GE Medical Systems
EAGLE 1000
Page 26
Information Technologies
Servicing Instructions 227 468 01 Rev.G
Connections
direct connection for measurement of one pressure
connection via adapter for measurement of two pressures
Display
pressure waveform with 25 mm/s sweep for both channels
pressure waveform for the following pressure sites: arterial (ART1, ART2), pulmonary
artery (PA), venous (VEN), left-atrial (LAP), intracranial (ICP) for ART, PA and LAP the
systolic, diastolic and mean values can be displayed
display update interval: 2 seconds
Signal processing
Pressure
measuring range -30 to +300 mmHg
resolution 0.25 mmHg, internal
cut-off frequency 12 Hz
accuracy ±2 mmHg ±1% of actual value
zero drift < ±0.1 mmHg/°C
5 minutes to warm up for full accuracy
Zero balancing
automatic compensation of the transducer value for a zero pressure in both channels,
every channel to be zeroed is checked for absence of pulse signals
balancing initiated after key press > 2 seconds
balancing range ±100 mmHg
storage of the balancing value for 5 minutes after power failure or when monitor is
switched off
deviation < ±1 mmHg
initiation only when signal modulation < ±1 mmHg