Chapter 10 - Care and Maintenance
10-7
P R E L I M I N A R Y
10-5-2
Functional Checks
NOTE:
Refer also to
Chapter 4 - General Procedures and Functional Checks
, for additional details about the
functional checks described in this section.
The functional checks take approximately 60 minutes to perform. Refer to the
Venue™ User Manual
whenever necessary.
10-5-2-1
System Checks
Table 10-5 System Functional Checks
Step
Item (or Mode)
Description
1
B-Mode
Verify basic B-Mode (2D) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
2
CF-Mode
Verify basic CF-Mode (Color Flow Mode) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
3
Doppler Modes
Verify basic Doppler operation (PW and CW if available).
Check the basic Ultrasound system controls that affect this mode of operation.
4
M-Mode
Verify basic M-Mode operation.
Check the basic Ultrasound system controls that affect this mode of operation.
5
Basic Measurements
Check Distance and Tissue Depth Measurement.
6
Applicable Software Options
Verify the basic operation of all optional modes such as Contrast. Check the basic
Ultrasound system controls that affect each option’s operation.
7
Monitor
Verify basic monitor display functions.
8
Peripherals
See: