8-16
Section 8-3 - Covers - Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-3-8
MPB Door Cover Replacement Procedure
8-3-8-1
Tools
Appropriate Phillips screwdriver
8-3-8-2
Time Required
5 min
8-3-8-3
Preparations
Shut down the Venue™ ultrasound unit, as described in
.
8-3-8-4
MPB Door Cover Removal Procedure
1) Remove all accessories.
2) Remove the following covers: Lower Front eTower cover, RS probe cover, Right Side eTower
Cover and the Left Side eTower Cover:.
3.) Loosen the four captive screws (two on each side), grip the MPB Door Cover at the bottom and pull
toward you to remove it.
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
Lower Front eTower Cover Removal Procedure
•
RS Probe Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
Figure 8-8 Removing MPB Door Cover
Summary of Contents for H45051VN