GE Medical Systems
EAGLE 1000
Page 43
Information Technologies
Servicing Instructions
227 468 01 Rev. G
2.2.10 Non-invasive blood pressure module
2.2.10.1 Function
The PCB NBP acquires and automatically processes the non-invasive blood pressure
signal. Via an internal asynchronous connection it passes the non-invasive blood pressu-
re data on to the PCB ECG Monitor. From here it receives a reset signal as well as a
separately controlled pump power supply for safety reasons. The PCB is not electrically
isolated as isolation is effected by the pneumatic signal input.
Warning: For safety requirements, when replacing the PCB it is important to use
the flat strip cable of correct length to match the version used.