Chapter 8 Replacement Procedures
8-35
8-3-3
Vivid 3N 17" Monitor Replacement - Procedure 2
NOTE:
This section describes the procedure for replacing a 17" monitor based on a MAG 17" monitor type.
8-3-3-1
Tools
Use the appropriate Phillips-type screw drivers as indicated in the Vivid 3N
monitor replacement
procedures.
8-3-3-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-3-3-3
Vivid 3N 17" Monitor Removal - Procedure 2
1) Unplug the VGA cable and AC power cable from the rear control console.
2) Remove the four screws that fasten the plastic cover of the base assembly, two screws from either
side. Remove the plastic cover.
3) Loosen the securing screw located at the rear of the monitor base. Once this screw is loose, you
can fully rotate and lift the monitor out from the machine.
Figure 8-36 Vivid 3N Monitor Release Screw
4) Remove the monitor by pulling it upwards. You may need to swivel the monitor slightly
right
and
left
while pulling it upwards.
NOTE:
Ensure that the Teflon washer remains in place when you remove the monitor.
5) Carefully lay the monitor on a soft surface and turn it upside down, so that its underside is
accessible, as shown in
Figure 8-37 "Vivid 3N Monitor showing Underside" on page 8-36
.