8-144
Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-7-8
MPB to BIB Control Cable Replacement Procedure
8-7-8-1
Tools
Appropriate Phillips screwdriver.
8-7-8-2
Time Required
10 min
8-7-8-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-8-4
MPB to BIB Control Cable Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower.
3) Remove MPB Front Metal Door.
4.) Remove the MPB module.
5.) Disconnect the cable from the BIB board.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
MPB Front Metal Door Removal Procedure
•
Figure 8-145 Disconnecting MPB to BIB Cable
Summary of Contents for H45051VN