GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-61
8-5-1-4 Back End Processor Installation Procedure
(cont’d)
Figure 8-56 System Configuration - Technical Support Tab
26.)Select the appropriate settings, as follows:
*
Ex Keyboard
-
Revision 4
*
Sound Board
-
Ver. 3
*
Number of Fans -
Set to
7
for BT00 systems
;
Set to
5
for BT01 systems
*
UPS
-
Change the UPS setting to
Used
or
Not Used
, as appropriate.
If the
Plug & Scan
option was purchased, the UPS setting can be
selected as
Used
. If this option not purchased, select
Not Used.
*
VIC
-
Set to
PC VIC
*
Nominal Voltage -
Set to the appropriate local voltage setting.
27) Perform a complete functional check of the system, as described in
Chapter 4 -
- Functional
Checks.
28) Install the Vivid 3N rear, left and right covers - refer to the
8-2-2-4 "Side Covers Installation
Procedure" on page 8-4
and
8-2-4-4 "Rear Cover Installation Procedure" on page 8-7
.