Chapter 8 Replacement Procedures
8-89
8-5-14
Hard Disk Replacement Procedure
8-5-14-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the Hard Disk
replacement procedure.
8-5-14-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-5-14-3
Hard Disk Removal Procedure
1) Remove the left cover, as described
8-2-2-3 "Side Covers Removal Procedure" on page 8-4
.
2.) Remove the BEP side cover as described in the appropriate procedure:
8-5-2-3 "BEP1 - Left and Right Side Cover Removal Procedure" on page 8-62
, or
8-5-3-3 "BEP2 -
Side Cover Removal Procedure" on page 8-64
.
3.) Remove the screw that holds the hard disk module in position (as shown in
Figure 8-74
below).
4) Carefully slide the module out from the unit.
5) Remove each of the four screws (two on each side), as shown in
Figure 8-77 "Hard Disk Removal
2" on page 8-90
.
Figure 8-76 Hard Disk Removal 1