GE Medical Systems
EAGLE 1000
Page 24
Information Technologies
Servicing Instructions 227 468 01 Rev.G
2.1.8 Non-invasive blood pressure
Signal input
Single-lumen tubing system, type CF signal input (IEC 601-1), patient is electrically not
connected to the system, thus the system is protected from defibrillation discharges and
electrosurgery voltages).
Signal processing
Oscillometric method:
stepwise pressure decrease, zero pressure reference prior to each measurement
Two measuring modes:
adult
neonate
mmHg
adult
systolic
25 to 255
diastolic
10 to 220
mean
18 to 255
neonate systolic
20 to 155
diastolic
5 to 110
mean
10 to 130
The specifications are met for pulse rates between 40 and 240 BPM (adult/neonate).
Safety circuits
Redundant monitoring of the cuff pressure by means of a second independent pressure
sensor; cuff deflation immediately on exceeding max. cuff pressure, on detection of faults
during a measurement and on power loss.
Max. cuff pressure
adult
300 mmHg
neonate 160 mmHg
Special features
When the cuff is properly applied the module recognizes in the adult mode whether the
cuff matches the selected range. If the wrong cuff is applied, the measurement is interrup-
ted.
Summary of Contents for Eagle 1000
Page 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 130: ......
Page 135: ......
Page 136: ......
Page 138: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......