GE Medical Systems
EAGLE 1000
Page 81
Information Technologies
Servicing Instructions
227 468 01 Rev. G
3.4 Safety Analysis Test
3.4.1 General introduction
The suggested Safety Analysis Tests refer to the international standard IEC 601-1.
The test are generallay performed with Safety Testers, on most of them, the measuring
circuits according IEC 601 are already implemented.
The following is a general description of the tests. For the handling of your Safety Tester
follow the user manual.
The tests may be performed under normal ambient conditions of temperature, humidity
and pressure and with line voltage. The leakage currents correspond to 110 % of rated
voltage for the tested unit. Most Safety Testers take this into account, otherwise the mea-
sured values have to be calculated.
In countries like Germany adequate tests according DIN VDE 0751 may be performed.
3.4.2 Recommended Test Equipment
- Safety Tester for measurements according to IEC 601.
- Testing connector according to the following drawings.