GE Medical Systems
EAGLE 1000
Page 99
Information Technologies
Servicing Instructions
227 468 01 Rev. G
4.5.5 Recorder service
Motor adjustment
What to adjust
What to mea-
How
Where to
How big and What else to
or to check?
sure with?
Test mode
turn?
exact?
consider?
Motor speed
Check mea-
Press Soft-
Speed ad-
<1 %
V = 25 mm/s
surements ofkey Print
juster
f
or
grid printout
for a moment
R 503
Eagle 1000
with ruler
Heating time adjustment
What to adjust What to mea-
How
Where to
How big and What else to printhead
or to check?
sure with?
Test mode
turn?
exact?
consider?
legend
Heating time
Connect oscil-
Press Soft-
Heating time referred to
Adjustment
loscope to
key Print
adjuster
printhead
made at
MP 2/1 and
for a moment R 504
temperature
25 mm/s.
MP 1/1 (GND)
22 °C:
Caution:
printhead
warms up
during
printing
450
µ
s ±1 %
R520
550
µ
s ±1 %
R650
Checking the +22 V supply voltage
What to adjust
What to mea-
How
How big and
What else to
or to check?
sure with?
Test mode
exact?
consider?
+22 V power
Voltmeter
B22 V
+22 V
±5 %
Recorder dis-
supply
and GND (C 500)
abled
Overtemperature protection
There is a protective circuit which protects the printhead from being overloaded. At a sub-
strate temperature exceeding approx. 58 °C the printhead switches off.
Summary of Contents for Eagle 1000
Page 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 130: ......
Page 135: ......
Page 136: ......
Page 138: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......