Section 8-6 - Lower Section Components Replacement
Section 8-6
Lower Section Components Replacement
8-6-1
AC Distribution Box Replacement Procedure
8-6-1-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter as indicated in the AC
Distribution Box replacement procedure.
8-6-1-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-6-1-3
AC Distribution Box Removal Procedure
1) Remove the Vivid 3N left, right and rear covers, as described in
8-2-2-3 "Side Covers Removal
Procedure" on page 8-4
, and
8-2-4-3 "Rear Cover Removal Procedure" on page 8-7
.
2) Unplug all the internal cables from the AC Distribution Box and the voltage setting plugs at Jumper1
and Jumper2.
3) Unscrew the Ground cable located at the top-right corner of the AC Distribution Box.
4) Remove the two screws that are located on the lower bracket of the AC Distribution Box.
Figure 8-78 Lower Bracket of the AC Distribution Box
5) Gently pull the AC Distribution Box towards you, pulling it out through the rear of the machine.
Connector
Primary
Connector
Secondary