GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
8-64
Section 8-5 - Back End Parts Replacement
8-5-3
BEP2 Cover and Octopus Card Holder Replacement Procedure
NOTE:
BEP2 refers to a Pentium 4-based BEP (newer type).
8-5-3-1
Tools
Use the appropriate flat and Phillips-type screw drivers, as indicated in the BEP cover replacement
procedure.
8-5-3-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-5-3-3
BEP2 - Side Cover Removal Procedure
1) Remove the two screws (that secure the side cover in position) from the rear of the BEP (see
Figure
8-59
below).
2) Slide the cover backwards and remove it.
8-5-3-4
BEP2 - Side Cover Installation Procedure
1) Return the cover to the side of the BEP and slide it forwards into position.
2) Replace and tighten the two screws to hold the cover firmly in position (
Figure 8-59 "Side Cover -
BEP2" on page 8-64
).
Figure 8-59 Side Cover - BEP2
Screw
Screw