GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
8-82
Section 8-5 - Back End Parts Replacement
8-5-12-4
BEP1 - ECG Module Installation Procedure
NOTE:
BEP1 refers to a Pentium III-based BEP (earlier type).
1) If necessary, assemble the side rails onto the new ECG Module as shown in
Figure 8-69
below.
Note:
In the event that screws or side rails are misplaced, one can use the screws, slides, and
rails from the ECG Module that was removed from the BEP.
2) Remove the protective layer from the Purpose Gaskets, and affix the gaskets in line with the
existing Purpose Gaskets on the bottom of the new ECG module.
3) Slide the new ECG Module into the front of the BEP, pushing it in until it is properly aligned with the
other modules at the front of BEP.
Figure 8-69 ECG Module with Rail Attached