Electrical Diagram for Patient Leakage Current Test
For protection of the test person, the following values of resistor R may be used:
Typ BF
22 kOhm (120 to 130 V)
47 kOhm (220 to 240 V)
Typ CF
100 kOhm (220 to 240 V)
3.4.4.3 Enclosure Leakage Current Test (System)
If the Eagle is connected to SynOpsis the enclosure leakage current has to be measured
additional according chapter 3.4.4.1.
Doing this measurement, the Eagle needs to be connected (hard wired) to SynOpsis.
R
UNIT
UNDER
TEST
POWER CORD
patient cable
or test plug
POWER CORD
Safety Tester
HIGH
LOW
GND
GND
NORM
S1
[M]
PATIN JACK
RVS