GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7-45
7-3-8-6 Pencil Probe Test
(cont’d)
Figure 7-24 Report Dialog Box - Pencil Probe Test (Close)
The
Report
dialog box includes the following:
•
Test Result Message:
Displays the test result.
•
Pencil Probe Relays:
Displays a graphic representation of the eight front board pencil probe
relays, as follows:
•
Each relay is represented by a button (highlighted in green) that displays the relay
number.
•
Each button is highlighted in yellow when selected.
•
Each button is highlighted in red if errors were detected for one or more relays.
•
Checkboxes:
Enable you to manipulate bad channels:
•
Show Only Bad Channels:
Select this checkbox to display only the bad channels in the
Display
area.
•
Mark bad Channel:
Select this checkbox to manually mark the currently selected vector
as bad.