Section 8-5 - Back End Parts Replacement
8-5-10
CDRW Drive Replacement Procedure
8-5-10-1
Tools
Use the appropriate flat and Phillips-type screw drivers and a wire cutter.
8-5-10-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-5-10-3
CDRW Drive Removal Procedure
1) Remove the Vivid 3N left and right side covers, and rear cover - see the
8-2-4-3 "Rear Cover
Removal Procedure" on page 8-7
and the
8-2-2-4 "Side Covers Installation Procedure" on page 8-4
.
2) Disconnect the power cable from the rear of the BEP.
3) Remove the BEP left and right side covers, as described in the appropriate procedure:
8-5-2-3 "BEP1 - Left and Right Side Cover Removal Procedure" on page 8-62
, or
8-5-3-3 "BEP2 -
Side Cover Removal Procedure" on page 8-64
.
4) Remove the BEP front cover as described in the appropriate procedure:
8-5-2-5 "BEP1 - Front Cover Removal Procedure" on page 8-63
, or
8-5-3-5 "BEP2 - Front Cover
Removal Procedure" on page 8-65
.
5) Remove the cables (data and power) that connect to the rear of the CDRW Drive.
6.) Carefully slide the CDRW Drive out from the front of the BEP as shown in
Figure 8-66
below.
Figure 8-66 CDRW Drive Removal