ИЛИ ПАЦИЕНТ ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ ДРУГОЕ ПОВРЕЖДЕНИЕ ТОВА СЕРВИЗНО РЪКОВОДСТВО Е НАЛИЧНО САМО НА АНГЛИЙСКИ ЕЗИК АКО ДОСТАВЧИКЪТ НА СЕРВИЗНИ УСЛУГИ НА КЛИЕНТ СЕ НУЖДАЕ ОТ ЕЗИК РАЗЛИЧЕН ОТ АНГЛИЙСКИ ЗАДЪЛЖЕНИЕ НА КЛИЕНТА Е ДА ПРЕДОСТАВИ ПРЕВОДАЧЕСКА УСЛУГА НЕ СЕ ОПИТВАЙТЕ ДА ИЗВЪРШВАТЕ СЕРВИЗНО ОБ...

Page 20: ... ILI PACIJENTA PROUZROČENE STRUJNIM UDAROM MEHANIČKIM I DRUGIM OPASNOSTIMA TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO ANGLIČTINU POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE ZODPOVEDNOSŤOU ZÁKAZNÍKA NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR AKO SI NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ ZANEDBANIE TOHTO UPOZORNENIA M...

Page 21: ...TERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE ETMEYİNİZ BU UYARININ GÖZ ARDI EDİLMESİ ELEKTRİK ÇARPMASI YA DA MEKANİK VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN OPERATÖRÜN YA DA HASTANIN YARALANMASINA YOL AÇABİLİR JA Traditional Chinese ...

Page 22: ...P R E L I M I N A R Y DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL ix 6 16 19 POC_PRECAUTIONS_LEGAL ZH CN KO ...

Page 23: ...trical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personnel of third party service companies with equivalent training or licensed electricians to perform electrical servicing on the equipment OMISSIONS ERRORS If there are any omissions errors or suggestions...

Page 24: ...o change without notice Every precaution has been taken in the preparation of this document Nevertheless GE assumes no responsibility for errors omissions or any damages including special or consequential resulting from the use of this information GE will issue updates to this information periodically as needed If there are any questions regarding the information contained in this manual please co...

Page 25: ...os and corrections 3 11 FEBURARY 2018 Support R2 configuration 4 16 JULY 2018 Added Kisyu codes ECG support and Upgrade from R1 to R2 procedure 5 10 APRIL 2019 Added R2 5 DOK 6 6 JUNE 2019 Added troubleshooting instructions Pages Revision Pages Revision Pages Revision 7 43 to 7 60 Add Troubleshooting instructions noise audio Un recognized probes system does not turn on MPB diagnostics 1 15 Add war...

Page 26: ... N A R Y Chapter 1 Introduction Section 1 1 Overview 1 1 1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Venue ultrasound scanner The service provider must read and understand all the information presented here before installing or servicing a unit ...

Page 27: ...s installation planners some parts of Chapter 2 Site Preparations Table 1 1 Contents in this Service Manual CHP NUMBER TITLE DESCRIPTION Chapter 1 Introduction Contains a content summary and warnings Chapter 2 Site Preparations Contains pre setup requirements for the Venue ultrasound scanner Chapter 3 System Setup Contains setup procedure with an setup checklist Chapter 4 General Procedures and Fu...

Page 28: ...n in Table 1 2 below NOTE When not specified otherwise the contents in this manual apply to all Venue models Table 1 2 Venue Models PSI Group Cat No Description PSI Code GP Code Kisyu Code Venue R1 H45051VN Venue ultrasound scanner UVENUA 24422A EXB0 Venue R2 H45281VN Venue R2 ultrasound scanner UVENUD 24422B EXB0 Venue R1 to R2 upgrade kit H45281SW Venue R1 to R2 Upgrade Kit UVENUE 24501A EXB1 ...

Page 29: ...ation in 2D Color Doppler M Mode Color M Mode and PW The fully digital architecture of the Venue system allows optimal usage of all scanning modes and probe types throughout the full spectrum of operating frequencies Signal flows from the Probe Connector Panel to the Front End and then over to the Back End Processor and finally to the cockpit and peripherals System configuration is stored on the V...

Page 30: ...e hardware versions on the boards Refer to HW Version tab To check the software versions on local software on the boards Refer to Software Version To check for installed options Refer to Options Setup on page 3 52 1 2 4 4 Purpose of Operator Manual s The Operator Manual s should be fully read and understood before operating the Venue system and also kept near the unit for quick reference ...

Page 31: ... message Known or potential hazards to personal are labeled in one of three ways DANGER WARNING CAUTION When a hazard is present that can cause property damage but has absolutely no personal injury risk a NOTICE is used NOTE Notes are used to provide important information about an item or a procedure NOTE Be sure to read the notes the information contained in a note can often save you time or effo...

Page 32: ...ed to make you aware of specific types of hazards that could possibly cause harm Refer to the User Manual for a complete list of icons used on the Ultrasound System not all icons may be listed in the table below Other hazard icons make you aware of specific procedures that should be followed Be sure to read the notes the information contained in a note can often save you time or effort NOTE The Ve...

Page 33: ...MOVED FROM AN OPERATING ULTRASOUND SYSTEM SOME METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED EVEN WHILE IN SHUTDOWN MODE WARNING WARNING BECAUSE OF THE LIMITED ACCESS TO CABINETS AND EQUIPMENT IN THE FIELD PLACING PEOPLE IN AWKWARD POSITIONS GE HAS LIMITED THE LIFTING WEIGHT FOR ONE PERSON IN THE FIELD TO 16 KG 35 LBS ANYTHING OVER 16 KG 35 LBS REQUIRES 2 PEOPLE WARN...

Page 34: ...OR AT LEAST 30 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO TEST POINTS TO VERIFY ISOLATION THE LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF ULTRASOUND SYSTEM COMPONENTS MAY BE ENERGIZED ALWAYS REFER TO THE ULTRASOUND SYSTEM S SERVICE MANUAL FOR LOTO WARNINGS AND CAUTIONS CAPACITORS ON ULTRASOUND SYSTEMS WITH THE SHEARWAVE OPTION CAN TAKE UP TO 5 MINUTES TO DISCHARGE WARNING WARNING FO...

Page 35: ...verview P R E L I M I N A R Y CAUTION USE PROTECTIVE GLASSES DURING DRILLING FILING AND DURING ALL OTHER WORK WHERE EYES NEED PROTECTION CAUTION USE SAFETY SHOES WHEN DOING WORK WHERE THERE IS ANY CHANCE OF FOOT DAMAGE CAUTION USE PROTECTIVE GLOVES WHEN DRILLING AND CUTTING ...

Page 36: ... WEIGHT OF INSTALLED PERIPHERALS WHEN READY FOR USE CARE MUST BE USED WHEN MOVING IT OR REPLACING ITS PARTS FAILURE TO FOLLOW THE PRECAUTIONS LISTED BELOW COULD RESULT IN INJURY UNCONTROLLED MOTION AND COSTLY DAMAGE USE THE HANDLE TO MOVE THE ULTRASOUND SYSTEM BE SURE THE PATHWAY IS CLEAR LIMIT MOVEMENT TO A SLOW CAREFUL WALK DO NOT LET THE ULTRASOUND SYSTEM STRIKE WALLS OR DOOR FRAME USE TWO PEOP...

Page 37: ...ile inside the vehicle Secure the system with straps or as directed otherwise to prevent motion during transport Prevent vibration damage by driving cautiously Avoid unpaved roads excessive speeds and erratic stops or starts NOTICE SPECIAL CARE SHOULD BE TAKEN WHEN TRANSPORTING THE ULTRASOUND SYSTEM IN A VEHICLE BEFORE TRANSPORTING PLACE THE ULTRASOUND SYSTEM IN ITS SPECIAL STORAGE CASE ENSURE THA...

Page 38: ... connector must meet international electrical standards WARNING WARNING CONNECTING A VENUE SCANNER TO INCORRECT POWER LEVEL WILL DESTROY THE SYSTEM CONNECT THE SYSTEM ONLY IN ACCORDANCE WITH THE VOLTAGE INDICATED ON THE PRODUCT LABEL WARNING WARNING SAFE PRACTICES FOLLOW THESE GUIDELINES TO MINIMIZE SHOCK HAZARDS WHENEVER YOU ARE USING THE ULTRASOUND SYSTEM TO MINIMIZE SHOCK HAZARD THE EQUIPMENT C...

Page 39: ...t Avoid pulling pinching or kinking the probe cable since a damaged cable may compromise the electrical safety of the probe To avoid the risk of a probe accidentally falling do not allow the probe cables to become entangled or to be caught in the system s wheels Follow these guidelines before connecting a probe to the scanner Inspect the probe prior to each use for damage or degradation to the hou...

Page 40: ...ice a distance of 2 5 m 8 2 ft above the floor on which the medical personnel stand and a horizontal distance of 1 5 m 4 9 ft have justified themselves as indicative of the dimensions of the Patient WARNING WARNING Do not attempt to use different peripherals accessories brand and model connected via USB ports other than approved and provided by GE Healthcare The Ultrasound system is extremely sens...

Page 41: ...e see the example in Figure 1 1 Patient Environment EN 60601 1 Sub clause 3 79 Patient Environment It is difficult for this standard to define dimensions for the volume in which diagnosis monitoring or treatment occurs The dimensions for the PATIENT ENVIRONMENT given in Figure 1 2 have been justified in practice Figure 1 1 Patient Safety Environment Figure 1 2 Example of Patient Environment ...

Page 42: ...d other heat sources Do not leave the battery in direct sunlight Do not pierce the battery with a sharp object hit it or step on it Do not use a damaged battery Do not apply solder to a battery Do not connect the battery to an electrical power outlet Do not immerse the battery in water or allow it to get wet Do not place the battery into a microwave oven or pressurized container If the battery lea...

Page 43: ...EMOVED FROM AN OPERATING VENUE SYSTEM SOME METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED EVEN WHILE IN SHUT DOWN MODE WARNING WARNING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD EQUIPMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF A FLAMMABL...

Page 44: ... All potentially hazardous stored or residual energy is relieved WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VENUE WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS 1 Follow LOCK OUT TAG OUT procedures 2 TURN OFF THE BREAKER 3 UNPLUG THE SYSTEM 4 MAINTAIN CONTROL OF THE SYSTEM POWER PLUG 5 WAIT FOR AT LEAST 30 SECONDS FOR CAPACITORS TO DISCHARGE A...

Page 45: ...explanation of label icons and symbols refer to Table 1 3 on page 1 23 A system Rating Label examples shown in Figure 1 3 and Figure 1 4 below is located at the rear of the system This indicates the ultrasound unit s basic power compliance In addition a General Label Figure 1 6 and Figure 1 7 below provides details regarding regulatory compliance as well as warnings and cautions Figure 1 3 Rating ...

Page 46: ...r 1 Introduction 1 21 P R E L I M I N A R Y When Upgrading the software from R1 to R2 an electronic label will be displayed under the About tab in the Config screen Figure 1 5 Electronic label eLabel for R1 to R2 upgrade Figure 1 6 General Label International R1 ...

Page 47: ...TION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 1 22 Section 1 1 Overview P R E L I M I N A R Y Figure 1 7 General Label International R2 Figure 1 8 Venue Warning Label Figure 1 9 Stored Energy Hazard Label ...

Page 48: ...e user documentation when complete information cannot be provided on the label Various WARNING Dangerous voltage the lightning flash with arrowhead in equilateral triangle is used to indicate electric shock hazards Various CE mark of conformity Rear of the system Read the Service Manual Intended to alert the user to refer to the operator manual or other instructions when complete information canno...

Page 49: ...ot push the unit sideways when the caster wheel brakes are in the locked position Instability may occur Rating Plate DO NOT place objects on the surface of the rear of the LCD Panel while folded Rating Plate Equipotentiality Indicates the terminal to be used for connecting equipotential conductors when interconnecting grounding with other equipment Peripherals Alternating current Various Protectiv...

Page 50: ...arded Bottom This symbol indicates the product contains hazardous materials in excess of the limits established by the Chinese standard GB T 26572 Requirements of concentration limits for certain restricted substances in electrical and electronic products The number in the symbol is the Environment friendly Use Period EFUP which indicates the period during which the hazardous substances contained ...

Page 51: ...en the Assembled in X statement is not shown on the label this indicates that the Customs country of origin is the same as the country of the legal manufacturer Every system has a unique marking for identification the Unique Device Identification UDI Label The UDI label consists of a series of alpha numeric characters and barcode which uniquely identify the Venue system as a medical device manufac...

Page 52: ...P R E L I M I N A R Y 1 7 3 Venue External Labels Location In addition to the labels described in the previous section an additional label may be found on the Venue ultrasound unit as described in the following section Figure 1 1 Venue System Configuration of Main Hardware Components ...

Page 53: ... or intended for use in patient care are regulated medical waste for transportation purposes and must be transported as a hazardous material NOTE The USER SERVICE staff should dispose of all the waste properly per federal state and local waste disposal regulations The Venue ultrasound scanner is not meant to be used for long term storage of patient data or images The user is responsible for the da...

Page 54: ...s supplied with the system NOTE Portable and mobile RF communications equipment can affect the Venue ultrasound scanner 1 9 2 Compliance The Venue ultrasound scanner conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge radiated and conducted RF fields magnetic fields and power line transient requirements NOTE For applicable standards refer ...

Page 55: ...TIONS WARNING WARNING RISK OF ELECTRICAL SHOCK SYSTEM MUST BE TURNED OFF AVOID ALL CONTACT WITH ELECTRICAL CONTACTS CONDUCTORS AND COMPONENTS ALWAYS USE NON CONDUCTIVE HANDLES DESIGNED FOR THE REMOVAL AND REPLACEMENT OF ESD SENSITIVE PARTS ALL PARTS THAT HAVE THE POTENTIAL FOR STORING ENERGY MUST BE DISCHARGED OR ISOLATED BEFORE MAKING CONTACT WARNING WARNING IF THE COVERS ARE REMOVED FROM AN OPER...

Page 56: ... capture cine loop 4D multi volume loop Restart the application before resuming clinical scanning Table 1 4 Phone Numbers for Customer Assistance 1 of 2 LOCATION PHONE NUMBER USA GE Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa WI 53226 USCAN Service On site Service Parts OLC Application Support 1 800 437 1171 1 800 558 2040 1 800 321 7937 or 1 262 524 5300 1 800 682 5327 or 1 262...

Page 57: ...GE Healthcare Pvt Ltd 4 Kadugodi Industrial Area Bangalore 560 067 India Phone 91 1 800 425 8025 91 1 800 425 7255 91 1 800 102 7750 Table 1 5 Phone and Fax Numbers for Manufacturer MANUFACTURER PHONE NUMBER FAX NUMBER GE Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa WI 53226 1 800 437 1171 1 414 721 3865 Table 1 4 Phone Numbers for Customer Assistance Continued 2 of 2 LOCATION PH...

Page 58: ...er 2 This chapter provides the information required to plan and prepare for the setup of a Venue ultrasound unit Included are descriptions of the electrical and facility requirements that must be met by the purchaser A worksheet is provided at the end of this chapter see Figure 2 3 on page 2 13 to help ensure that all the required network information is available prior to setup ...

Page 59: ... needed for Ultrasound system installation updates and repairs However operator and patient comfort may be optimized if the room light is subdued and indirect Therefore a combination lighting system dim bright is recommended Keep in mind that lighting controls and dimmers can be a source of EMI which could degrade image quality These controls should be selected to minimize possible interference 2 ...

Page 60: ... conduit 2 2 5 1 Venue Power Requirements Electrical specifications for the Venue system are as follows 2 2 5 2 Inrush Current Inrush current is not a factor for consideration due to the inrush current limiting properties of the power supplies 2 2 5 3 Site Power Outlets A dedicated AC power outlet must be within reach of the unit without requiring the use of extension cords Other outlets adequate ...

Page 61: ...there is no guarantee that interference will not occur in a particular installation NOTE Possible EMI sources should be identified before the unit is installed and should not be on the same line as the ultrasound system A dedicated line should be used for the ultrasound system Electrical and electronic equipment may produce EMI unintentionally as the result of a defect Some of these sources includ...

Page 62: ... magnet wrap at each end Install all covers Loose or missing covers or Radio Frequency gaskets allow radio frequencies to interfere with the ultrasound signals Replace broken RF gaskets If more than 20 or a pair of the fingers on an Radio Frequency gasket are broken replace the gasket Do not turn on the Ultrasound system until any loose metallic part is removed Do not place labels where RF gaskets...

Page 63: ...on as possible if possible six weeks before delivery to allow enough time to make any changes WARNING WARNING FOR CONSOLE ULTRASOUND SYSTEMS AND FOR ULTRASOUND SYSTEMS MOUNTED ON A DOCKING ISOLATION CART HAVE TWO PEOPLE AVAILABLE TO DELIVER AND UNPACK THE ULTRASOUND SYSTEM CAUTION Attempts to move the Ultrasound system considerable distances or on an inclive by one person could result in injury or...

Page 64: ...le electrical codes The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment The desire to use a non listed or customer provided product or to place an approved product further from the Ultrasound system than the interface kit allows presents challenges to the installation team To avoid delays during installation such variances should be ...

Page 65: ...heral are within 2 m 6 5 ft of each other with peripheral within 1 m of the Venue system to connect cables Power outlets for other medical equipment Power outlets for test equipment within 1 m 3 2 ft of the ultrasound unit Clean and protected space to store probes in their cases or on a rack Material to safely clean probes done using a plastic container never metal In the case of a network option ...

Page 66: ... full size ground wire from the distribution panel to the Ultrasound outlet Sites with a mains power system without a defined Neutral The dedicated line shall consist of one phase two lines not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Figure 2 1 Minimal floor plan 2 5 m x 3 m 8 by 10 foot 1 Door at least 762 mm 30 inches 8 Stool...

Page 67: ...binet 11 Network Interface 3 Film Viewer 12 457 mm 18 inches distance of Ultrasound system from wall or objects 4 Counter Top 13 Stool 5 Counter Top and Sink with hot and cold water 14 Footswitch 6 Overhead Lights Dimmer Dual Level Lighting bright and dim 15 Storage for Linens and Equipment 7 Emergency Oxygen 16 Examination Table 1930 x 610 mm 76 x 24 inches 8 Suction Line 17 Lavatory and Dressing...

Page 68: ...ide Circuit breaker for dedicated power outlet is easily accessible Sink with hot and cold water Receptacle for bio hazardous waste like used probe sheaths Emergency oxygen supply Storage for linens and equipment Nearby waiting room lavatory and dressing room Dual level lighting bright and dim Lockable cabinet ordered by GE for its software and proprietary manuals ...

Page 69: ...servers for storage and to workstations for viewing images As an added benefit transferring images in this manner enables viewing to be done on reviewing station while scanning continues 2 3 3 4 DICOM Option Pre Installation Requirements To configure the Venue ultrasound unit to work with other network connections the network administrator must provide the required information which should include...

Page 70: ... 2 3 Connectivity Installation Worksheet Site System Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept Venue SN Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone E Mail Address Site Scanner IP Settings IP Address Subnet Mask Default Gateway Name AE Title ...

Page 71: ...ksheet for DICOM Network Information Venue Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses Default DICOM APPLICATION INFORMATION NAME Store 2 Store 3 Store 1 MAKE REVISION IP ADDRESSES PORT AE TITLE Store 5 Store 6 Store 4 Storage Commit MPPS Work list ...

Page 72: ...iances from the basic setup Make sure system and probes have been subject to acclimation period Environmental cooling is sufficient Lighting is adjustable to adapt to varying operational conditions of the scanner Electrical facilities meet system requirements EMI precautions have been taken and all possible sources of interference have been removed Mandatory site requirements have been met If a ne...

Page 73: ...2 16 Section 2 4 Connectivity Installation Worksheet DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y This page was intentionally left blank ...

Page 74: ...epared setup can proceed as described in this chapter Included in this chapter are guidelines for transporting the unit to a new site as well as procedures that describe how to receive and unpack the equipment and if necessary how to file a damage or loss claim Instructions for checking and testing the unit probes and external peripherals for electrical safety are also provided NOTE A Venue is rea...

Page 75: ...e 3 5 3 2 2 1 System Acclimation Time When unpacking the Ultrasound system allow the temperature of the Ultrasound system to stabilize before powering up The following table describes guidelines for reaching operational temperatures from storage or transport temperatures Following transport the Venue system may be very cold or hot Allow time for the system to acclimate before being switched ON Acc...

Page 76: ...K THE SYSTEM AS IT IS HEAVY TWO PEOPLE ARE ALWAYS REQUIRED WHENEVER A PART WEIGHING 16KG 35 LB OR MORE MUST BE LIFTED CAUTION TO PREVENT ELECTRICAL SHOCK CONNECT THE UNIT TO A PROPERLY GROUNDED POWER OUTLET DO NOT USE A THREE PRONG TO TWO PRONG ADAPTER AS THIS DEFEATS SAFETY GROUNDING CAUTION DO NOT WEAR THE ESD WRIST STRAP WHEN YOU WORK ON LIVE CIRCUITS WHERE MORE THAN 30 V PEAK IS PRESENT CAUTIO...

Page 77: ...or wooden crate mounted on a raised wooden base see Figure 3 5 on page 3 5 CAUTION Two people are needed to unpack the Ultrasound system because of its weight Attempts to move the Ultrasound system considerable distances or on an incline by one person could result in injury or damage or both Two people are required whenever a part weighing 16 KG 35 LBS or more must be lifted CAUTION Remember to us...

Page 78: ...n Venue Table 3 10 Shipping Carton Dimensions and Weights Description Height Width Depth Weighta a Weight is approximate and will vary depending upon the supplied peripherals Venue scanner with peripherals and accessories 160 cm 63 ins 79 cm 31 ins 77 5 cm 30 5 ins 121 kgs 267 lbs Figure 3 6 Additional Labels on Shipping Carton ...

Page 79: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 3 6 Section 3 3 Receiving and Unpacking the Equipment P R E L I M I N A R Y Figure 3 7 Shipping Packages ...

Page 80: ...l inspection of the shipping carton proceed to the unpacking instructions below as appropriate NOTE When using utility a knife to cut the nylon wrap take extra care not to scratch the system or any peripheral 1 Cut the securing strips 2 Remove the nylon wrap from the lower part of the package to gain access to the accessories boxes Figure 3 8 Removing Nylon Wrap ...

Page 81: ...he Equipment P R E L I M I N A R Y 3 Remove the accessories boxes 4 Remove the protective foams from the top of the system Figure 3 9 Removing Accessories Boxes Figure 3 10 Removing Protective Foams Probes boxes 3 and 4 Small basket TV_TR tray Probes box 2 Probes box 1 Additional accessories box ...

Page 82: ...EVISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 9 P R E L I M I N A R Y 5 Remove the wooden ramps and then remove the nylon wrap from the system Figure 3 11 Removing Nylon Wrap from the System Wooden ramps Nylon wrap ...

Page 83: ...eiving and Unpacking the Equipment P R E L I M I N A R Y 6 Unlock the four butterfly locks and remove the two wooden brackets 7 Remove the silica gel bags Figure 3 12 Removing Wooden Brackets Figure 3 13 Removing Silica Gel Bags Wooden brackets Butterfly locks Silica gel bags ...

Page 84: ... completely from the shipping carton 10 Make sure the system is standing firmly on the floor before applying the locking brake 11 Before commencing the installation procedure follow the instructions for Physical Inspection on page 3 17 12 Remove the packing slip shipping consignment note detailing the contents of the shipping carton and keep close to hand ready to mark the check list refer to Phys...

Page 85: ...from the safety latches 2 Remove the locks and open the four safety latches located on both sides of the crate 3 The shipping crate contains the following items Venue system Power cord Manuals Accessories Figure 3 15 Removing Locking Pin CAUTION WHEN USING SHARP TOOLS TO OPEN PACKING MATERIALS TAKE CARE TO AVOID CUTTING OR DAMAGING ANY OF THE CONTENTS Figure 3 16 Opening the Safety Latches ...

Page 86: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 13 P R E L I M I N A R Y Figure 3 17 Shipping Crate Contents ...

Page 87: ...CE MANUAL 3 14 Section 3 3 Receiving and Unpacking the Equipment P R E L I M I N A R Y 4 Lay the front wooden crate door ramp and secure it with the 2 metal pins located inside the crate Figure 3 18 Wooden crate ramp 5 Remove the silica gel bags ...

Page 88: ...NUE SERVICE MANUAL Chapter 3 System Setup 3 15 P R E L I M I N A R Y 6 Remove the anti static foams 7 Remove the wooden planks to gain access to the system Figure 3 19 Removing the Anti Static Foams Figure 3 20 Removing the Wooden Planks ...

Page 89: ...packing the Equipment P R E L I M I N A R Y 8 Remove the plastic bags from the system 9 Release the caster locking on the front and rear casters 10 Carefully roll the scanner forwards down the unloading ramp to remove it completely from the shipping crate Figure 3 21 Removing Plastic Bags ...

Page 90: ...gnetic Interference EMI Many of the covers shields and screws are provided primarily to protect the system from image artifacts caused by this interference For this reason it is imperative that all covers and hardware are installed and secured before the unit is put into operation WARNING WARNING CONNECTING A VENUE SCANNER TO INCORRECT VOLTAGE LEVEL WILL MOST LIKELY DESTROY IT CONNECT THE SYSTEM O...

Page 91: ...plugged Clean the console 2 Probes Check all probes for wear and tear on the lens cable and connector Look for bent or damaged pins on the connector and in the connector socket on the unit Verify that the EMI fingers around the probe connector socket housing are intact Check the probe locking mechanism and probe switch 3 Cockpit Monitor Clean the Cockpit monitor by gently wiping with a dry soft li...

Page 92: ...ew of the Venue Ultrasound Unit Figure 3 22 Front View of the Venue Ultrasound Scanner Item 1 Cockpit monitor Tilts up and down and swivels left and right 2 eTower 3 Basket holders 4 Caster 1x front with free movement and no locking 5 Swivel and full lock caster 1x front 6 Front Handle 7 Auxiliary monitor 1 2 5 7 3 4 3 6 ...

Page 93: ...push the scanner and place it in the desired position 10 On Off Switch Cover 3 Probe locking levers 11 Interface Panel Provides ports for LAN Insulated USB and Dual USB see Figure 3 24 4 Caster 1x back with free movement and no locking 12 Probe Connectors for 4 RS probes 5 Swivel and full lock caster 1x back 13 Articulated Arm 6 Power inlet connector 14 Probe holders 7 System Base 15 Gel Cup Holde...

Page 94: ...r not insulated 3 HDMI connector not insulated 3 4 4 EMI Protection The Venue has been designed to minimize the effects of Electro Magnetic Interference EMI Many of the covers shields and screws are provided primarily to protect the Venue from image artifacts caused by this interference For this reason it is imperative that all covers and hardware are installed and secured before the Venue is put ...

Page 95: ...s right corner at 1 30 m above the floor and in a distance of 1 meter at all four sides 1 meter above the floor 3 5 3 Electrical Specifications 3 5 3 1 Verification of the Venue Voltage Setting Verify that the mains voltage specified for the Venue is available on site The voltage setting for the Venue is found on a label near the Mains Power Circuit Breaker on the rear of the Venue 3 5 3 2 Electri...

Page 96: ...e valid version of the system standard IEC60601 1 1 Everybody who connects additional equipment to the signal input part or signal output part of Venue configures a medical system and is therefore responsible that the Ultrasound system complies with the requirements of the valid version of IEC60601 1 1 If in doubt consult the technical service department or your local representative for GE 3 5 4 1...

Page 97: ...can be connected or changed any time as described below regardless of whether the system is powered ON or OFF The Venue has four RS connectors Figure 3 25 Probe Connectors on Venue Front End Door CAUTION HANDLE THE PROBE GENTLY WHILE CONNECTING AND DISCONNECTING DO NOT TOUCH THE PATIENT AND ANY OF THE CONNECTORS ON THE ULTRASOUND UNIT SIMULTANEOUSLY INCLUDING ULTRASOUND PROBE CONNECTORS ...

Page 98: ...ting the probe a Do a visual check of the probe pins and system sockets b Remove any dust or foam rests from the probe pins c Verify the probe and the probe cable for any visual damage CAUTION Do not allow the probe head to hang freely Excessive impact to the probe will result in irreparable damage CAUTION To prevent probe connector pins damage or PCB board damage do not use excessive force when c...

Page 99: ...robe ensure that the connector locking handle is positioned to the right 3 Align the connector with the probe port and carefully push into place 4 Push the connector locking handle to the left to secure the probe connector 5 Carefully position the probe cord so it is free to move and is not resting on the floor 3 5 5 3 Disconnect Probes Follow these steps to disconnect the RS probes as applicable ...

Page 100: ...tion describes how to configure the Venue 3 6 2 Venue Configuration 3 6 2 1 EZ Config Setup Wizard The Venue Configuration Wizard enables the user to easily configure the system after SW installation 1 After the system reboots the Venue setup wizard welcome screen appears 2 Select the interface language and tap OK to proceed Figure 3 27 Venue EZ Config Welcome Screen ...

Page 101: ... installation The settings screen appears The Local tab opens by default 4 Wait while system initializes the setup wizard Note For software version 302 x x the initialization duration might take up to 15 minutes due to automatic firmware updates Figure 3 28 Venue Installation Wizard Run Wizard Figure 3 29 Venue Setup Wizard Initialization ...

Page 102: ...red details and tap Save Then proceed to the next tab 6 On the Network tab define the required settings and tap Save Then proceed to the next tab 7 On the Thermal Printer tab configure the printer and tap Save Then proceed to the next tab If Figure 3 30 Venue Installation Wizard Local Tab Figure 3 31 Venue Installation Wizard Network Tab ...

Page 103: ...0 REVISION 6 VENUE SERVICE MANUAL 3 30 Section 3 6 Configuration P R E L I M I N A R Y the printer is already configured skip this step and proceed to the next one Figure 3 32 Venue Installation Wizard Thermal Printer Tab ...

Page 104: ...apter 3 System Setup 3 31 P R E L I M I N A R Y 8 On the InSite tab define the required settings and tap Submit Changes Make sure you fill all mandatory fields highlighted in bold Then proceed to the next tab Figure 3 33 Venue Installation Wizard InSite Tab ...

Page 105: ...items list by moving the On Off sliders to On Then select each Connectivity item to define its properties Tap Apply and then Save Then proceed to the next tab 10 On the OptionKey tab enter the key string to enable the purchased options and tap Save Then proceed to the next tab Figure 3 34 Venue Installation Wizard Connectivity Tab Figure 3 35 Venue Installation Wizard OptionKey Tab ...

Page 106: ...Then proceed to the next tab 12 On the About tab view all settings defined on each tab and click Save to save the entire set of defined settings Then click Exit to exit the Setup Wizard NOTE To reinitialize the EZ Config wizard access the Service tab under the System Configuration menu refer to Open Service Screen on page 3 41 Figure 3 36 Venue Installation Wizard Report Tab Figure 3 37 Venue Inst...

Page 107: ...2 Accessing System Configuration Settings All settings configured by the EZ Config wizard can be modified via System Configuration Settings To access these settings perform the following steps 1 On the Home screen tap Settings Config Figure 3 38 Home Screen Settings Menu 2 Log on as ADM Figure 3 39 Operator Login ...

Page 108: ...5771498 100 REVISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 35 P R E L I M I N A R Y 3 From the Config side menu select System Settings The System Settings screen is displayed Figure 3 40 System Settings ...

Page 109: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 3 36 Section 3 6 Configuration P R E L I M I N A R Y 3 6 2 3 Enter Location Figure 3 41 Enter Location ...

Page 110: ...start The 24 first characters of this name are displayed on the scanning screen s title bar All 64 are displayed on the image properties on saved images 2 Select the Department field and type the name of the department max 64 characters After restart This name will be displayed on the image properties on saved images 3 Select the Echolab field and type the name After restart This name will be disp...

Page 111: ...e Format see 3 in the figure 24 the 24 hour format is used 12 the 12 AM PM hour format is used 4 Adjust the date see 1 in the figure New date is displayed 5 Adjust the time see 2 in the figure New time is displayed 6 Select Default Century 1900 2000 or None see 5 in the figure 1900 the number 19 is automatically displayed when entering the year in the patient date of birth To edit century press BA...

Page 112: ...498 100 REVISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 39 P R E L I M I N A R Y 3 6 2 5 Select Language for User Interface and Online Manuals Figure 3 43 Select Language Units and Controls Sound 2 4 6 1 3 5 ...

Page 113: ...preferred language for the online manual The selected language will be used as soon as the unit has been restarted 4 Select the Secondary check box to enable the multi language keyboard characters support Then from the pulldown menu select the required language The selected language will be used as soon as the unit has been restarted 5 See Table 3 18 on page 3 40 See Table 3 18 on page 3 40 6 See ...

Page 114: ... 1 SW Recovery In cases of abnormal system behavior as a part of troubleshooting process it is recommended to reload the system software To reload the software without affecting patient data tap the SW Recovery button on the Service screen and follow the displayed instructions SW installation media is not required for this process 3 6 3 1 2 Re initiate EZ Config Wizard The EZ Config wizard can be ...

Page 115: ...on or a combination of software options The Software Option Key is specific for each unit NOTE There may be more than one Software Option Key in use depending on the installed options 3 6 5 2 To Install a Software Option Follow these steps to install the Software Option Key 1 Log on as adm 2 Select Admin lower part of window 3 Select the System Admin tab 4 Select New to open the New Key dialog whe...

Page 116: ...possible connection methods as outlined below 3 7 2 Stand alone Venue No network connection needed 3 7 3 Sneaker Net Environment No network connection needed Use removable media to move data from the Venue to another unit 3 7 4 Wireless Connection from Venue to DICOM Server Wireless network adapter must be connected to the Venue system in order to enable wireless network connection ...

Page 117: ...ork printer you should use default delivery settings 3 8 1 Introduction To be able to use the network functions when connected to a hospital network the Venue must have a proper network address Before you can set up the Venue you need to collect some information The Worksheet see sample Connectivity Installation Worksheet on page 2 13 can be used for gathering this information Typical source for t...

Page 118: ...NNNN is the scanner s serial number AE Title Venue Port No Default port number 104 2 Server Config Servers List of servers Buttons Use the buttons to Add Modify or Remove servers 3 Remote Path Used for Save As Export from Q Analysis and for exporting Error Logs with Alt D 4 Configurable Remote Path User Add Secondary Log in Credential 5 Save Settings Select Save Settings to archive any changes you...

Page 119: ... Log on as adm 2 Tap Settings Config Connectivity TCPIP 3 Tap the Network Settings button The Network Connections window opens 4 Open the Control Panel and go to Network and Internet Network and Sharing Center 5 Click or tap the following link Set up a new connection or network The Set Up a Connection or Network wizard is started Figure 3 46 Set Up a New Connection or Network ...

Page 120: ...Security type field select the type of security used by the hidden wireless network Some routers may name this authentication method Depending on the security type you select Windows 10 may or may not ask you to also specify an encryption type In the Security key field enter the password used by the hidden WiFi If you do not want others to see the password you type select the Hide characters check...

Page 121: ...privacy at risk because skilled professionals can intercept this search for the hidden network NOTE If you select the Connect even if the network is not broadcasting check box Windows 10 searches for the hidden network every time it is not connected to a network even if the hidden network is not in your area This may put your privacy at risk because skilled professionals can intercept this search ...

Page 122: ...N 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 49 P R E L I M I N A R Y complete the setup If you are in the range of the hidden WiFi your Windows 10 device automatically connects to it Figure 3 49 Entering Security Information ...

Page 123: ... I M I N A R Y 3 8 4 Changing the AE Title and or Port Number Port No Figure 3 50 AE Title and Port No 1 To change AE Title and or Port No edit the respective fields 2 Select Save settings to store your changes 3 Reboot Venue to activate the settings or continue with other Tcpip set up tasks 4 ...

Page 124: ...vironment This procedure is required for every new wireless network NOTE The WiFi configuration is available only if wireless adapter is connected to the system Do not use any type of wireless network adaptor other than a GE approved adaptor 1 From the Connectivity menu select TcpIp and then tap Wireless Settings 2 In the Wifi Configuration window view the list of available wireless networks Figur...

Page 125: ...onnected to the USB port located on the rear side of the cockpit monitor NOTE It is possible to connect the Wireless Network adapter both when the system is powered ON and after it is powered OFF All software drivers for the Wireless Network are pre installed for the designated Wireless Network adaptor only NOTICE IMPORTANT During Stand by mode it is NOT recommended to introduce or remove USB devi...

Page 126: ... Setup 3 53 P R E L I M I N A R Y Section 3 10 InSite Default Machine Contact Setup 3 10 1 Time Required 5 min 3 10 2 Preparations No dongle or service password needed 3 10 3 Setup Procedure 1 On the main screen tap Settings Service InSite Figure 3 52 Accessing InSite SW ...

Page 127: ...n 3 6 Configuration P R E L I M I N A R Y 2 From the InSite ExC menu select Request for Service 3 In the Contact GE Service Browser window open the Users tab Figure 3 53 InSite ExC Main Menu Request for Service Figure 3 54 Contact GE Service Browser Window Users Tab ...

Page 128: ...er 5 Enter person s contact information Red asterisk indicates required field Then tap Add User 6 Select the check box for the person who will be the Default Contact Tap Set Default Machine Contact Figure 3 55 Users Tab Add User Figure 3 56 Users Tab Adding User Figure 3 57 Users Tab Selecting Default Machine Contact ...

Page 129: ...56 Section 3 6 Configuration P R E L I M I N A R Y 7 Tap Set Default Machine Contact 8 The Default Machine Contact will be indicated in red text The setup is complete Figure 3 58 Users Tab Default Machine Contact Selected Figure 3 59 Users Tab Setup Complete ...

Page 130: ... Installation Acceptance Test Criteria A Venue is ready for use after the system has been configured successfully in accordance with the information provided in Chapter 3 System Setup this chapter 3 11 2 User s Manual s Check that the correct User Manual s or storage media with User Manuals per software SW revision and language for the system is included 3 11 3 Product Locator Installation Card NO...

Page 131: ...udes the General Procedures and the Functional Checks General Procedures is a collection of commonly used procedures that are available by cross references from other parts of this manual Functional Checks is a collection of procedures for quickly checking major functions of the Venue scanner and diagnostic instructions using the built in service software These checks can be a great asset in deter...

Page 132: ...ely in place The covers are required for safe operation good Ultrasound system performance and cooling purposes WARNING WARNING Energy Control and Power Lockout for Venue When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts 1 Follow LOCK OUT TAG OUT procedures 2 Turn off the breaker 3 Unplug the Ultrasound system 4 Maintain control of the Ultrasoun...

Page 133: ...ot disconnect any USB mass storage device from the system unless a Software Installation procedure is required and the appropriate software installation storage device is connected DANGER ALWAYS CONNECT THE ULTRASOUND SYSTEM TO A FIXED POWER SOCKET WHICH HAS THE PROTECTIVE GROUNDING CONNECTOR DANGER NEVER USE A THREE TO TWO PRONG ADAPTER THIS DEFEATS THE SAFETY GROUND DANGER ENSURE THAT THE POWER ...

Page 134: ...ng label on the rear of the Venue ultrasound scanner 5 Connect the AC Power Cable female plug to the Power Inlet AC IN at the rear of the unit see Figure 4 1 6 Screw the plastic cable clamp into the hole provided in the support column centrally located at the rear of the system 7 Connect the other end of the AC Power Cable male plug to a hospital grade mains power outlet with the proper rated volt...

Page 135: ...own after the application is loaded b The Back End Processor and the rest of the scanner starts with the sequence listed in the next steps 1 Back End Processor is turned ON and starts to load the software 2 The Start Screen is displayed on the monitor 3 A start up progress bar indicating the time used for software loading is displayed 4 The software initiates and sets up the Front End electronics ...

Page 136: ...e Dongle Select this button when you want to exit to the Windows Desktop NOTE If you need to restart Venue when logged on to the Windows Desktop ensure that you do a complete power down Shut Down This is required to power up the Front End Processor Shutdown Use this button to shut down the system The entire system will shut down If the Shutdown button is greyed out or the system cannot be shutdown...

Page 137: ...ower switch located on the rear lower part of the system to the OFF position Wait for 10 seconds and re set the power switch to the ON position Wait for about 15 Sec before the System is operational Cancel Use this button to exit from the System Exit menu and return to the previous operation Figure 4 5 Power ON OFF Switch and Power AC IN Socket Rear Panel AC IN Power ON OFF Switch Power Inlet ...

Page 138: ... screen In order to disable enable the functionality of the On Off physical button open Settings Config Imaging Global Settings and check or uncheck the Disable physical power button box When the physical power button functionality is disabled pressing the physical On Off button on the cockpit will trigger the following message Figure 4 6 Disable Physical Power Button in Config Screen Figure 4 7 P...

Page 139: ... VENUE SERVICE MANUAL Chapter 4 General Procedures and Functional Checks 4 9 P R E L I M I N A R Y To switch off the system tap Settings Off This action will trigger the display of the System Exit window Figure 4 8 Software Off Button ...

Page 140: ...cess to the setup menus Example To select a printer As default no password has been set for USR Just type the name USR and select Login If you log on as ADM you will have access to all setup menus and service adjustments on the Venue system Example Adjust network and connectivity settings NOTE It is possible for the administrator ADM to establish new users and set unique passwords for each user in...

Page 141: ... can be adjusted To tilt the cockpit monitor use the lever handle and adjust the cockpit monitor to the desired angle 4 2 8 Moving and Transporting the Venue Ultrasound Scanner 4 2 8 1 The Casters Wheels Control The pedals located between the left front and right back wheels of the Venue control the swivel and brake on the wheels Examine the wheels frequently for defects to avoid breaking or jammi...

Page 142: ...work 4 Secure the unit s power cable 5 Place all probes in the probe holders Ensure that the probe cables do not protrude from the unit or interfere with the wheels 6 Ensure that no loose items are left on the unit 7 Unlock the casters Related information Cockpit Monitor Position Adjustment on page 4 11 Power Shutdown on page 4 6 CAUTION Unlock the brakes on the front and back wheels when preparin...

Page 143: ...en transporting the Venue by vehicle In addition to the precautions listed earlier follow the steps below Center the cockpit monitor and move it to its down position Disconnect all probes and secure them in their boxes Ensure that the transporting vehicle is appropriate for the unit s weight Park the vehicle on a level surface for loading and unloading Secure the Venue while it is on the lift to p...

Page 144: ...ue operates as intended The functional checks may also be used during troubleshooting 4 3 2 Performance Checks 4 3 2 1 Test Phantoms The use of test phantoms is only recommended if required by your facility s customer s QA program 4 3 3 2D Mode B Mode Checks 4 3 3 1 Introduction The 2D Mode is the system s default mode 4 3 3 2 Preparations 1 Connect one of the probes 2 Turn ON the Venue The 2D Mod...

Page 145: ... Depth control to adjust the range to be imaged Use the Modes to toggle between the scanning system s main modes Use the Frequency control move to lower frequency to increase penetration Press to activate the Needle Visualization mode Gain Angle and Tilt can be activated The needle function only applies to linear probes and the C1 5 RS convex probe Use Split to split the screen into two allowing t...

Page 146: ...natomical M Mode can also be used with previously acquired digitally stored 2D images More than one heart cycle should be stored if performing M Mode in post processing 2 Position the cursor over the required area of the image 3 Rotate the solid cursor line to the desired direction 4 3 4 3 3 Curved Anatomical M Mode 1 In M Mode press Curved AMM 2 Position the starting point of the time motion curv...

Page 147: ... 3 4 5 1 Color 2D 1 From an optimized 2D image press Color 2 Position the ROI frame over the area to be examined 3 Press Select The instruction Size should be highlighted 4 Adjust the dimension of the ROI 4 3 4 5 2 Color M Mode 1 From M Mode press Color 2 Position the color area in the M Mode display 3 Press Select The instruction Size should be highlighted 4 Adjust the dimension of the color area...

Page 148: ... limit settings Adjust the Compress setting to balance the effect of stronger and weaker echoes and obtain the desired intensity display Adjust Frequency to optimize flow display Higher setting will improve resolution and the lower setting will increase the depth penetration Adjust Frame rate to a higher setting to improve motion detection or to a lower setting to improve resolution NOTE Frequency...

Page 149: ...h and low Pulse Repetition Frequency PRF When the Doppler PRF is raised beyond a certain limit more than one Doppler gate is displayed on the screen Press Auto on the Control panel to activate Automatic Spectrum Optimization ASO ASO is used to automatically adjust baseline and scale of the PW CW spectrum to optimize the spectral display It will avoid the display of a folded spectrum and stretch th...

Page 150: ...erator Panel A list of the connected probes will pop up on the screen 2 Select the desired probe An application menu for the desired probe is listed on the screen 3 Browse to the desired application Press Select to launch the application To change application without changing the current probe press Appl on the Operator Panel The selected application starts 4 Verify no missing channels All channel...

Page 151: ...eeze to freeze the cineloop 4 Scroll through the acquisition and find the sequence of interest 5 Adjust Cycle select to move from heart beat to heart beat and select the heart cycle of interest 6 Adjust Num cycles to increase or decrease the number of heart beats to be played back 7 In Freeze press Set left or Set right to set the corresponding cineloop boundary to the current frame 8 Adjust Left ...

Page 152: ...Audio Check 4 3 8 1 Introduction The system default settings is to play audible feedback while pressing any action key i e storing an image changing mode typing text etc This functionality can be disabled in the Config screen under system settings tab Figure 4 11 UDI label attached to Software installation media Label ...

Page 153: ...alled and hooked up at the Windows level 1 On the Venue application main screen tap Settings Config Connectivity Additional Outputs Figure 4 12 In the Printer Setup section verify that the selected printer is Sony UP D711MD Verifying Selected Printer Table 4 20 Printer Checks Step Task Expected Result s 1 When scanning in 2D Color Mode press Freeze to stop image acquisition Image scanning stops wi...

Page 154: ...3 In the Sony UP D711MD window select Printer Properties Printer Window Figure 4 14 In the Sony UP D711MD Properties window tap Print Test Page this will send a print to the printer bypassing all of the Scanner software Printing Test Page Figure 4 15 The message A test page has been sent to your printer appears Sony Printer Notification ...

Page 155: ...ill see an incomplete Test Page printed out This is normal 4 3 9 3 ECG Checks The system automatically detects the ECG module once connected The indication for the user will be displayed in the Peripherals status area Figure 4 16 Peripheral Status Area In order to check the functionality of the ECG device the ECG leads need to be connected to person body or to an ECG simulator ...

Page 156: ...Ensure that you do not apply too much force to move the cockpit and that the movement is smooth During the movement up and down make sure the image displayed on the monitor does not present any disturbance 3 Move down the cockpit to its maximum lowest position until it is locked in place Make sure the arm is locked and no movement is observed when moderate force is applied to the cockpit upwards a...

Page 157: ...ck handle it is located under the cockpit on the front of the Venue system to release the tilt locking mechanism Make sure the locking mechanism is locked when reaching 45o The tilt locking mechanism is released 7 Check the arm pan as follows Hold the cockpit monitor and move it to its maximum pan left and then to its maximum pan right positions Ensure that you do not apply too much force to move ...

Page 158: ...lock 1 Push down on the upper grey lever to unlock the brake and swivel on the casters Push and pull the unit right left backwards and forwards Ensure that the wheels move freely in all directions Check the wheels for wear and tear and replace if necessary 2 Press the foot brake red lever down on each caster wheel to lock the swivel and full lock wheels in position Push and pull the unit right lef...

Page 159: ... 4 29 P R E L I M I N A R Y 4 3 11 Electrical Tests 4 3 11 1 Ground Continuity Test 1 Using multimeter ensure there is a ground continuity between the ground prong of the IPP power connector and the arm ground cable connection Figure 4 17 Ground Continuity test Ground prong IPP Power Connector System back ...

Page 160: ...Theory 5 1 P R E L I M I N A R Y Chapter 5 Venue Components and Function Theory Section 5 1 Overview 5 1 1 Purpose of Chapter 5 This chapter explains Venue system concepts component arrangement and sub system functions It also describes the power distribution system the cabling system and probes ...

Page 161: ...tile and depending upon the installed software can be used for various imaging modes These include 2D Gray Scale and 2D Color Flow imaging M Mode Gray Scale imaging Color M Mode Doppler Different combinations of the above modes The Venue system main hardware components are configured as illustrated in Figure 5 1 NOTE For a detailed description of Venue system operating modes refer to the Venue Use...

Page 162: ...lectronics The Venue system internal electronics are divided into three Front End Unit see page 5 18 Back End Processor see page 5 22 System Power Distribution see page 5 26 Venue system internal electronics comprise an eTower that contains both the Front End Unit and the Back End Unit The Interconnecting signals between the two sections are routed internally via PCIe cable All the interconnection...

Page 163: ...ly unit and batteries compartment Section 5 3 Options For a list of the available options see Optional Peripherals on page 9 18 For descriptions of the options see the User Manual For Installation instructions see Options Setup on page 3 52 For configuration see Options Setup on page 3 52 Figure 5 2 Venue Ultrasound Scanner System Design ...

Page 164: ... ways Connection via a Local Area Network LAN set up for this special use only Connection via the hospital network Local Area Network LAN or Wide Area Network WAN The database on the review station is used as the master and images are stored directly to the review station s storage In this configuration the ultrasound scanner is just an intermediate acquisition unit which after completion of a stu...

Page 165: ...on InSite ExC is your direct link with a GE Online Service Engineer or Applications Support Engineer or a Request for Service via the InSite ExC link at the bottom of the display screen 5 5 2 InSite ExC Icon The InSite ExC icon is located under the Service Settings Figure 5 3 InSite ExC Icon Clicking on the Icon brings up the InSite ExC menu ...

Page 166: ... the customer to generate a Service or Apps request Connect to GE Increase polling rate Click this icon to increase the contact Poll Rate from 15 minutes to 15 seconds The increased polling rate continues for 15 minutes Connect Clinical Lifeline Changes the polling rate the same way as Connect to GE and in addition Switches to disruptive mode Starts Virtual Console Observation does not provide acc...

Page 167: ...spatch directly to GE Service after you fill in the following information Items with a red asterisk Problem type Problem area Problem description Send 3 After you have completed filling in all of this information press Send to initiate the Request for Service Figure 5 5 Request for Service Contact Information After you press Send the following pop up appears Figure 5 6 Request for Service Confirma...

Page 168: ... The Venue system automatically submits a Request for Service once a software update package was downloaded successfully Applicable for software version 302 X X only All RFS requests can be monitored under the Machine Queue tab Figure 5 8 Machine Queue In addition you can use the Users screen to identify your institution s point of contact for service dispatches ...

Page 169: ...your system but cannot perform any service related functions depending on whether InSite has connected or not connected There are two Non Disruptive states If you see a black and white icon InSite ExC is activated but not open for Technical Support access If you see a yellow icon InSite ExC is activated and the Technical Support person can look around on your system but cannot perform any service ...

Page 170: ...10 IoT 64 bit The cockpit monitor module contains the following elements Four probe holders and gel cup holder Power ON OFF touch button SLEEP mode indicator AC Power LED Indicator Speaker Auxiliary display Mechanical Interface to Arm module Light Sensors x2 LCD display 19 5 4 1280x1024 pixels Touch module multi touch surface 5 6 1 1 Probe Holders The cockpit monitor contains four probe holders an...

Page 171: ...m power on Orange indicates system power off The On Off touch button is sensitive to A bare finger A Latex gloved hand A Latex gloved hand with ultrasound gel 5 6 1 3 Sleep Mode Indicator The Sleep Mode indicator is located in the upper left corner of the cockpit monitor to prevent accidental activation Figure 5 10 Sleep Mode Indicator The system Sleep Mode button has one color for sleep status in...

Page 172: ...or Possible States System State LED Indicators No Power AC main power switch is off or power cord is disconnected batteries are drained or disconnected System is Off AC main power switch is on and power cord is connected System is Off AC main power switch is off or power cord is disconnected System is On on AC AC main power switch is on and power cord is connected System is On on Batteries AC main...

Page 173: ...th scrolling browsing and scaling of the display as required To facilitate comfortable positioning for the operator the height of the Main Display can be adjusted as required For optimal viewing the screen angle is adjustable tilting may be between 0o and 50o At the tilt of 45o the tilt mechanism is automatically locked and the cockpit can be tilted by additional 5o only To release the tilt clocki...

Page 174: ... LCD Controller and Touch Module Configuration Interface 5 6 1 9 1 Input Video Interface A Display Port standard receptacle connector is used for carrying video signals and preparatory none video signals Table 5 4 Touch Screen LCD Display Optical Characteristics Item Value Contrast Ratio Min 600 Horizontal Viewing Angle Min 170 degrees CR 10 Vertical Viewing Angle Min 170 degrees CR 10 ...

Page 175: ... Battery Status Indication When the battery is charging If the AC power cable is plugged in the AC mains power indicator is lit green and the battery is being charged until reaching full charge The background color of the battery status display changes according to the remaining operational time as follows Green remaining operation time is more than 90 minutes Yellow remaining operation time is be...

Page 176: ...Venue ultrasound unit interface panel showing external peripheral accessory connectors 1 Ethernet LAN connector 1000 Base TX Ethernet IEEE 802 3 3kV insulation 2 Dual USB 3 0 connector not insulated 3 HDMI connector not insulated NOTE Non insulated I O can be populated either by certified medical devices or a self powered device powered by the system not by external AC All other devices should be ...

Page 177: ...e following modules Front End Power Supply T FEPS See Front End Power Supply T FEPS on page 5 20 Control Front End T CFE Module includes four Transmit and Receive Modules TRX32 See Front End T CFE Board on page 5 20 and Back End Processor on page 5 22 Probe Selector Module T PSB For interconnection of probes see Probe Selection Board T PSB on page 5 21 Figure 5 15 Front End Unit Location of Compon...

Page 178: ...E Module The PSB is connected to the T CFE via 3 connectors The Interface between the T CFE and the PSB modules consists of the following Dedicated I2C channel for Board ID VPD Control Signals I2C channel for Probe ID Reading and PSB Debug Information LVDS channels for Matrix Probe Setup and Status Reading PSB DC Supply Probe Low Voltage and High Voltage Supply 5 8 3 Signal Flow The TRX32 Boards i...

Page 179: ...ransmit receive all the channels data signals to from the four TRX32 boards assembled on the T CFE board In addition it contains all the control and status lines required for HV MUX Setup Probe selection and reading Probe status and information 5 8 5 1 T TRX32 Boards Transmitter Receiver 5 8 5 1 1 General Description The Venue system has 128 Receive Transmit channels divided into 4 boards each con...

Page 180: ...a mechanical and electrical interface for 4 RS probes The T PSB supports the following Probes with up to 128 elements directly 192 element probes using on board HV multiplexers The PSB receives all the required control lines from the T CFE board The analog receive transmit lines are routed to the T CFE board using two dedicated connectors A third connector is used for power and control 5 8 6 1 T P...

Page 181: ... Back End Processing BEP Module Solid State Hard Drive SSD 5 9 2 Back End Processing BEP Module The Main BEP module is the System Computer responsible for operating the Venue Scanner application under a Windows 10 operating system The BEP board contains the following Interfaces Expansion PCI Express x16 USB USB 2 0 USB 3 0 Video Digital video outputs Audio Digital audio output and analog audio inp...

Page 182: ... signals between the BEP and the Front End BEP Interface Panel this panel allows external connections of multiple interfaces by extension cables that connects the internal ports to the IP panel The BEP Interface Panel is located on the rear side of the Venue ultrasound scanner and contains the following connectors HDMI port supports connection of external monitor 2 USB 3 0 ports support for extern...

Page 183: ...DIRECTION 5771498 100 REVISION 5 VENUE SERVICE MANUAL 5 24 Section 5 9 Back End Processor P R E L I M I N A R Y 5 9 3 BEP Block Diagram Figure 5 18 BEP Block Diagram ...

Page 184: ...the boards in the system and redirect them to their relevant destinations The Back End Board BIB module responsible for the scanner Interfaces comprises three main Interfaces Back End mITX standard via the Docking Interface Csound Front End CFE Board Main Power Board MPB 5 9 5 Solid State Hard Drive The Hard Disk is the main storage device of the Venue ultrasound unit The Hard Disk which is contro...

Page 185: ...t supplies Power to the system when AC power line is not present or in the event of AC power failure 12 16 8V Input filter unit IPP before PSU for filtering AC line noises 5 10 2 AC Distribution The PSU module converts the AC power line 100 240V AC to 18V DC 400 W this module is the main power supply for the Venue ultrasound scanner 5 10 3 DC Power Distribution The main DC Power supplied from the ...

Page 186: ...nd battery power 12v 16 8v The Embedded Controller SOM is responsible for the following functions Control the power up and wake up sequences Battery Monitoring and Battery Charger control Interface to the PSU Interface to the T CFE Power sequencing and Monitoring Device Back End Thermal Management 5 10 5 Rechargeable Battery Pack The battery pack consists of Li Ion cells protection and fuel gauge ...

Page 187: ...erature is between 0 C and 40 C 32 F and 104 F and discharge the batteries between 20 C and 50 C 4 F and 122 F Do not short circuit the battery by directly connecting the battery terminals with metal objects Do not heat the battery or incinerate Do not expose the battery to temperature over 60 C 140 F Keep it away from fire and other heat sources Do not charge the battery near a heat source e g fi...

Page 188: ...CFE Fan located on the top of the Front End module MPB Blower Fan located on the on the rear side of the system behind the MPB module All fans are controlled and have variable speed The cooling requirement for the Venue ultrasound scanner with monitor and on board peripherals is up to 2000 BTU h This figure does not include cooling needed for lights people or other equipment in the room NOTE Each ...

Page 189: ...sible for the acquisition of the ECG analog signal inputs 5 12 2 External Peripherals 5 12 2 1 Wi Fi Adapter The Wi Fi adapter allows the user to connect the system to network wireless The Wi Fi adapter is connected at the back side of the cockpit monitor via a L shape USB adapter 5 12 2 2 Bar code reader The bar code reader allows the user scanning patient ID from a printed bar code label and sca...

Page 190: ... come from its integration with iLinq The following sections contain a brief introduction of iLinq s features 5 13 4 1 Service Desktop The Service Platform and other Service software use the iLinq web server and the Internet Explorer browser 5 13 4 2 Request for Service NOTE This feature that allow the customer to contact the GE OnLine Center are available for Warranty and Contract customers only ...

Page 191: ...em 5 13 5 3 Access Security The service interface has different access and security user levels Each user is only granted access to the tools that are authorized for their use NOTE A Service Dongle part number 066E0703 is necessary for use by GE Service when performing proprietary level diagnostics OnLine Center access to the scanner requires the password and they must have Disruptive permission a...

Page 192: ...5 33 P R E L I M I N A R Y Every access request whether successful or not will be logged into a service access log that is viewable to authorized users Related information Data Management on page 4 10 See Chapter 7 Diagnostics Troubleshooting for more information This page was intentionally left blank ...

Page 193: ...Venue Section 6 2 Power Supply Adjustments There are no adjustments on the power supply The DC Power is self regulated If a voltage is outside the specified range it means that something is wrong either with the power supply itself or with one or more of the units connected to that specific power outlet When an error occurs the power will be turned OFF immediately Section 6 3 Cockpit Monitor Adjus...

Page 194: ...ncreased decreased in the following axes red arrow shows location of the controlling screw per axis 6 4 2 Tilt Resistance Adjustments 1 At the back of the cockpit disconnect the Wi Fi dongle if exists release four screws and remove the transparent cover and the Wi Fi dongle 2 Remove the cockpit rear cover if exists disconnect four screws using appropriate Phillips screwdriver Figure 6 20 Optional ...

Page 195: ... Chapter 6 Service Adjustments 6 3 P R E L I M I N A R Y 3 Using 8mm Allen key and 19mm open wrench release tighten the Allen screw NOTE Ensure there is no tilt autonomic movement Figure 6 22 Cockpit Rear Cover Removal Figure 6 23 Tilt Movement Adjustment ...

Page 196: ...s 1 Tighten Release the 8mm socket at the back of NOTE Ensure there is no pan swivel autonomic movement 6 4 4 Arm Vertical Movement Adjustments 1 Move the cockpit to one side until the Alen screw inside the arm is accessible 2 Using a 5mm Alen key tighten release the Alen screw Figure 6 24 Pan Movement Adjustment Figure 6 25 Arm Vertical Movement Adjustment ...

Page 197: ...re there is no arm autonomic movement 6 4 5 Arm Rotate Resistance Adjustments 1 At the back of the system release the nut under the riser cover 2 Tighten or release the 4mm Allen screw to adjust the required arm movement resistance NOTE Ensure there is no arm autonomic movement Figure 6 26 Arm Rotate Movement Adjustment ...

Page 198: ...gnostics Troubleshooting Section 7 1 Overview 7 1 1 Purpose of Chapter 7 This chapter describes how to setup and run the tools and software that help maintain image quality and system operation Very basic host system and board level diagnostics are run whenever power is applied Some Service Tools may be run at the application level ...

Page 199: ...OF CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING If the covers are removed from an operating Venue some metal surfaces may be warm enough to pose a potential heat hazard if touched even while in shutdown mode Use all Personal Protection Equipment PPE such as gloves safety shoes safety glasses and kneeling pad to reduce the risk of injury ...

Page 200: ...e The Venue Companion is a Windows based application that provides visual guides for customer replaceable units CRU This application is distributed as part of the software installation media 7 3 2 System Diagnostics The following steps are different for R1 and R2 configuration For R1 systems with software version 301 X X Figure 7 27 Venue Companion Application ...

Page 201: ...7 2 Service Safety Considerations P R E L I M I N A R Y Press and hold ON OFF Standby buttons simultaneously For R2 systems with software version 302 X X On the Home screen tap Settings Service Diag Figure 7 28 Venue R1 Systems Accessing Diagnostics Screen Figure 7 29 ...

Page 202: ...Select Config F2 About screen Applications Software Application Software revision Software medium s part number System Software System Software revision Software medium s part number 7 4 4 Collect a Trouble Image with Logs For R1 systems with software version 301 X X If the system should malfunction press the Alt Logs Alt D for external keyboard keys simultaneously This will collect a screen captu...

Page 203: ...ion 302 X X On the Home screen tap Settings Log Logs NOTE You MUST select one of the available devices as the destination device if it is to be different than the default Export directory on the hard drive The screen capture is a bitmap which eliminates the possibility of artifacts from compression Figure 7 30 Venue R2 Systems Accessing Diagnostics Screen ...

Page 204: ... creation of a log file based on a selected bookmark or for a user configurable time frame Different type of information can be selected to be part of the log file The disk drive installed in the Venue system is based on Solid State technology SSD and does not need to be defragmented 1 Type description of issue here 2 Select if you ve had a system lockup after restart 3 Select where to store the r...

Page 205: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 7 8 Section 7 4 Gathering Troubleshooting Data P R E L I M I N A R Y 7 4 4 2 Figure 7 32 Exiting Service Diagnostic Tests ...

Page 206: ...ion as possible about the conditions when the noise appear Is the noise present all the time after some time of use After how long time at special times of the day or night When at all locations in the hospital or only in one room area from time to time no special pattern of time is observed 7 5 4 Overview of Types of Noise There are different types of noise Use the information next to classify th...

Page 207: ...ial occasions 7 5 4 4 Self generated Noise Generated inside the Ultrasound system Example Color Noise in the near field Self generated noise will not change if you touch the scanner or the probe Self generated noise may be due to either heat problems hardware problems software problems 7 5 4 5 Heat Problems Heat problems are usually starting when the Venue has been ON for some time If the Venue ha...

Page 208: ... shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Sites with a mains power system without a defined Neutral The dedicated line shall consist of one phase two lines not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet 7 5 6 Different System Try another Venue scanner at the sam...

Page 209: ...ooting 7 6 1 Purpose of this Section In this section you will find Audio troubleshooting procedures and hints to be used if there is no Doppler or system sound 7 6 2 Audio Troubleshooting Procedure 1 In the application 1 Enter config system settings 2 Make sure that the mute control sounds is not checked Figure 7 33 Verify Mute Control is NOT Checked ...

Page 210: ...100 REVISION 6 VENUE SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 13 P R E L I M I N A R Y 3 Enter CW mode and select volume menu right side increase the volume 4 Exit to windows Figure 7 34 Increase the volume ...

Page 211: ...R E L I M I N A R Y 2 In the windows screen 1 Tap the speaker icon move the slider to 100 make sure it is not muted 2 Tap and hold the speaker icon and select playback devices make sure the default device is as seen below Figure 7 35 Increase the Windows volume Figure 7 36 Verify selected playback device ...

Page 212: ... SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 15 P R E L I M I N A R Y 3 Tap and hold the default device and select test 4 While the system is turned on disconnect and reconnect the cockpit black connector Figure 7 37 Select Test ...

Page 213: ...s not Recognized Procedure 1 Reseat the PSB TCFE and TRX boards according to Venue service manual 2 If the issue is not resolved this may be a probe issue In this case perform the following steps for each of the probes individually 1 disconnect all probes except one probe and restart the system 2 If the issue persists continue checking this probe on all the other ports to decide if it is a probe i...

Page 214: ...n This section manages the most common scenarios where the system does not startup 7 8 2 Contents in this Section Cockpit ON OFF Button turns orange BEP BIOS Jumper Setup 7 52 BEP Fan does not Spin ON OFF button is Orange MPB to BIB Connection 7 53 Application error Abnormal behaviour Front End Connections 7 54 No Auxiliary Display 7 56 System Stuck in Stand By Mode verify T_CFE Board Revision 7 5...

Page 215: ...om The system does not turn ON cockpit on off button turns green for a few seconds and then turns orange Suggested solution Reset BEP BIOS jumper 1 Reset BIOS by changing the jumper position as shown in the picture move from the 2 outer pins to the 2 inner pins 2 Turn on the system and then off again 3 Reset the jumper to its original position Figure 7 38 Change Jumper Position to Reset BIOS ...

Page 216: ...tion Symptom T CFE and T FEPS LEDs turn On and then off BEP fan does not spin The On Off button on the cockpit turns green for a second and then turns orange Suggested solution Reset MPB to BIB connection 1 Verify BEP jumper is in place 2 Reseat MPB to BIB cable Figure 7 40 MPB to BIB cable Reseat 3 Check BIB connector for bent pins 4 Replace BIB board Figure 7 39 T CFE LEDs T FEPs LEDs ...

Page 217: ...al system behavior occurred Suggested solution Reset front end connections 1 Inspect the BIB blue cable if damaged replace the BIB board 2 Reset the BIB blue cable on the T CFE side disconnect and re connect Figure 7 41 Reset BIB cable 3 Inspect and re connect MPB to T FEPS power cable Replace if damaged Figure 7 42 Reset T FEPS power cable 4 Check power cable to Frontend If damaged replace power ...

Page 218: ...splay LEDs are off when turning on the system Suggested solution Reset front end connections 1 If fans are spinning a Reset cockpit cables b Reset MPB c Replace cockpit 2 If fans are NOT spinning a Turn off the switch located at the rear part of the system for 30 seconds and turn on again b Reset cockpit cables c Reset MPB d Replace MPB Figure 7 43 No auxiliary display ...

Page 219: ...and By Mode verify T_CFE Board Revision Symptom System Stuck in Stand By Mode Suggested solution verify T_CFE board revision 1 Verify T CFE board P N 5764675_X revision X is 5 or higher the label is found under the black metal bracket on the back side of CFE board 2 If needed install R2 302 40 0 SW version or higher Figure 7 44 CFE board revision 5 ...

Page 220: ...tify if there is a real failure of the MPB perform the following procedure 7 9 2 MPB Diagnostics Procedure Troubleshooting 3 When running MPB Diagnostics if the following screen appears indicating the remaining scan time is less than 20 secRun diagnostic test again with fully charged batteries Note This message will be displayed only in service level diagnostics with service dongle When running us...

Page 221: ...nt procedures for Venue system parts as outlined below NOTE The illustrations provided in this chapter are for illustration purposes only and are subject to change without notice 8 1 2 Visual Guide The Venue Companion is a Windows based application that provides visual guides for customer replaceable units CRU This application is distributed as part of the software installation media Figure 8 45 V...

Page 222: ...enue ultrasound unit as described in Power Shut Down 2 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 2 1 4 Basic Storage Basket Large Storage Basket Removal Procedure 1 Pull the basket upwards and remove it as shown below 8 2 1 5 Basic Storage Basket Large Storage Basket Installation Procedure 1 Install the new Basic Storage Basket Large Stor...

Page 223: ...s described in Power Shut Down 2 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 2 2 4 Power Cable Holder Removal Procedure 1 Move the Power Cable Holder to the center of the Halo Handle and remove the Power Cable Holder by pushing it upwards 8 2 2 5 Power Cable Holder Installation Procedure 1 Install the new Power Cable Holder FRU Part Refer t...

Page 224: ... Item Label Item 1 Cover MPB Door 10 Cover Front Base FRU 2 Cover Mid Thermal Baffle FRU 11 Cover Left Side Base FRU 3 Cover Scanner Door FRU 12 Cover Right Side Base FRU 4 Cover Right Side eTower FRU 13 Cover Right Base Bottom FRU 5 Cover Left Side eTower FRU 14 Cover Riser FRU 6 Cover Upper eTower Front FRU 15 Cover RS Probe Connector FRU 7 Cover Lower Front eTower FRU 16 Cover Riser Thermal FRU...

Page 225: ...ECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 8 Halo Handle FRU 17 Cover Left Base Bottom FRU 9 Cover Printer Insert FRU 18 Cover Cockpit Rear Cables CRU Table 8 2 Venue Ultrasound Unit Covers Label Item Label Item ...

Page 226: ...cement cover it is necessary to fit these in the appropriate positions Refer to the illustrations provided in the specific Cover Replacement procedure DANGER ELECTRICAL HAZARDS EXIST AT SEVERAL POINTS IN THE SYSTEM FAMILIARIZE YOURSELF WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT LEVELS BEFORE REMOVING ANY OF THE COVERS DANGER DO NOT WEAR THE ESD WRIST BAND STRAP WHEN REMOVING PARTS FROM THE POWER...

Page 227: ...s described in Power Shut Down 8 3 2 4 Lower Front eTower Cover Removal Procedure 1 Gripping the recess at the center of the cover pull the lower front eTower cover out towards you to remove it from the system 8 3 2 5 Lower Front eTower Cover Installation Procedure 1 Holding the recess with one hand push the top of the cover upwards and inwards with the other hand until clicks into place FRU Part ...

Page 228: ...arations Shut down the Venue ultrasound unit as described in Power Shut Down 8 3 3 4 Left Side eTower Cover Removal Procedure 1 Remove all accessories 2 Remove the RS Probe Cover 3 Grip the rubber holder and pull the cover toward you to remove it from the system FRU Part Refer to Table 9 8 on page 9 7 Accessories Replacement Procedures RS Probe Cover Removal Procedure Figure 8 3 Removing the Left ...

Page 229: ...side eTower cover and make sure the eTower pins go inside the dedicated holes on the inner side of the cover 2 Install the RS Probe Cover 3 Install all accessories Figure 8 4 Installing the Left Side eTower Cover RS Probe Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in Lower Front eTower Cover Replacement Procedure on page 8 216 eTower pin ...

Page 230: ...he right side eTower cover removal procedure is identical to the left side eTower cover removal procedure 8 3 4 5 Right Side eTower Cover Installation Procedure 1 The right side eTower cover installation procedure is identical to the left side eTower cover installation procedure 2 Install all accessories FRU Part Refer to Table 9 8 on page 9 7 Accessories Replacement Procedures Left Side eTower Co...

Page 231: ...Venue ultrasound unit as described in Power Shut Down 8 3 5 4 Mid Thermal Baffle Cover Removal Procedure 1 Grip the cover at its bottom and pull the cover toward you to remove it from the system 8 3 5 5 Mid Thermal Baffle Cover Installation Procedure 1 Reinstall the cover FRU Part Refer to Table 9 8 on page 9 7 Figure 8 5 Removing the Mid Baffle Thermal Cover Perform the checks listed in Mid Therm...

Page 232: ...nt Cover Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower cover Right Side eTower Cover and Left Side eTower Cover 3 Remove the Upper eTower Cover 4 Release the two Phillips screws and remove the Printer Insert Cover or Printer Frame Cover depending on system configuration to separate it from the Upper eTower Front Cover FRU Part Refer to T...

Page 233: ...ps screws 2 Install the Upper eTower Cover on the scanner 3 Refit the following covers Lower Front eTower cover Right Side eTower Cover and the Left Side eTower Cover and RS Probe Cover 4 Install all accessories Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Accessori...

Page 234: ...wer Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 3 7 4 Printer Insert Cover Removal Procedure 1 Remove all accessories 2 Remove the Upper eTower Front Cover 3 Release the two Phillips screws and remove the Printer Insert Cover FRU Part Refer to Table 9 8 on page 9 7 Accessories Re...

Page 235: ...ion Procedure 1 Install the Printer Insert Cover on the Upper eTower Cover and secure with two Phillips screws 2 Install the Upper eTower Cover on the scanner 3 Install all accessories Upper eTower Front Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in Printer Insert Cover Replacement Procedure on page 8 216 ...

Page 236: ...emoval Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower cover RS probe cover Right Side eTower Cover and the Left Side eTower Cover 3 Loosen the four captive screws two on each side grip the MPB Door Cover at the bottom and pull toward you to remove it FRU Part Refer to Table 9 8 on page 9 7 Accessories Replacement Procedures Lower Front eTower Cover Removal Proc...

Page 237: ...llowing covers Lower Front eTower cover Right Side eTower Cover and the Left Side eTower Cover 3 Install all accessories Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in MPB Door Cover Replacement Procedure...

Page 238: ...h the wheels in the locked position 8 3 9 4 RS Probe Cover Removal Procedure 1 Grip the RS Probe Cover on the recess on its bottom and gently pull it toward you to disconnect it from the holding magnets 2 Slide the cover down to remove it 8 3 9 5 RS Probe Cover Installation Procedure 1 Slide the RS Probe Cover upward and align it in order to insert the pins into the riser assembly 2 Gently push th...

Page 239: ...on 8 3 10 4 Riser Thermal Cover Removal Procedure 1 Remove the following covers Lower Front eTower Right Side and Left Side eTower covers 2 Open the door 3 Hold the cover and pull it down 4 Remove the cover 8 3 10 5 Riser Thermal Cover Installation Procedure Attach the Riser Thermal Cover to scanner and secure with two Phillips screws FRU Part Refer to Table 9 9 on page 9 8 Lower Front eTower Cove...

Page 240: ... all accessories 2 Remove the following covers Lower Front eTower RS Probe Cover Right Side and Left Side eTower covers Upper Front eTower and Riser Thermal Cover 3 Remove cockpit monitor 4 Remove Scanner Arm 5 Loosen and remove four screws securing the riser cover to the riser access screws from underneath FRU Part Refer to Table 9 9 on page 9 8 Accessories Replacement Procedures Lower Front eTow...

Page 241: ...ocedures 8 21 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Remove the riser cover Note Ensure the pressure pin is not dropped if exists Figure 8 11 Removing Riser Cover Figure 8 12 Pressure Pin ...

Page 242: ...the riser cover 2 Install four rubber cable guides two short guides on the back side and two long guides in the front side on the riser cover Use a flat screwdriver to gently affix the rubber guide in place if required 3 Install new riser cover on riser and secure the four securing screws Figure 8 13 Installing Riser Cover Rubber Cap Short rubber guide Figure 8 14 Installing Rubber Cable Guides ...

Page 243: ...ermal Cover 7 Install all accessories Arm Installation Procedure Cockpit Monitor Installation Procedure Lower Front eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Upper eTower Front Cover Removal Procedure Riser Thermal Cover Installation Procedure RS Probe Cover Installation Procedure Accessories Replacement Procedu...

Page 244: ... level surface with the wheels in the locked position 8 4 1 4 Base Module Removal Procedure 1 Remove all accessories 2 Remove the cockpit monitor 3 Remove the following covers Lower Front eTower RS Probe cover Right Side and Left Side eTower covers and Front Base Cover 4 Position the system on its side to gain access to the system base 5 Unscrew the four base mounting screws FRU Part Refer to Tabl...

Page 245: ...TION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Remove the base module 7 Remove the four caster wheels Figure 8 15 Removing the Base Module Casters Removal Procedure Four screws that hold Base Module Base Module Removed ...

Page 246: ...e mounting screws and tighten them 4 Refit the following covers Right Side and Left Side eTower covers Lower Front eTowe and Front Base Cover 5 Install the cockpit monitor 6 Install all accessories Figure 8 16 Installing the Base Module Front Base Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Insta...

Page 247: ... Front End Metal Door Removal 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower and Mid Thermal Baffle Cover 3 Remove full front end 4 Remove the riser thermal cover 5 Leave the door open by approximately 30 degrees to make enough space for the FE door to be removed FRU Part Refer to Table 9 7 on page 9 6 Accessories Replace...

Page 248: ...4 System Modules Replacement Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Slightly lift and slide the door upwards until completely removed Figure 8 17 Removing the FE Door ...

Page 249: ...29 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 7 Remove the eight holding screws to separate the front end metal door from the scanner door cover Figure 8 18 Separating the scanner door cover from FE Door ...

Page 250: ...an cable 6 Close the FE door and close the two fastening latches on the left side of the Front End door assembly 7 Tighten the captive screw 8 Refit the following covers Mid Thermal Baffle Cover Lower Front eTower left side eTower right side eTower and RS Probe Cover 9 Install all accessories 10 Turn ON power to the system Riser Thermal Cover Installation Procedure Mid Thermal Baffle Cover Install...

Page 251: ... a level surface with the wheels in the locked position 8 4 3 4 Full Front End Removal 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower and Right Side eTower 3 Open the two fastening latches on the left side of the FE door assembly and loosen the captive screw Then carefully open the door FRU Part Refer to Table 9 12 on page 9 12 Accessories...

Page 252: ...ENUE SERVICE MANUAL P R E L I M I N A R Y 4 Disconnect the following cables MPB to T FEPS BIB to T CFE Blue cable Fan cable 5 While supporting the FE module with one hand unscrew the four Phillips screws and remove the full FE module Figure 8 20 Removing the Full FE Module from the FE Door ...

Page 253: ... with cable clips Fan cable 5 Close the FE door and close the two fastening latches on the left side of the Front End door assembly Figure 8 19 6 Tighten the captive screw 7 Refit the following covers Lower Front eTower left side eTower right side eTower and RS Probe Cover 8 Install all accessories 9 Turn ON power to the system 10 Perform Software Update Procedure to make sure the FW is updated Le...

Page 254: ... 4 SSD Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower 3 Open the two fastening latches on the left side of the Front End door assembly and loosen the captive screw Then open the Front End door 4 Fix the FE door with door support 5 Release the SSD Module captive screw and extract the SSD Module FRU...

Page 255: ...the Front End door assembly 3 Tighten the captive screw 4 Refit the following covers Lower Front eTower left side eTower right side eTower and RS Probe Cover Ins 5 Install all accessories 6 Turn ON power to the system Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Removal Procedure Accessor...

Page 256: ...ltrasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 4 5 4 Cockpit Monitor Removal Procedure 1 On the back of the cockpit release four screws and remove the transparent Wi Fi dongle cover Disconnect the Wi Fi dongle 2 Remove four Phillips screws securi...

Page 257: ...pit rear cables cover if exists see below 4 Disconnect the following cables connected to cockpit monitor BE TO COCKPIT CABLE MPB TO COCKPIT CABLE use flat head screwdriver Figure 8 23 Cockpit Monitor Securing Screws and Cockpit Rear Cover Removal Figure 8 24 Cockpit Monitor Cables System without Cable Cover System with Cable Cover ...

Page 258: ...the dedicated hole and press to release the securing spring The click is sounded indicating the securing spring is released 7 Lift and remove the cockpit monitor Figure 8 25 Cockpit Ground Cables Removal CAUTION Raise the Articulated Arm to its up position to prevent the unintentional injury to personnel from the stored mechanical energy Figure 8 26 Cockpit Monitor Securing Screws Insert screwdriv...

Page 259: ...re 1 Mount the cockpit monitor on the mounting hooks and secure with four Phillips screws 2 Reconnect the cockpit monitor cables BE TO COCKPIT CABLE MPB TO COCKPIT CABLE Ground cables x2 Rear cable cover if exists 3 Reinstall the Gel Holder Figure 8 27 Installing Cockpit Monitor Perform the checks listed in Cockpit Monitor Replacement Procedure on page 8 217 ...

Page 260: ...tanding securely on a level surface with the wheels in the locked position 8 4 6 4 PSU Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower cover Right Side eTower Cover and the Left Side eTower Cover 3 Remove the MPB door 4 Remove the two battery packs FRU Part Refer to Table 9 10 on page 9 10 Accessories Replacement Procedures RS Probe...

Page 261: ...ive screw at the bottom of the PSU and pull the PSU out gently 8 4 6 5 PSU Module Installation Procedure 1 Slide the new PSU into position 2 Secure the captive screw at the bottom of the PSU 3 Install the two battery packs Figure 8 28 Removing the PSU Module Figure 8 29 Installing the PSU Module Battery Module Installation Procedure ...

Page 262: ... Side eTower Cover Left Side eTower Cover and RS Probe Cover 6 Install all accessories MPB Door Cover Installation Procedure Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in PSU Module Replacement Procedure...

Page 263: ...ls in the locked position 8 4 7 4 Battery Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower cover Right Side eTower Cover Left Side eTower Cover and MPB Door Cover 3 Release the two captive screws securing each battery pack 4 Pull out the top and bottom battery packs FRU Part Refer to Table 9 10 on page 9 10 Accessories Replacement Pr...

Page 264: ...screws for each battery 3 Refit the following covers Lower Front eTower cover Right Side eTower Cover Left Side eTower Cover MPB Door Cover and RS Probe Cover 4 Install all accessories MPB Door Cover Installation Procedure Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedur...

Page 265: ...is standing securely on a level surface with the wheels in the locked position 8 4 8 4 BE Module Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower cover Left Side eTower cover and Right Side eTower cover 3 Remove the SSD module 4 Remove MPB Front Metal Door FRU Part Refer to Table 9 13 on page 9 13 Accessories Replacement Procedures RS Probe Cover Removal ...

Page 266: ...bly and loosen the captive screw Then carefully open the door 6 Secure the door with door support 7 Disconnect the following cables connected to the BE module MPB to BIB MPB to BEP white USB cable Cockpit to BEP split cable green and black Printer cable yellow ECG USB cable 8 Disconnect the MPB to BE power cable Figure 8 31 Opening Scanner Door Figure 8 32 Disconnecting Cables Connected to BE Modu...

Page 267: ...E L I M I N A R Y 9 Disconnect the BIB blue cable from the CFE plastic cable guide Disconnect the MPB to BIB cable 10 Loosen the two captive screws on both sides of the BE module 11 Gently slide out the BE module Figure 8 33 Disconnecting MPB to BE Power Cable Figure 8 34 Removing Back End Module ...

Page 268: ...BEP white Cockpit to BEP split cable green and black Printer cable yellow Use the following diagram for reference 5 Reconnect the BIB blue cable to the CFE plastic cable guide 6 Reconnect the MPB to BE power cable 7 Install the SSD module 8 Close the FE door and close the two fastening latches on the left side of the Front End door assembly 9 Tighten the captive screw 10 Install the MPB Front Meta...

Page 269: ...sories 13 Turn ON power to the system Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in Back End BE Module Replacement Procedure on page 8 217 ...

Page 270: ... wheels in the locked position 8 4 9 4 BEP PCB Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Perform BE module removal procedure 4 Remove omega bracket release 5 screws connecting the omega bracket 5 Remove the supporting bracket using an appropriate Phillips screwdriver Accessories Replacement Procedures RS Pro...

Page 271: ...m of the bracket 6 Release the following cable connectors SATA DATA BE TO SSD A red cable press the latch and pull the cable connector out SSD PWR cable Gently pull the connector towards you 7 Disconnect the following cables at the back of the BE Two USB connectors Network cable Figure 8 37 Removing BE Supporting Bracket Figure 8 38 Disconnect SATA DATA and SSD PWR Cables ...

Page 272: ...P PCB Release four screws in the BEP PCB corners using Phillips screwdriver Note If the MiTx to BIB cable is preventing easy removal of the BEP PCB disconnect the MiTx to BIB cable connector latch see yellow circle in the above figure Gently take out the BEP PCB 9 The BEP PCB is released Figure 8 39 Disconnect BE Cables from BE Back Figure 8 40 Remove BEP PCB ...

Page 273: ...ocedure 1 Reconnect the BEP PCB using four screws in the PCB corners using Phillips screwdriver 2 Reconnect the following cables at the back of the BE Two USB connectors Network cable Display port 3 Reconnect the following cable connectors SATA DATA BE TO SSD A red cable Figure 8 41 Reconnect BEP PCB Figure 8 42 Reconnect BE Cables from BE Back ...

Page 274: ...VENUE SERVICE MANUAL P R E L I M I N A R Y SSD PWR cable 4 Reconnect the supporting bracket using an appropriate Phillips screwdriver Two captive screws Two screws at the bottom of the bracket Figure 8 43 Reconnect SATA DATA and SSD PWR Cables Figure 8 44 Reconnecting BE Supporting Bracket ...

Page 275: ...r eTower Front Cover 9 Install all accessories 10 Turn ON power to the system Figure 8 45 BE Omega Bracket Repositioning MPB Module Installation Procedure MPB Front Metal Door Installation Procedure Upper eTower Front Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cov...

Page 276: ...heels in the locked position 8 4 10 4 MPB Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower cover Right Side eTower cover Left Side eTower cover and MPB Door Cover 3 Remove the Battery Packs 4 Remove the PSU module 5 Release the captive screw at the top center of the MPB FRU Part Refer to Table 9 10 on page 9 10 Accessories Replacemen...

Page 277: ...SION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Release the captive screw at the bottom center of the MPB 7 Pull the MPB halfway out 8 Tilt the top toward you and pull the MPB out all the way Figure 8 46 MPB Module Mounting Screws Figure 8 47 MPB Module Removal ...

Page 278: ... the rails until the docking of the MPB module 2 Tighten the two mounting screws to secure the MPB module 3 Return the PSU module 4 Return the battery packs 5 Refit the following covers MPB Door Cover Lower Front eTower cover Right Side and Left Side eTower covers and RS Probe cover 6 Install all accessories 7 Turn ON power to the system PSU Module Installation Procedure Battery Module Installatio...

Page 279: ...E SERVICE MANUAL P R E L I M I N A R Y 8 Check PM_FW 9 If PM_FW is not up to date the Comment section states expected version is 3 xx perform software reload 10 Figure 8 48 Check if PM_FW is up to date Perform the checks listed in MPB Module Replacement Procedure on page 8 217 ...

Page 280: ...cessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower and Right Side eTower 3 Open and remove the full Front End door assembly and place it on a stable surface 4 Remove the T FEPS module CAUTION When performing these procedures take precautions to avoid damage of electrostatic sensitive components Always have the ESD wrist strap connected either to the DIB chas...

Page 281: ...N 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 5 Loosen and remove the two retaining screws and remove the TRx box 6 Remove the T TRx 32 modules 7 Loosen and remove the six screws Figure 8 49 Removing TRx Box T TRx Module Removal Procedure Figure 8 50 Removing TRx Box ...

Page 282: ...the TTRX BOX and secure with two screws previously removed 4 Install the T FEPS module 5 Install the full Front End assembly 6 Refit the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe Cover Figure 8 51 Separating the T CFE Board TRx Module Installation Procedure Front End Power Supply T FEPS Installation Procedure Full Front End Installation Right Side eTower C...

Page 283: ...N 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 7 Install all accessories 8 Turn ON power to the system Accessories Replacement Procedures Perform the checks listed in T CFE cFront End Module Replacement Procedure on page 8 217 ...

Page 284: ...t side eTower and Right Side eTower 3 Remove the T FEPS module 4 Open and remove the full Front End assembly and place it on your laps in order not to damage the probe locker levers 5 Remove the T CFE module The T PSB module is now separated from other parts CAUTION When performing these procedures take precautions to avoid damage of electrostatic sensitive components Always have the ESD wrist str...

Page 285: ...owing covers Lower Front eTower Left side eTower Right Side eTower and RS Probe Cover 6 Install all accessories 7 Turn ON power to the system Plastic Cable Guide for CFE Removal Procedure T CFE Module Installation Procedure Front End Power Supply T FEPS Installation Procedure Full Front End Installation Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lo...

Page 286: ...ure the system is standing securely on a level surface with the wheels in the locked position 8 5 1 4 BIB Board Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower Left side eTower and Right Side eTower 3 Remove the Back End Module 4 Remove the supporting bracket using an appropriate Phillips screwdriver Release two captive screws FRU Part Ref...

Page 287: ...et 5 Disconnect BIB TO BE SPLIT CABLE connector that connects the BIB board to the BE module 6 Disconnect the PCIe cable connector that connects the BIB board to the BE module Figure 8 52 Removing BE Supporting Bracket Figure 8 53 Removing BIB Board Disconnecting BIB TO BE SPLIT CABLE Connector Figure 8 54 Removing BIB Board Disconnecting PCIe Cable Connector ...

Page 288: ...he two fastening latches on the left side of the Front End door assembly 6 Tighten the captive screw 7 Refit the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe cover 8 Install all accessories 9 Turn ON power to the system Figure 8 55 Removing BIB Board BE Module Installation Procedure Right Side eTower Cover Installation Procedure Left Side eTower Cover Install...

Page 289: ...in the locked position 8 5 2 4 CFE Board Release Arm Removal Procedure The lever shaped CFE Release Arm is located on the PSB Board 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower and Right Side eTower 3 Remove the T CFE Module 4 Unscrew the two lever holding screws and remove the CFE Release Arm Lever from the PSB Board Figure 8 56 Locatio...

Page 290: ...5 Install the Front End door assembly 6 Refit the following covers Lower Front eTower Left side eTower and Right Side eTower 7 Install all accessories 8 Turn ON power to the system Figure 8 57 Removing CFE Release Arm Front End Power Supply Removal Procedure Front End Metal Door Replacement Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lower Front eTo...

Page 291: ...ocedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower 3 Open the two fastening latches on the left side of the FE door assembly and carefully open the door 4 Disconnect the MPB TO T FEPS Cable 5 Loosen the 2 retaining screws that secure the T FEPS to the Scanner door assembly CAUTION When performing these procedures take...

Page 292: ...r 3 Reconnect the cables previously disconnected 4 Close the Front End door and fasten securely with the two latches 5 Refit the following covers Lower Front eTower Left side eTower and Right Side eTower 6 Install all accessories 7 Turn ON power to the system Figure 8 58 Removing the T FEPS Metal Cover Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Low...

Page 293: ... covers RS Probe Cover Lower Front eTower Left side eTower and Right Side eTower 3 Open the two fastening latches on the left side of the FE door assembly and loosen the captive screw Then carefully open the door 4 Disconnect the MPB TO T FEPS Cable 5 Disconnect the fan cable CAUTION When performing these procedures take precautions to avoid damage of electrostatic sensitive components Always have...

Page 294: ...previously disconnected 3 Close the Front End door and fasten securely with the two latches 4 Refit the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe Cover 5 Install all accessories 6 Turn ON power to the system Figure 8 59 Removing the T TRX Box Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lower Front eTower Cov...

Page 295: ...and Right Side eTower 3 Open the two fastening latches on the left side of the FE door assembly and loosen the captive screw Then carefully open the door 4 Remove the T TRx Box 5 Secure the door by clipping the door with door support 6 Disconnect the following cables from the FE assembly Fan cable MPB to T FEPS cable CAUTION When performing these procedures take precautions to avoid damage of elec...

Page 296: ...way of a white plastic holder at each end of the T TRx module These are opened by gently pulling them outwards to tilt the holder away from the edge of the module as shown in Figure 8 60 8 Open each white plastic holder located on either end of the first TRx Module to release the module then slide the module towards you to remove it 9 Repeat Steps 4 6 to remove each of the remaining TRx Modules Fi...

Page 297: ...t the cables previously disconnected 6 Close the Front End door fasten securely with the two latches and tighten the captive screw 7 Refit the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe Cover 8 Install all accessories 9 Turn ON power to the system CAUTION When performing these procedures take precautions to avoid damage of electrostatic sensitive components...

Page 298: ...he Venue ultrasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 6 1 4 MPB Front Metal Door Removal Procedure 1 Remove all accessories 2 Remove the Lower Front eTower cover 3 Release four captive screws and remove the MPB Front Metal Door FRU Part Refer ...

Page 299: ...llation Procedure 1 Install the new MPB Front Metal Cover and tighten the four captive screws 2 Refit the Lower Front eTower cover 3 Install all accessories 4 Turn ON power to the system Lower Front eTower Cover Removal Procedure Accessories Replacement Procedures Perform the checks listed in MPB Front Metal Door Replacement Procedure on page 8 217 ...

Page 300: ...ut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 6 2 4 MPB Guide L R Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower Side Covers MPB Power Door 3 Remove the MPB Module 4 Unscrew the two mounting screws that hold the Guide L R FR...

Page 301: ... 8 6 2 5 MPB Guide L R Installation Procedure 1 Attach the MP Guide L R 2 Tighten the two mounting screws to secure the guide to the receptical box 3 Install the MPB Module Figure 8 64 MPB Guide L R Mounting Screws Figure 8 65 MPB Guide L R Mounting Screws MPB Module Installation Procedure MPB Guide L R Mounting Screws MPB Guide L R ...

Page 302: ...ower Cover and RS Probe cover 5 Install all accessories 6 Turn ON power to the system MPB Door Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in MPB Guide L and Guide R Replaceme...

Page 303: ...n 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 6 3 4 Halo Handle Removal Procedure 1 Remove all accessories 2 Remove cockpit monitor 3 Remove the arm 4 Remove the Riser Thermal Cover 5 Remove the Riser Cover FRU Part Refer to Table 9 6 on page 9 5 Accessories Replacement Procedures Cockpit ...

Page 304: ...6 Remove the Halo Handle mounting screws four x4 front screws and four x4 side screws two from each side 7 Lift the handle upwards and remove it Figure 8 66 Removing the Halo Handle Mounting Screws Figure 8 67 Removing the Halo Handle Halo Handle Side Screws Halo Handle Front Screws Lift and Remove the Halo Handle Halo Handle Removed ...

Page 305: ...s and tighten them 3 Install the Riser Cover 4 Install the Riser Thermal Cover 5 Install the articulated arm 6 Install the cockpit monitor 7 Install all accessories 8 Turn ON power to the system Riser Cover Installation Procedure Riser Thermal Cover Installation Procedure Arm Installation Procedure Cockpit Monitor Installation Procedure Accessories Replacement Procedures Perform the checks listed ...

Page 306: ... securely on a level surface with the wheels in the locked position 8 6 4 4 Arm Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower cover Left Side eTower cover Right Side eTower cover and Riser Thermal Cover 3 Remove cockpit monitor 4 Carefully raise the arm to its up position FRU Part Refer to Table 9 6 on page 9 5 Accessories Replacement Pr...

Page 307: ...d through the arm axis refer to MPB TO Cockpit Cable Removal Procedure and to BE to Cockpit Cable Removal Procedure 7 Using Allen key screwdriver remove the two securing screws of the arm side cover 8 Slide the arm side cover upwards and remove it 9 Perform steps 4 5 for the other arm side cover Figure 8 68 Removing Arm Base Cover Figure 8 69 Removing Arm Side Covers ...

Page 308: ...ll out the cockpit monitor cables routed inside the riser 11 Disconnect the ground cable from the riser 12 Using Forged Steel Ring Pliers remove the C clip and then the Teflon ring Figure 8 70 Removing Cockpit Monitor Cables Figure 8 71 Removing Ground Cable Figure 8 72 Removing Teflon Ring and C clip Teflon ring C clip C clip Teflon ring ...

Page 309: ... cables through the riser opening refer to MPB TO Cockpit Cable Installation Procedure and BE to Cockpit Cable Installation Procedure 4 Route each cockpit monitor cable through the arm side and install the arm side covers 5 Re connect the ground cable to the riser 6 Secure each side cover with two securing screws 7 Install the arm base cover 8 Install the following covers Lower Front eTower cover ...

Page 310: ...s 11 Turn ON power to the system Riser Thermal Cover Removal Procedure Left Side eTower Cover Removal Procedure Right Side eTower Cover Removal Procedure Lower Front eTower Cover Removal Procedure RS Probe Cover Installation Procedure Cockpit Monitor Installation Procedure Accessories Replacement Procedures Perform the checks listed in Articulated Arm Replacement Procedure on page 8 218 ...

Page 311: ...ser Removal Procedure 1 Remove all accessories 2 Remove cockpit monitor 3 Remove the following covers RS Probe cover Lower Front eTower cover Left Side eTower cover Right Side eTower cover Riser Thermal Cover and Upper Front eTower Cover 4 Remove scanner arm 5 Loosen and remove four screws securing the riser cover to the riser FRU Part Refer to Table 9 5 on page 9 4 Accessories Replacement Procedu...

Page 312: ...100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Remove the riser cover 7 Remove the Teflon sliding top ring Note Ensure the pressure pin is not dropped if exists Figure 8 75 Removing Riser Cover Figure 8 76 Removing Teflon Sliding Top Ring Figure 8 77 Pressure Pin ...

Page 313: ...er 8 6 5 5 Riser Installation Procedure 1 Install new riser on the eTower 2 Apply LOCTITE 243 on the riser securing screws and tighten the screws 3 Install the Teflon sliding top ring 4 Remove the masking tape holding the pressure pin ensure the pin is not dropped Figure 8 78 Removing Riser Figure 8 79 Installing Teflon Sliding Top Ring Figure 8 80 Pressure Pin ...

Page 314: ...over and RS Probe cover 8 Install the cockpit monitor 9 Install all accessories 10 Turn ON power to the system Arm Installation Procedure Upper eTower Front Cover Installation Procedure Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Riser Thermal Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Pr...

Page 315: ... 6 4 Casters Removal Procedure 1 Using a 6mm Allen key loosen and remove the screw that secures the wheel shaft in the wheel securing socket 2 Lift the chassis sufficiently to allow the wheel to drop down out of the wheel securing socket 3 Remove the wheel 8 6 6 5 Casters Installation Procedure 1 Carefully lift the chassis sufficiently to allow insertion of the replacement wheel shaft into the whe...

Page 316: ...ure the screw is tightened sufficiently to secure the wheel on the chassis Do not over tighten the screw NOTE When installing wheels make sure you install the correct type of the wheel The wheels position is crossover 5 Carefully position the system on the ground Figure 8 82 Wheels Position in the System Perform the checks listed in Casters Replacement Procedure on page 8 218 ...

Page 317: ...l Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower Left side eTower Right Side eTower Riser Thermal Cover 3 Open the two fastening latches on the left side of the FE door assembly loosen the captive screw and carefully open the door See Figure 8 19 4 Secure the door by clipping the door with door support Figure 8 19 5 Remove the BIB cable from the ...

Page 318: ...SERVICE MANUAL P R E L I M I N A R Y Gently pull down the Plastic Cable Guide for CFE and disconnect the cable connector 6 Loosen and remove the two screws securing the cable guide to CFE board and remove it Figure 8 83 Disconnecting the PCIe Cable Figure 8 84 Removing Plastic Cable Guide for CFE ...

Page 319: ... 4 Close the FE door and close the two fastening latches on the left side of the Front End door assembly 5 Tighten the captive screw 6 Refit the following covers Lower Front eTower left side eTower Right side eTower and RS Probe cover 7 Install all accessories 8 Turn ON power to the system Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTow...

Page 320: ...evel surface with the wheels in the locked position 8 6 8 4 MPB Blower Fan Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remove MPB Front Metal Door 4 Disconnect MPB to BIB Control Cable from the fan cable connector FRU Part Refer to Table 9 7 on page 9 6 Accessories Replacement Procedures RS Probe Cover Removal...

Page 321: ...edures 8 101 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 5 Loosen the elastic spring slings that hold the fan and carefully remove the MPB fan Figure 8 86 Removing the MPB Fan Elastic spring sling ...

Page 322: ...on Procedure 1 Carefully insert the new MPB fan into the fan sleeve 2 Tighten the elastic spring slings to secure the fan in its place 3 Reconnect the MPB to BIB Control Cable to the fan cable connector 4 Install the MPB Front Metal Door Figure 8 87 Tightening Elastic Spring Slings Figure 8 88 Installing the MPB Fan MPB Front Metal Door Installation Procedure ...

Page 323: ...Side eTower and RS Probe cover 6 Install all accessories 7 Turn ON power to the system Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in MPB Blower Fan Replacement Procedure on page 8 218 ...

Page 324: ...he wheels in the locked position 8 6 9 4 IPP Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower Left side eTower Right Side eTower and MPB Front Metal Door 3 Remove the MPB Power Door 4 Remove battery modules 5 Remove the PSU module 6 Remove the MPB module FRU Part Refer to Table 9 10 on page 9 10 Accessories Replacement Procedures RS ...

Page 325: ... the three screws that hold the IPP metal frame 8 Using the nose pliers or flat head screwdriver squeeze the white plastic connector holders and push them out to remove the power connector Figure 8 89 Removing IPP Holding Screws Figure 8 90 Removing Power Connector Connector holder Connector holder System back System front ...

Page 326: ...ts Replacement Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 9 Using the open end wrench disconnect the ground cable 10 Remove the IPP module Figure 8 91 Disconnecting Ground Cable Ground cable ...

Page 327: ...Power Door 8 Refit the following covers MPB Front Metal Door Left side eTower Right Side eTower Lower Front eTower and RS Probe cover 9 Install all accessories 10 Turn ON power to the system MPB Module Installation Procedure PSU Module Installation Procedure Battery Module Installation Procedure MPB Door Cover Installation Procedure MPB Front Metal Door Installation Procedure Right Side eTower Cov...

Page 328: ...in the locked position 8 7 1 4 MPB Rear USB Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remove MPB Front Metal Door 4 Remove the battery packs to gain access to the MPB module 5 Remove the PSU module 6 Unscrew the two mounting screws and remove the MPB module FRU Part Refer to Table 9 14 on page 9 14 Acc...

Page 329: ...2 Connector 8 Using Phillips screwdriver disconnect the two screws that hold the cable connector to the rear panel and remove the cable Figure 8 92 Disconnecting MPB Rear USB Cable To MPB J12 Connector Figure 8 93 Disconnecting MPB Rear USB Cable To Rear Panel Connector MPB to Rear USB Cable To MPB J12 Connector MPB to Rear USB Cable To Rear Panel Connector side 1 MPB to Rear USB Cable To Rear Pan...

Page 330: ...battery pack units 6 Install MPB Front Metal Door 7 Refit the following covers Lower Front eTower left side eTower Right side eTower and RS Probe cover 8 Install all accessories 9 Turn ON power to the system MPB Module Installation Procedure PSU Module Installation Procedure Battery Module Installation Procedure MPB Front Metal Door Installation Procedure Left Side eTower Cover Installation Proced...

Page 331: ...eels in the locked position 8 7 2 4 MPB TO Cockpit Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower Right Side eTower and Riser Thermal covers 3 Perform MPB removal procedure 4 For systems with rear cockpit cable cover installed perform the following steps a Disconnect the Wi Fi adapter Accessories Replacement Procedures RS Probe Co...

Page 332: ... Y b Release the four Phillips screws and remove the rear cable cover 5 Disconnect the MPB TO COCKPIT CABLE from the monitor Disconnect the cable connector use flat head screwdriver Release Ground cable use Phillips screwdriver Figure 8 94 Removing Rear Cable Cover Figure 8 95 MPB TO COCKPIT Cable and Cockpit Ground Cables Removal ...

Page 333: ...ly raise the arm to its up position b Remove the arm base cover c Using Phillips screwdriver remove the two securing screws of the arm side cover d Slide the right side arm side cover upwards and remove it 8 Cut the tie wraps securing the MPB TO COCKPIT cable to the arm right side and release the cable from the arm Figure 8 96 Removing Arm Base Cover Figure 8 97 Removing Arm Side Covers Figure 8 9...

Page 334: ...M I N A R Y 9 Route the cable and its connector downwards through the arm hole 10 Release plastic clips that hold the cable using a flat screwdriver Three clips in the Front End Two clips between the connectors and the MPB receptacle side Figure 8 99 Route Cable Through Arm Hole Figure 8 100 Plastic Clips Holding the Cable ...

Page 335: ... not to pull and cause damage to the other cables 13 Open cable cover door Open the captive screw of the cable cover door using a Phillips screwdriver Slide the cable cover door towards you to free the cable Slide the cable connectors upwards and gently route them one after the other through the door with the rubber grommet 14 Pull the MPB TO COCKPIT CABLE The cable is released Figure 8 101 Discon...

Page 336: ...le Installation Procedure 1 Remove ferrite clip from MPB to Cockpit cable 2 Route the MPB to Cockpit cable through the articulated arm so that the length of the exposed cable toward the cockpit is minimum 44 cm Do not use tie wraps to secure cable in place Figure 8 103 Removing Ferrite from Cable Figure 8 104 Routing MPB to Cockpit Cable Measuring 44 cm 44 cm ...

Page 337: ...force and route the MPB to Cockpit cable through the tilt axis keeping the routed cable on the most left and the connector facing to the right Make sure it is routed above the tilt release cable wire Figure 8 105 Removing Upper and Lower Cockpit Covers Figure 8 106 Routing the MPB to Cockpit Cable through Tilt Axis Upper cover removed Lower cover removed Connector facing to the right ...

Page 338: ...cable approximately 3 cm below the cable connector 6 Connect the MPB to Cockpit cable to the cockpit and fasten the flat screws 7 Reconnect the ground cable 8 Make sure the MPB to Cockpit cable is routed under the pin that is located on the rear side of the Figure 8 107 Attaching Ferrite to MPB to Cockpit Cable Figure 8 108 Connecting the MPB to Cockpit Cable to Cockpit 3 cm ...

Page 339: ...e lower covers 10 Route the MPB to Cockpit cable through the cable cover door and the arm hole 11 Connect the MPB TO COCKPIT CABLE to the monitor Connect the cable connector use flat head screwdriver Connect Ground cable use Phillips screwdriver Figure 8 109 Routing the MPB to Cockpit Cable under the Pin Figure 8 110 Routing the MPB TO COCKPIT CABLE Route cable under the pin ...

Page 340: ...cure cable to arm using tie wraps 14 Gently slide the MPB TO COCKPIT cable through the MPB receptacle side towards the front 15 Reconnect J1 and J2 connectors ensuring the connectors arrow is pointing upwards 16 Secure the cable to the plastic clips Three in the front end Figure 8 111 Measuring 30 cm of the Cable Figure 8 112 Securing the Cable to Arm Figure 8 113 Reconnecting J1 and J2 Connectors...

Page 341: ...se the cable cover door Slide the cable cover door towards the system and fix the rubber grommet in its position Close the captive screw of the cable cover door using a Phillips screwdriver 18 Install the arm side cover 19 Secure side cover with two securing screws 20 Install the arm base cover Figure 8 114 Plastic Clips Holding the Cable Figure 8 115 Closing Cable Cover Door ...

Page 342: ...t the Wi Fi dongle if exists 25 nstall all accessories 26 Turn ON power to the system MPB Module Installation Procedure Riser Thermal Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Wi Fi Adapter Installation Procedure For Systems with No E...

Page 343: ...E to Cockpit Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Lower Front eTower Left side eTower Right Side eTower Riser Thermal and MPB Door cover 3 Disconnect the BE TO COCKPIT CABLE from the monitor Disconnect the cable connector use flat head screwdriver Release Ground cable use Phillips screwdriver FRU Part Refer to Table 9 14 on page 9 14 Accessories R...

Page 344: ...following steps a Disconnect the Wi Fi adapter b Release the four Phillips screws and remove the rear cable cover c Gently pull out the BE to Cockpit cable from the axis tilt 5 Release the BE to Cockpit cable from the arm a Carefully raise the arm to its up position b Remove the arm base cover Wi Fi Adapter Removal Procedure Figure 8 117 Removing Rear Cable Cover Figure 8 118 Removing Arm Base Cov...

Page 345: ...m side cover d Slide the left side arm side cover upwards and remove it 6 Cut the tie wraps securing the BE to cockpit cable to the arm left side and release the cable from the arm 7 Route the cable and its connector downwards through the arm hole Figure 8 119 Removing Arm Side Covers Figure 8 120 Arm Side Cable Removing Tie Wraps Figure 8 121 Route Cable Through Arm Hole ...

Page 346: ... side 9 Disconnect the Cockpit to BEP split cable connected to BE module green USB connector and black display port connector 10 Open cable cover door a Open the captive screw of the cable cover door using a Phillips screwdriver b Slide the cable cover door towards you to free the cable c Slide the cable connectors upwards and gently route them one after the other through the Figure 8 122 Plastic ...

Page 347: ...er 8 Replacement Procedures 8 127 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y door with the rubber grommet The cable is released Figure 8 124 Releasing the BE to Cockpit Cable ...

Page 348: ...le Installation Procedure 1 Route BE to Cockpit Cable through arm so that the cable length that remains exposed to the cockpit direction is minimum 35cm 2 Remove the lower and then the upper cockpit covers Figure 8 125 Routing BE to Cockpit Cable Measuring 35 cm Figure 8 126 Removing Upper and Lower Cockpit Covers 35 cm Upper cover removed Lower cover removed ...

Page 349: ... trough cockpit tilt axis For easier routing make sure the cable is aligned to the left 4 Connect the BE to Cockpit cable DP connector to the cockpit 5 Reconnect the ground cable 6 Refit the cockpit upper and then the lower covers Figure 8 127 Routing BE to Cockpit Cable through Tilt Axis Figure 8 128 Connecting BE to Cockpit Cable to the Cockpit ...

Page 350: ... use flat head screwdriver Connect Ground cable use Phillips screwdriver 9 Gently slide the BE TO COCKPIT cable through the MPB receptacle side towards the front 10 Reconnect the Cockpit to BEP split cable connected to BE module green and black Ensure the cables are connected to the correct locations see cable diagram below 11 Secure the cable to the plastic clips Figure 8 129 Route Cable Through ...

Page 351: ...cable cover door towards the system and fix the rubber grommet in its position Close the captive screw of the cable cover door using a Phillips screwdriver 13 Install the arm side cover 14 Secure side cover with two securing screws 15 Install the arm base cover 16 Refit the following covers Left Side eTower Riser Thermal Right Side eTower Lower Front Figure 8 132 Plastic Clips Holding the Cable Fi...

Page 352: ...l accessories 20 Turn ON power to the system Riser Thermal Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Wi Fi Adapter Installation Procedure For Systems with No Existing Wi Fi Adapter Accessories Replacement Procedures Perform the checks...

Page 353: ...bed in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 7 4 4 PSU to MPB DC Docking Cable Removal Procedure 1 Remove all accessories 2 Remove the MPB module 3 Remove the fan 4 Disconnect J8 to lower PSU connectors see connectors diagram below FRU Part Refer to Table 9 14 on page...

Page 354: ... a On each side of each connector use a flat screwdriver to push one of the plastic flaps towards the hole and then push the other flap through the hole to release the connector b From the front side release both cable edges by pulling the rubber towards you Figure 8 135 Releasing the Connector Figure 8 136 Releasing the Cable Edges ...

Page 355: ...ensuring the orientation is appropriate by checking the plastic marks are pointing to the correct direction Push the connectors plastic hinges through the designated holes until both hinges are fully inside Connect the ground cable using an appropriate Phillips screwdriver 2 Perform Fan installation procedure 3 perform MPB installation procedure 4 Turn ON power to the system Figure 8 137 Releasing...

Page 356: ...h is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 7 5 4 MPB to BE PWR Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower Side Covers MPB Power Door 3 Remove the MPB Module FRU Part Refer to Table 9 14 on page 9 14 Accessories Replacement Procedures RS Probe Cover Remo...

Page 357: ...L I M I N A R Y 4 Disconnect the two lower cable connectors J4 J5 by sliding them upwards see J connectors map that follows 5 Release the cable from the two plastic cable clips The cable is released Figure 8 139 Disconnecting Two Lower Connectors Figure 8 140 Releasing Cable from Plastic Cable Clips ...

Page 358: ...lowing covers MPB Door Cover Left and Right Side eTower Covers Lower Front eTower Cover and RS Probe cover 6 Install all accessories 7 Turn ON power to the system MPB Module Installation Procedure MPB Front Metal Door Installation Procedure MPB Door Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Ins...

Page 359: ...heels in the locked position 8 7 6 4 ON OFF Switch Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remove the battery packs to gain access to the MPB module 4 Unscrew the two mounting screws and remove the MPB module FRU Part Refer to Table 9 14 on page 9 14 Accessories Replacement Procedures RS Probe Cover ...

Page 360: ... Disconnect the ON OFF Switch cable connector from the MPB J10 connector Disconnect the ON OFF Switch cable connector with the ON OFF switch from the MPB module by pressing the securing clips on both sides of the connector 6 Remove the ON OFF Switch Cable Figure 8 142 ON OFF Switch Cable To MPB J10 Connector Figure 8 143 ON OFF Switch Cable with ON OFF Switch Connection to MPB Module To MPB J10 Co...

Page 361: ...he MPB module and tighten the two mounting screws 4 Install the battery pack units 5 Refit the following covers Lower Front eTower left side eTower Right side eTower and RS Probe cover 6 Install all accessories 7 Turn ON power to the system Battery Module Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Ins...

Page 362: ...rface with the wheels in the locked position 8 7 7 4 PCIe Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remove the BE module 4 Disconnect one side of the PCIe cable from the BIB board and the other side from the BE module 5 Remove the PCIe cable FRU Part Refer to Table 9 14 on page 9 14 Accessories Replace...

Page 363: ... BE module 3 Refit the following covers Lower Front eTower Left side eTower and Right Side eTower 4 Install all accessories 5 Turn ON power to the system BE Module Installation Procedure Right Side eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure Accessories Replacement Procedures Perform the checks listed in PCIe Cab...

Page 364: ...securely on a level surface with the wheels in the locked position 8 7 8 4 MPB to BIB Control Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remove MPB Front Metal Door 4 Remove the MPB module 5 Disconnect the cable from the BIB board FRU Part Refer to Table 9 14 on page 9 14 Accessories Replacement Procedu...

Page 365: ...NUAL P R E L I M I N A R Y 6 Disconnect MPB to BIB Control Cable from the fan cable connector 7 Release plastic clips that hold the cable Figure 8 146 Disconnecting Fan Cable Connector Figure 8 147 Releasing MPB to BIB Control Cable from Plastic Clips MPB to BIB Control Cable Fan Connector ...

Page 366: ...the blower cable connector 3 Secure the cable with plastic clips 4 Install the MPB module 5 Install the MPB Front Metal Door 6 Install the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe cover 7 Install all accessories 8 Turn ON power to the system Figure 8 148 Disconnecting MPB to BIB Control Cable from J6 Connector MPB Module Installation Procedure MPB Front M...

Page 367: ...ith the wheels in the locked position 8 7 9 4 Printer USB Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower Upper eTower Front Cover 3 Remove the MPB Front Metal Door FRU Part Refer to Table 9 19 on page 9 18 Accessories Replacement Procedures RS Probe Cover Removal Procedure Upper eTower Front Cover ...

Page 368: ...ive screw NOTE Make sure the Peripheral Cable Cover is properly aligned and affixed 4 Install MPB Front Metal Door 5 Refit the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower Upper eTower Front Cover 6 Install all accessories 7 Turn ON power to the system Figure 8 150 Disconnecting the Printer Cable MPB Front Metal Door Installation Procedure Upper eTower Fron...

Page 369: ...is standing securely on a level surface with the wheels in the locked position 8 7 10 4 BEP Cables Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Perform BE module removal procedure 4 Remove omega bracket release 5 screws connecting the omega bracket Accessories Replacement Procedures RS Probe Cover Removal Proce...

Page 370: ...captive screws Release two screws at the bottom of the bracket 6 Release the following cable connectors SATA DATA BE TO SSD A red cable press the latch and pull the cable connector out SSD PWR cable Gently pull the connector towards you 7 Disconnect the following cables at the back of the BE Two USB connectors Network cable Figure 8 152 Removing BE Supporting Bracket Figure 8 153 Disconnect SATA D...

Page 371: ...able is preventing easy removal of the PCB disconnect the MiTx to BIB cable connector latch see yellow circle in the above figure Gently take out the PCB 9 The cables are now exposed and easy to access Disconnect the relevant cable that is being replaced USB split cable use socket wrench 5 5mm HDMI cable use socket wrench 5 5mm Network cable use socket wrench 5 5mm Figure 8 154 Disconnect BE Cable...

Page 372: ...PWR cable SATA DATA BE to SSD cable MiTx to BIB cable release one connector on the BIB and pull upwards two connectors on the BEP PCB 8 7 10 5 BEP Cables Installation Procedure 1 Reconnect the cable s that was were removed USB split cable HDMI cable Network cable SATA PWR cable Figure 8 156 Remove PCB Figure 8 157 MiTx to BiB Cable Removal ...

Page 373: ... R Y 2 Route the cables on the BEP bottom bracket as shown below 3 Reconnect the PCB using four screws in the PCB corners using Phillips screwdriver 4 Reconnect the following cables at the back of the BE Two USB connectors Network cable Figure 8 158 Route cables on BEP Bottom Bracket Figure 8 159 Reconnect PCB ...

Page 374: ...R E L I M I N A R Y Display port 5 Reconnect the following cable connectors SATA DATA BE TO SSD A red cable SSD PWR cable 6 Reconnect the supporting bracket using an appropriate Phillips screwdriver Two captive screws Figure 8 160 Reconnect BE Cables from BE Back Figure 8 161 Reconnect SATA DATA and SSD PWR Cables ...

Page 375: ...wing covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower Upper eTower Front Cover Figure 8 162 Reconnecting BE Supporting Bracket Figure 8 163 BE Omega Bracket Repositioning MPB Module Installation Procedure MPB Front Metal Door Installation Procedure Upper eTower Front Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Instal...

Page 376: ...res DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 11 Install all accessories 12 Turn ON power to the system Accessories Replacement Procedures Perform the checks listed in Back End BE Module Replacement Procedure on page 8 217 ...

Page 377: ...ultrasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 7 11 4 MPB TO T FEPS Cable Removal Procedure 1 Perform MPB removal procedure 2 Disconnect the following cables connected to BE module MPB to BIB MPB to BEP white Cockpit to BEP split cable green and...

Page 378: ...the MPB TO T FEPS cable edge 3 Connect the following cables to BE module MPB to BIB MPB to BEP white Cockpit to BEP split cable green and black Printer cable yellow 4 Install the MPB module 5 Install the MPB Front Metal Door 6 Install the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe cover Figure 8 165 Disconnecting MPB to T FEPS Cable Edge MPB Module Installa...

Page 379: ...5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 7 Install all accessories 8 Turn ON power to the system Accessories Replacement Procedures Perform the checks listed in Front End Power Supply T FEPS Replacement Procedure on page 8 217 ...

Page 380: ...uired 15 min 8 7 12 3 Preparations 1 Shut down the Venue ultrasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 7 12 4 MPB To Peripherals Harness FRU Cable Removal Procedure 1 Perform MPB removal procedure 2 Release the cable from the plastic cable clip...

Page 381: ...he cable from the upper plastic cable clips 4 If a printer is used Release the printer plastic cable clips routed upwards Loosen the Phillips screw and remove the Peripheral Cable Cover Figure 8 167 Disconnecting Cable Clips Figure 8 168 Release Cable from Upper Plastic Cable Clips Figure 8 169 Disconnecting the Printer Cable ...

Page 382: ...VENUE SERVICE MANUAL P R E L I M I N A R Y Disconnect the printer cable yellow USB connector from the Back End assembly 5 Disconnect the MPB to BEP white cable connected to BE module Figure 8 170 Release Cable according to Configuration Figure 8 171 Disconnecting Cables Connected to BE Module ...

Page 383: ...e clips 4 Install the MPB module 5 Install the MPB Front Metal Door 6 Install the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe cover 7 Install all accessories 8 Turn ON power to the system Figure 8 172 Release Cable from J3 Connector MPB Module Installation Procedure MPB Front Metal Door Installation Procedure Left Side eTower Cover Installation Procedure Rig...

Page 384: ...he wheels in the locked position 8 8 1 4 Printer Removal Procedure NOTE Perform these steps only for Venue systems with printer already installed on them For systems with no previously installed printer skip this section and proceed to section Printer Installation Procedure For Systems with No Existing Printer 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower...

Page 385: ...ng Printer 1 Remove the Printer Insert Cover from the Upper eTower Cover 2 Install the Printer Frame Cover and secure with two Phillips screws 3 Remove the metal bracket Use Phillips screwdriver Printer Insert Cover Removal Procedure Figure 8 173 Installing Printer Frame Cover Figure 8 174 Removing Metal Bracket Upper eTower Cover Printer Frame Cover Phillips screws Metal bracket ...

Page 386: ...clip the white plastic cable holder and release the two printer cables Printer Power Cable and Printer USB Cable 6 Attach bracket to the Printer and tighten the four Phillips screws the bracket and the screws are supplied with the printer kit MPB Front Metal Door Removal Procedure Figure 8 175 Releasing Printer Cables Figure 8 176 Attaching Bracket to Printer Unclip cable holder ...

Page 387: ... MANUAL P R E L I M I N A R Y 7 Insert the Printer Power Cable and Printer USB cable through the dedicated opening 8 Connect the Printer USB Cable and Printer Power Cables to the Printer Figure 8 177 Printer Cables Routed Through Dedicated Opening Figure 8 178 Connecting Printer Cables ...

Page 388: ... L I M I N A R Y 9 Attach the Printer to the eTower and secure with two bottom screws 10 Attach the metal cable cover and secure the Phillips screw 11 Install MPB Front Metal Door Figure 8 179 Attaching Printer to the eTower Figure 8 180 Attaching Metal Cable Cover MPB Front Metal Door Installation Procedure Metal cable cover ...

Page 389: ...nter Installation Procedure For Systems with Existing Printer 1 Perform Steps 8 14 of the section Printer Installation Procedure For Systems with No Existing Printer Upper eTower Front Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Installation Procedure Accesso...

Page 390: ...ake sure the system is standing securely on a level surface with the wheels in the locked position 8 8 2 4 Wi Fi Adapter Removal Procedure NOTE Perform these steps only for Venue systems with Wi Fi adapter already installed on them For systems with no previously installed Wi Fi adapter skip this section and proceed to section Wi Fi Adapter Installation Procedure For Systems with No Existing Wi Fi ...

Page 391: ...iously installed Wi Fi adapter For systems with existing Wi Fi adapter skip this step and proceed to Step 2 2 Connect the L shape USB adapter to the back of the Cockpit 3 Attach the transparent Wi Fi dongle cover and secure with four screws 8 8 2 6 Barcode Replacement Procedure 1 Connect the Barcode to the system 2 Wait for a few minutes for the system to accept the Barcode 3 Restart the system Fi...

Page 392: ... Installation Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y Note The barcode device is detected by the system as a COM Port under windows device manager Figure 8 184 Barcode detection in Device manager ...

Page 393: ... is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 8 3 4 ECG Installation 1 Remove the following covers Lower Front eTower Upper Front eTower and MPB Front Metal Door Printer Insert Cover 2 Replace the printer insert cover with the one provided in the ECG installation kit FRU Part Refer to Table 9 20 on page 9 19 Lower Front eTowe...

Page 394: ...ws and 2 washers verify correct orientation 4 Using two Thumb screws install the bracket on the eTower 5 Connect the ECG USB cable to the USB 2 0 port on the BEP above the Green connector 6 Route the ECG USB cable through the dedicated opening in the eTower Figure 8 186 Attaching ECG Bracket to Printer Bracket Figure 8 187 Installing Bracket on the eTower Figure 8 188 Connecting ECG USB Cable ...

Page 395: ...CG USB cable 8 Fold the excessive cable and use the cable clips to hold it 9 Attach the metal bracket to close the cable opening 10 Route the USB cable as shown below and connect the ECG USB cable to the ECG Figure 8 189 Attaching Grommet to the ECG USB Cable Figure 8 190 Attaching Metal Bracket Figure 8 191 Connecting ECG to ECG USB Cable ...

Page 396: ...t the ECG module and tighten the screw to fix the ECG module in place 12 Install the following covers MPB Front Metal Door Upper Front eTower Lower Front eTower with cutout for ECG Figure 8 192 Fixing ECG Module MPB Front Metal Door Installation Procedure Lower Front eTower Cover Installation Procedure Upper eTower Front Cover Installation Procedure ...

Page 397: ...Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 8 4 4 ECG Installation 1 Remove the following covers Lower Front eTower Upper Front eTower and MPB Front Metal Door 2 Disconnect printer USB and Power cables FRU Part Refer to Table 9 20 on page 9 19 Lower Front eTower Cover Remov...

Page 398: ...o thumbscrews holding the printer bracket and remove the printer with the bracket 4 Remove four screws connecting the printer to the bracket 5 Attach the ECG bracket to the printer bracket with 2 screws and 2 washers verify correct orientation Figure 8 194 Disconnecting Printer Figure 8 195 Disconnecting Printer Figure 8 196 Attaching ECG Bracket to Printer ...

Page 399: ...e printer using four screws 7 Connect ECG USB cable to the USB 2 0 port on the BEP on top of the green connector 8 Route the USB cable through the dedicated opening in the eTower and affix it along with the printer USB cable Figure 8 197 Attaching Bracket to Printer Figure 8 198 Connecting ECG USB Cable Figure 8 199 Connecting ECG USB Cable ...

Page 400: ...the metal bracket 10 Connect the USB cable and power cable to the printer and the USB cable to the ECG module 11 Using two Thumb screws install the bracket on the eTower 12 Tighten the screw to fix the ECG module in place Figure 8 200 Installing Metal Bracket Figure 8 201 Connecting Printer and ECG Cables Figure 8 202 Connecting Printer and ECG Cables ...

Page 401: ...cable and use the cable clips to hold it 14 Install the following covers MPB Front Metal Door Upper Front eTower Lower Front eTower with cutout for ECG Figure 8 203 Connecting Printer and ECG Cables MPB Front Metal Door Installation Procedure Lower Front eTower Cover Installation Procedure Upper eTower Front Cover Installation Procedure ...

Page 402: ...s described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 8 5 4 ECG Removal Procedure 1 Remove the following covers Lower Front eTower and Upper Front eTower 2 Release the screw that secures the ECG module in place FRU Part Refer to Table 9 20 on page 9 19 Lower Front eTow...

Page 403: ...and remove the ECG 8 8 5 5 ECG Installation Procedure 1 Connect ECG to the USB cable 2 Insert the ECG module all the way in 3 Tighten the ECG securing screw 4 Install the following covers Upper Front eTower Lower Front eTower Figure 8 205 Pulling Out ECG Figure 8 206 Removing ECG Upper eTower Front Cover Installation Procedure Lower Front eTower Cover Installation Procedure ...

Page 404: ...allow the program to make changes to your computer 3 The ImageUSB by PassMark Software main window appears NOTICE Burning of the disk on key media Mem stick is performed using ImageUSB utility downloaded from the GE portal Table 8 3 ImageUSB System Requirements Requirement Description Operating System Windows XP SP3 Windows Server 2003 Windows Vista Windows Server 2008 Windows 7 Windows 8 and Wind...

Page 405: ...e check box 6 In the Step 3 section click Browse and select the image file in BIN format downloaded from the GE Portal 7 In the Step 4 section click Write to start the burning process Figure 8 208 Disk on Key Burn Process Select Target USB Figure 8 209 Disk on Key Burn Process Write Image to USB Drive Figure 8 210 Disk on Key Burn Process Write Image to USB Drive Figure 8 211 Disk on Key Burn Proc...

Page 406: ...e burning process 9 View the burning progress in the Overall progress status bar 10 If the software burning process is successfully completed the Imaging Completed message appears Click OK to close the notification window Figure 8 212 Disk on Key Burn Process Confirmation and Warning Messages Figure 8 213 Disk on Key Burn Process Progress Status Bar Figure 8 214 Disk on Key Burn Process Imaging Co...

Page 407: ...I M I N A R Y 11 If the burning process fails a notification window appears Click OK to close the window Repeat steps 3 9 using different USB media 12 Exit the ImageUSB utility 13 Proceed with the Software Update Procedure section page 8 188 Figure 8 215 Disk on Key Burn Process Imaging Failed Message ...

Page 408: ...ations in accordance with different software versions 1 Disconnect all probes 2 Plug the Software Installation Media into the USB port located on the Interface Panel of the Venue 3 Turn on the system The opening screen is displayed 4 Select Update Venue SW to install software without affecting user data For R1 software version 301 X X the following Warning message appears Figure 8 216 Venue Softwa...

Page 409: ...N A R Y For R2 software version 302 X X the following Warning message appears 5 Click Yes in the Warning message box 6 Automatic installation process runs At the end of the process the following message appears Figure 8 219 Venue Software Installation Options Figure 8 218 Venue Partition C Warning Message R2 ...

Page 410: ...ion media and tap OK in the message box 9 In the Confirmation dialog box that appears tap Yes to continue the reboot 10 Tap Yes to continue the reboot 11 After the system reboots the Venue setup wizard welcome screen appears 12 Select the interface language and tap OK to proceed Figure 8 220 Venue Installation Process Information Dialog Box Figure 8 221 Venue Installation Process Confirmation Dial...

Page 411: ... Y 13 Tap Run wizard to continue the installation 14 Enter the system serial number and tap OK to confirm 15 Tap OK to confirm the serial number Figure 8 223 Venue Installation Wizard Run Wizard Figure 8 224 Venue Installation Process Set Serial Number Figure 8 225 Venue Installation Process Confirm Serial Number ...

Page 412: ... regain the initial storage capacity of the media used for burning software downloaded from GE portal 1 On your PC insert the Disk on Key media into any available USB port 2 Browse to the folder that contains the files downloaded from GE portal 3 Run the RevertFlashDisk cmd command NOTE If the Windows Protected Your PC window pops up click More Info and then click Run Anyway If prompted to allow a...

Page 413: ...or 2 Tap Settings and select Config 3 On the Connectivity tab select Dataflow and make sure the USB Storage option is enabled WARNING WARNING WHILE THE SOFTWARE UPGRADE PROCEDURE IS DESIGNED TO PRESERVE DATA YOU SHOULD SAVE ANY PATIENT DATA IMAGES SYSTEM SETUPS TO A BACKUP MEDIA BEFORE PERFORMING A SOFTWARE INSTALLATION IT IS RECOMMENDED TO WRITE DOWN THE TCP IP ADDRESS AND THE WIRELESS NETWORK SE...

Page 414: ...MANUAL P R E L I M I N A R Y 4 Connect USB memstick to any available USB port in the system 5 From the side menu select Admin and then Backup 6 Select the System Configuration checkbox 7 Select USB HD Memstick media as a Destination Device 8 Tap Start Backup Figure 8 228 Backup Options Tab Figure 8 229 Backup Options to Select ...

Page 415: ...the backup is complete the following message appears 9 Eject the USB memstick when prompted 8 9 3 2 Backup of Patient Exams 1 Connect an external USB 3 0 hard drive with at least 100GB of free space 2 Tap the Home button and select Archive Management Figure 8 230 Backup Complete Message Figure 8 231 Archive Management Tab ...

Page 416: ... L I M I N A R Y 3 Select the Select All checkbox to select all patients for backup 4 Tap Send Selected 5 In the Copy window select USB Storage and click OK Figure 8 232 Select All Patients for Backup Figure 8 233 Send Selected Button Figure 8 234 Copy Window NOTICE IMPORTANT Make sure the selected option is USB Storage and not DICOM USB ...

Page 417: ...s complete proceed to Software Upgrade Procedure 8 9 3 3 Software Upgrade Procedure 1 Disconnect all probes 2 Plug the Venue R2 Software Installation Media into the USB port located on the Interface Panel of the Venue 3 Turn on the system The opening screen is displayed 4 Select Upgrade Venue SW to install software without affecting user data Figure 8 235 Copying Patients Dialog Figure 8 236 Venue...

Page 418: ...installation media and tap OK in the message box 9 In the Confirmation dialog box that appears tap Yes to continue the reboot 10 Tap Yes to continue the reboot 11 After the system reboots the Venue setup wizard welcome screen appears Figure 8 237 Venue Partition C Warning Message NOTICE IMPORTANT Venue R1 user presets are not compatible with R2 software The backed up presets can only be used if re...

Page 419: ...K to proceed 13 Tap Run wizard to continue the installation 14 The settings screen appears The Local tab opens by default NOTE Due to FPGAs programming as part of the initialization process it might take up to 15 minutes before the Local tab is displayed Figure 8 240 Venue Installation Wizard Welcome Screen Figure 8 241 Venue Installation Wizard Run Wizard ...

Page 420: ...low the procedure displayed in the window If the following errors appear proceed with T CFE replacement 15 Type in the option key and configure the network settings 16 Continue with the EZ Config installation as specified in Section 3 6 2 1 17 Open Config screen Imaging Advanced Tools and make sure the Auto zone increment Figure 8 242 Venue Setup Wizard Firmware Update Errors Figure 8 243 Venue Se...

Page 421: ...ION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y check box is NOT selected Figure 8 244 Auto zone increment check box Perform the checks listed in Operating System and Application Software Installation Procedure on page 8 219 ...

Page 422: ... Application Software Loading Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 18 Verify that an e label is displayed in Config screen under the About tab as shown below Figure 8 245 Venue Upgrade Label ...

Page 423: ... the CRM system is updated correctly see Figure 8 246 NOTE This step is applicable only for upgrades that were obtained electronically eDelivery 20 In case the purchased upgrade kit was delivered as a physical media scan the UDI label that is attached to the installation media Refer to Figure 8 247 as reference only always use the actual label Figure 8 246 Venue Upgrade UDI Label ...

Page 424: ...9 Operating System and Application Software Loading Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y Figure 8 247 Example of UDI label attached to Software installation media Label ...

Page 425: ...ware Option keys are available prior to commencing the installation process Do not use the Software Option keys from a previous system version At the site perform the following steps before you start the upgrade procedure 1 Perform Disk Management for all database records and select the Copy option for instructions refer to the Venue User Manual on your Document Media NOTE It is recommended that D...

Page 426: ...erating system and the Venue application software This excludes preparation time backup etc and performing functional checks following the installation procedures Throughout the installation processes the Venue system will be unavailable for scanning Section Section 8 9 on page 8 184 provides step by step instructions for installation of Windows operating system software and Venue ultrasound scann...

Page 427: ... port located on the Interface Panel of the Venue 3 Turn on the system The opening screen is displayed 4 For new installation after SSD replacement tap Format SSD and Install If patient information is stored on the SSD perform one of the following Select Update SW to install software without affecting user data The Warning message appears 5 Click Yes in the Warning message box 6 Automatic installa...

Page 428: ...n media and tap OK in the message box 9 In the Confirmation dialog box that appears tap Yes to continue the reboot 10 Tap Yes to continue the reboot 11 After the system reboots the Venue setup wizard welcome screen appears 12 Select the interface language and tap OK to proceed Figure 8 250 Venue Installation Process Information Dialog Box Figure 8 251 Venue Installation Process Confirmation Dialog...

Page 429: ...EVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 13 Tap Run wizard to continue the installation 14 Enter the system serial number and tap OK to confirm Figure 8 253 Venue Installation Wizard Run Wizard Figure 8 254 Venue Installation Process Set Serial Number ...

Page 430: ...5 Tap OK to confirm the serial number 16 The settings screen appears The Local tab opens by default 17 Continue with the EZ Config installation as specified in Section 3 6 2 1 on page 3 27 Figure 8 255 Venue Installation Process Confirm Serial Number Perform the checks listed in Operating System and Application Software Installation Procedure on page 8 219 ...

Page 431: ...eplacement Procedures 8 211 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 8 9 6 NOTE The downgrade procedure is only applicable for R1 systems in which the R2 upgrade procedure failed ...

Page 432: ...erences or variations in accordance with different software versions The software recovery procedure allows the user to reload the operating system and the application Drive C partition without affecting any user information or PHI data This procedure can be initiated from the configuration screen under the Service tab 1 Boot the system into the application screen and open the Config menu 2 Open t...

Page 433: ...ISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 4 A pop up window will be displayed notifying the user about the next steps 5 Select OK Windows recovery options will be displayed 6 Select Troubleshoot Figure 8 257 Information window Figure 8 258 Recovery Options ...

Page 434: ...nue SW reload 8 The system will reboot automatically and the Venue SW reload window will be displayed 9 Select Reload Venue SW The software reload process will start 10 Once the process is complete The Venue SW reload window will be displayed again 11 Select Reboot Venue Scanner 12 The system will reboot Figure 8 259 Troubleshoot Window Figure 8 260 Venue SW Reload ...

Page 435: ...ON 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 13 Continue with Venue GO EZ Config Setup Wizard EZ Config Setup Wizard Perform the checks listed in Operating System and Application Software Installation Procedure on page 8 219 ...

Page 436: ...required Left Side eTower Cover Replacement Procedure No functional checks are required Right Side eTower Cover Replacement Procedure No functional checks are required Mid Thermal Baffle Cover Replacement Procedure No functional checks are required Upper eTower Front Cover Replacement Procedure Mechanical Functions Checks Front Base Cover Replacement Procedure Mechanical Functions Checks Upper eTo...

Page 437: ...edure Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG Electronic Boards BIB Board Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG SSD Module Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG BIB Board Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System...

Page 438: ...ics Select Full System Diagnostics without JIG MPB Blower Fan Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG IPP Module Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG Ground Continuity Test Cables MPB Rear USB Cable Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System Diagnost...

Page 439: ...only Installation Procedure Note TheaboveproceduresaredescribedinSoftware Installation Procedure Power on Boot up Power Shut Down Probe Connectors Check 2D Mode B Mode Checks M Mode Checks PW CW Doppler Mode Checks Audio Check Peripheral Checks Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG Table 8 4 Replacement Procedures Functional Checks Required Continued Repl Part...

Page 440: ...erview of replacement parts for the Venue ultrasound scanner NOTE In the detailed Parts lists illustrations are accompanied by FRU names corresponding Part Numbers and a compatibility matrix This Repl Proc icon indicates refer to the instructions in Chapter 8 Replacement Procedures NOTE The illustrations provided in this chapter are for illustration purposes only and are subject to change without ...

Page 441: ... P R E L I M I N A R Y Section 9 2 List of Abbreviations Assy Assembly T BEP Back End Processor T CFE Control Front End CRU Customer replaceable Unit Ctrl Control T FEPS Front End Power Supply LCD Liquid Crystal Display Int Internal I O Input Output T PSB Probe Selection Board TS Touch Screen ...

Page 442: ...E L I M I N A R Y Section 9 3 Main Assemblies and Sub Assemblies Figure 9 1 Venue System View PSU MPB Batteries Covers eTower System Power Distribution Cockpit Monitor Parts Front End Back End Back End Cables System Power Cables Mechanical Hardware Parts Accessories Optional Peripherals ...

Page 443: ... Parts Diagram 1 Table 9 5 Mechanical Hardware Parts Diagram 1 Item Part Name Part Number Can Replace Part Can Be Replace d by Part Not Compatib le With Rep l Pro 101 CFE RELEASE ARM S5430823 102 Plastic Cable Guide for CFE FRU S5432232 103 MPB GUIDES L and R FRU S5737527 104 FRONT END METAL DOOR FRU S5731310 105 Riser FRU S5759997 106 Mechanical RISER with Friction Adj hole FRU S5759997 1 101 103...

Page 444: ...eplaced by Part Not Compatible With Repl Proc 107 MPB FAN SLEEVE FRU S5755990 108 Base Module FRU S5759996 109 Articular Arm Module FRU S5759998 S5759999 1 110 Articular Arm Module with Locking FRU Applicable only for systems with serial number 140 S5759998 1 S5759999 111 Halo Handle FRU S5759999 S5759998 1 112 Halo Handle with Locking FRU Applicable only for systems with serial number 140 S575999...

Page 445: ... 9 7 Mechanical Hardware Parts Diagram 3 Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 112 NO LOCK CASTOR S5755602 1 113 DIRECTIONAL and BRAKE LOCK CASTOR S5755602 114 Module MPB RECEPTACLE FRU S5730133 115 T TRX BOX Assy CRU S5730759 116 FE DOOR ASSY S5730539 117 MPB BLOWER CRU S5755992 112 115 114 116 113 117 RP RP RP RP RP ...

Page 446: ...ompatible With Repl Proc CRU 201 Cover MPB Door S5759973 202 Cover Mid Thermal Baffle FRU S5759974 203 Cover Scanner Door FRU S5759978 204 Cover Right Side eTower FRU S5759982 205 Cover Left Side eTower FRU S5759983 206 Cover Upper eTower Front FRU S5759985 207 Cover Lower Front eTower FRU S5759986 208 Cover Lower Front with ECG Support S5759986 2 209 Cover Printer Insert FRU S5759987 201 203 207 ...

Page 447: ...h Repl Proc CRU 210 Cover Front Base FRU S5759988 211 Cover Left Side Base FRU S5759989 212 Cover Right Side Base FRU S5759990 213 Cover Right Base Bottom FRU S5759991 214 Cover Riser FRU S5759992 215 Cover Riser with Friction Adj Hole FRU S5759992 1 216 Cover RS Probe Connector FRU S5759993 217 Cover Riser Thermal FRU S5759994 218 Cover Left Base Bottom FRU S5759995 219 MPB Front Metal door FRU S...

Page 448: ...RECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL Chapter 9 Renewal Parts 9 9 P R E L I M I N A R Y 9 4 3 System Power Distribution Figure 9 7 Power Parts 302 303 304 308 305 311 301 310 307 306 309 312 ...

Page 449: ...Y S5737214 5737214 5737214 302 MODULE MPB MAIN POWER BOARD CRU S5731214 303 AC DC PS CRU S5730963 1 304 MODULE IPP FRU S5730761 305 Cable PSU to MPB DC Docking FRU S5721128 306 Cable ON OFF switch FRU S5737035 307 Cable MPB TO T FEPS FRU S5721141 308 Cable MPB To peripherals harness FRU S5721126 309 Cable MPB TO COCKPIT FRU S5721144 310 Cable MPB TO BIB CONTROL FRU S5721137 311 Cable MPB TO BE PWR...

Page 450: ...al Parts 9 11 P R E L I M I N A R Y 9 4 4 Cockpit Monitor Parts Figure 9 8 Cockpit Monitor Parts Table 9 11 Cockpit Monitor Parts Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc CRU 401 MODULE COCKPIT CRU S5737034 401 RP ...

Page 451: ... Cards and Assemblies Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc CRU 501 Module T TRX FRU S5764764 502 Module T PSB Probe Selection Board FRU S5726584 503 Module T FEPS Front End Power Supply FRU S5721459 504 Module T CFE cSound Front End FRU S5730757 505 T TRX BOX Assy CRU S5730759 506 FULL FE MODULE CRU S5721151 501 502 504 503 505 506 RP RP...

Page 452: ... 10 Back End Table 9 13 Back End Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 601 SSD MODULE FRU S5728757 602 Module BIB BackEnd Interface FRU S5727393 603 BIB PLASTIC BRACKET S5728756 604 MODULE BEP PCB FRU S5729618 605 BACK END ASSY CRU S5728755 601 602 603 605 604 RP RP RP RP ...

Page 453: ...e Part Can Be Replaced by Part Not Compatible With Repl Proc 701 Cable MITX TO SSD PWR FRU S5721127 702 Cable BIB TO BE SPLIT CABLE FRU S5721145 703 Cable BE TO IO USB3 FRU S5715745 704 Cable BE TO IO LAN FRU S5715747 705 Cable BE TO IO DP HDMI FRU S5715746 706 Cable BE to Cockpit FRU S5715748 707 Cable Printer USB FRU S5721142 708 Cable PCIe FRU S5715744 709 Cable SATA DATA BE TO SSD A FRU S57157...

Page 454: ...9508 803 Probe L12n RS 5505771 804 Probe 12L RS 5499501 805 Probe E8C RS 5499516 806 Probe 9L RS 5499511 807 Probe C1 5 RS 5499608 808 Probe 6Tc RS KN100106 809 Probe 6S RS 47236956 810 Probe L4 12t RS 5435010 811 Probe L8 18i RS 5499609 Figure 9 13 Software Table 9 16 Software Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 901 Venue DOK SW media ...

Page 455: ...Replaced by Part Not Compatible With 1001 Power Cable Australia R2415383 7 1002 Power Cable Britain R2415383 4 1003 Power Cable China R2415383 6 1004 Power Cable Denmark 5439668 1005 Power Cable Europe R2418616 1006 Power Cable Israel R2415383 1 1007 Power Cable Japan R2415383 5 1008 Power Cable Switzerland R2415383 8 1009 Power Cable USA Canada R2269460 2 1010 Power Cable Brazil 5405959 1011 Powe...

Page 456: ...Accessories Figure 9 15 Accessories Table 9 18 Accessories Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 1101 Venue Basic Storage Basket FRU S5767134 1102 Venue Large Storage Basket FRU S5767133 1103 Venue Power Cable Holder FRU S5766312 1103 1101 1102 RP RP RP ...

Page 457: ...6 Optional Peripherals Table 9 19 Optional Peripherals Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 1201 Printer UP D711MD Kit 5449734 1202 Printer Support Kit Spare Part S5767756 1203 USB Barcode Reader FRU S5767381 1204 Wi Fi adapter 5728576 1205 Wi Fi Support Kit S5767760 1204 1201 1202 1203 1205 RP RP RP RP ...

Page 458: ...e Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 1301 ECG Detachable cable AHA Type USA 5146056 1302 ECG detachable cable IEC type EURO and ASIA 5146739 1303 ECG module 5129487 1304 Norav German Version ECG Module 5800116 1305 USB Cable for ECG 1 5m length 5146055 CAUTION USE ONLY APPROVED DEFIBRILLATION PROOF ECG PATIENT CABLES 1301 1302 1305 1303 1304 RP RP ...

Page 459: ...ire additional tasks and or inspections at a different frequency than listed in this manual 10 1 2 Purpose of Chapter 10 This chapter describes the Care and Maintenance procedures for the Venue ultrasound scanner and its peripherals These procedures are intended to maintain the quality of the Ultrasound system s performance Read this chapter completely and familiarize yourself with the procedures ...

Page 460: ... CIRCUIT BOARDS WHILE MAINS POWER TO THE SYSTEM IS ON CAUTION PRACTICE GOOD ESD PREVENTION WEAR AN ANTI STATIC STRAP WHEN HANDLING ELECTRONIC PARTS AND WHEN DISCONNECTING CONNECTING CABLES CAUTION DO NOT OPERATE THIS ULTRASOUND SYSTEM UNLESS ALL BOARD COVERS AND FRAME PANELS ARE SECURELY IN PLACE SYSTEM PERFORMANCE AND COOLING REQUIRE THIS CAUTION TO ENSURE MUTUAL PROTECTION AND SFAETY OF GE SERVI...

Page 461: ...our GE Service Representative can help you with establishing performing and maintaining records for a Quality Assurance program Contact GE for coverage and or price for service 10 3 3 Maintenance Task Schedule How often should care maintenance tasks be performed The Customer Care Maintenance Task Schedule provided in Table 10 2 on page 10 4 specifies how often the Venue ultrasound scanner should b...

Page 462: ...hly Per Facility s QA Program Notes Clean Probes or before each use Clean Probe Holders Inspect AC Mains Cable Mobile Ultrasound system Check Weekly Inspect Cables and Connectors Clean Console Clean Cockpit Inspect Wheels Casters Brakes and Swivel Locks Mobile Unit Check Daily Check Articulated Arm Movement Mobile Unit Check Daily Functional Checks Also after corrective maintenance Safety Test twi...

Page 463: ...s are recommended 2 Bit TX 15 M3 3 Bit TX 20 M4 4 Bit TX 25 M5 5 Bit TX 30 M6 6 Bit TX 45 M10 7 Socket Set must include 7mm socket 8 Side cutter diagonal 5 or 6 inch 9 Flat Blade Driver 3 2 mm 10 Flat Blade Driver 4 mm 11 Flat Blade Driver 6 mm 12 Phillips Driver PH1 13 Phillips Driver PH2 14 Phillips Driver PH3 15 Hex Key 1 5 mm Unbrako Key Allen Key 16 Hex Key 2 mm Unbrako Key Allen Key 17 Hex K...

Page 464: ...Record all probes and Ultrasound system options 3 Power up Turn the Ultrasound system power ON and verify that all fans and peripherals turn On Watch the displays during power up to verify that no warning or error messages are displayed Where applicable confirm that the battery is charged If no AC Input present use the internal battery 4 Probes Verify that the Ultrasound system properly recognizes...

Page 465: ...t affect this mode of operation 2 CF Mode Verify basic CF Mode Color Flow Mode operation Check the basic Ultrasound system controls that affect this mode of operation 3 Doppler Modes Verify basic Doppler operation PW and CW if available Check the basic Ultrasound system controls that affect this mode of operation 4 M Mode Verify basic M Mode operation Check the basic Ultrasound system controls tha...

Page 466: ...ite properly 2 B W Printer Verify hardcopy output of the B W video page printer Clean heads and covers if necessary 3 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 4 ECG Verify basic operation with customer Table 10 7 Mains Cable Inspection As Appropriate Step Item Description 1 Unplug Cord Disconnect the mains cable from the wall outlet and from the Ultrasound sy...

Page 467: ... stop the Ultrasound system from moving and release mechanism Where applicable check all wheel locks and wheel swiveling for proper operation 11 Cables and Connectors Check all internal cable harnesses and connectors for wear and secure connector seating Pay special attention to probe strain or bend reliefs 12 Shielding and Covers Check to ensure that all EMI shielding internal covers air flow pan...

Page 468: ...e ethanol or methanol alcohol abrasive cleaners or other strong solvents Only use isopropyl alcohol when instructed to do so Do not spray any liquid directly onto the Venue covers Do not allow any liquid to drip or seep into the system DO NOT scratch or press on the panel with any sharp objects such as pencils or pens as this may result in damage to the panel Make sure not to spill or spray any li...

Page 469: ...e touch panel display without affecting the system operation tap Settings and then tap Clean The screen turns black allowing you to use a soft cloth with glass cleaning solution to clean the panel 2 Tap together with two hands the two buttons appearing above the pointing hand symbol to return to normal operation Figure 10 261 Settings Menu Clean Button Figure 10 262 Cleaning Mode ...

Page 470: ...e noted Do not drop a probe or strike it against a hard surface as this may damage the probe elements and the acoustic lens or may crack the housing CAUTION TO HELP PROTECT YOURSELF FROM BLOOD BORNE DISEASES WHEN CLEANING AND HANDLING PROBES WEAR APPROVED NON ALLERGIC DISPOSABLE GLOVES THESE ARE MADE OF NITRILE DERIVED FROM VEGETABLE STARCH TO PREVENT ALLERGIC LATEX REACTIONS CAUTION FAILURE TO FO...

Page 471: ...bes NOTE The most recent up to date information on probes and probe care is available at http www gehealthcare com usen ultrasound products probe_care html Always perform a visual check of the probe pins and system sockets before plugging in a probe When handling probes always observe the precautions listed in Probe Handling on page 10 12 The Interoperative probes often have special usage consider...

Page 472: ...e being visibly clean and properly disinfected The purpose of the regulation is to protect employees in the transportation industry as well as the persons who will receive and or open the package NOTE The US Department of Transportation DOT has ruled that items that were saturated and or dripping with human blood that are now caked with dried blood or which were used or intended for use in patient...

Page 473: ...owing service or repair activities They differ somewhat from the standards that are used for design verification and manufacturing tests e g IEC 60601 1 including national deviations which require a controlled test environment and can place unnecessary stress on the Ultrasound system DANGER TO AVOID ELECTRICAL SHOCK THE ULTRASOUND SYSTEM UNDER TEST MUST NOT BE CONNECTED TO OTHER ELECTRICAL EQUIPME...

Page 474: ...ks holes and similar defects WARNING WARNING Users must ensure that safety inspections are performed whenever damage is suspected and on a regular basis in accordance with local authorities and facility procedures DO NOT use the Venue or individual probes which fail any portion of the safety test WARNING WARNING To minimize risk of electric shock only trained persons are allowed to perform the ele...

Page 475: ...ems found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition where the Neutral grounded supply conductor and the Grounding protective earth conductor are reversed If later tests indicate high leakage currents this should be suspected as a possible cause and the outlet wiring should be visually inspected Figure 10 263 ...

Page 476: ... to the exposed metal parts of the case The ground wire resistance should be less than 0 2 ohms Reference the procedure in the IEC60601 1 DANGER ELECTRIC SHOCK HAZARD THE PATIENT OR OPERATOR MUST NOT COME INTO CONTACT WITH THE EQUIPMENT DURING THIS TEST Figure 10 264 Ground Continuity Test 1 GROUND PIN 2 OHMMETER 3 Venue 4 ACCESSIBLE METAL PART MONITOR HOUSING REAR PANEL CONNECTOR ANY CASTER WHEEL...