8-56
Section 8-4 - System Modules- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-10
MPB Module Replacement Procedure
8-4-10-1
Tools
•
Appropriate Phillips screwdriver.
8-4-10-2
Time Required
15 min
8-4-10-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-10-4
MPB Module Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe cover, Lower Front eTower cover, Right Side eTower
cover, Left Side eTower cover and MPB Door Cover.
3.) Remove the Battery Packs.
4.) Remove the PSU module.
5.) Release the captive screw at the top center of the MPB.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
MPB Door Cover Removal Procedure
•
Battery Module Removal Procedure
•