GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
7-12
Section 7-3 - Performing Front End (FE) Diagnostics
7-3-2 Radio Frequency Interface (RFI) Diagnostic Tests (for RFI Configuration)
(cont’d)
6.) Review the test results in the Comment area and proceed as follows:
-
If no problems were identified, proceed to re-boot the system, as prompted (Figure 7-6,
above).
-
If no problems were identified but the problem persists, contact your local GE Medical
Systems field engineer for assistance.
-
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration section, on page 6-14, and then repeat the RFI test.
-
If errors were identified and the results were not marginal, verify that the RFI cables are
connected and functioning properly, as described in Chapter 5 - Components and Function
(Theory). If the problem persists, contact your local GE Medical Systems field engineer for
assistance.
Figure 7-6
RFI Tab - Self Test Complete