GE Medical Systems
EAGLE 1000
Page 20
Information Technologies
Servicing Instructions 227 468 01 Rev.G
2.1.4 Power supply
* from the power line, monitor design in IEC protection class I
Potential equalization
* potential equalization connection according to DIN 42 801(measurement in or at the
heart) or connection for function ground (interference elimination during general cardio-
logical examinations)
Rated voltage range from 95 to 240 V AC
* operating voltage range from 85 to 264 V, 49 to 65 Hz
* rated current from 0.15 to 0.3 A
Internal power supply (rechargeable battery)
* automatic toggling from line to battery operation and vice versa
* integrated charge circuit with automatic adaption of charge characteristics
* fast battery charging to 80% of its capacity within 3.5 hours, afterwards
* normal charging to 100% of its capacity within 8 hours, afterwards
* continuous trickle charging by means of charge pulses to prevent the memory effect
Charge balance
measurement of the charge and load current
compensation for battery self-discharge
graphic indication of battery capacity
Alarms
warning when battery provides power for only 5 minutes of operation (without
recorder)
automatic switch-off at 10.0 V to prevent overdischarging
Capacity of the rechargeable battery
2000 mAh, 12 volts
Charging indicator
LED
Operating time
approx. 3 hours (typical operating time with fully charged battery at 23 °C, at a
battery temperature of 60 °C the capacity may be reduced to 70%)
(min. 2 hours at 25 °C ambient temperature and new, fully charged battery)
Summary of Contents for Eagle 1000
Page 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 130: ......
Page 135: ......
Page 136: ......
Page 138: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......