3.2 Checkout Procedures
Note:
The F keys F1-F5 change their function depending on the selected menu.
When the following testing instructions refer to such a text this is displayed in square
brackets
[....]
.
The numbers given in parentheses
(..)
of operating controls and socket connections refer
to the diagram on the front and rear panels of the monitor in the users manual.
3.2.1 Recommended testing/calibration equipment and accessories:
For monitor models
10112301 - 10112307
10112401 - 10112407
1X Multiparameter simulator
Lionheart
ECG
1X 3-lead patient cable
223 287 01
1X 3-lead patient trunk cable
223 288 01
1X Set of electrode leads
384 011 34
Invasive blood pressure
1X Blood pressure simulation cable
Lionheart with
Hellige connector
1X Dual transducer cable
220 102 01
Non-invasive blood pressure
1X Digital pressure manometer Diptron 3 plus
Measuring range 500 mmHg
Accuracy 0.04% FS
Parts for the pneumatic circuit comprising:
1m Tubing
822 046 30
1X Fastening coupling
929 166 35
1X Coupling
929 166 34
2X T-piece
929 167 02
1X Pumpball with deflation valve, connecting nipple 3 mm - 5mm
1X Connecting tubing
216 118 01
1X Blood pressure cuff
217 319 04
SpO2
1X SpO2 simulator
408 610 - 001
1X Simulator cable SPO2
223 417 01
1X Finger probe
701 240 21
1X Connection cable for SpO2 probe
303 443 58
Temperature
1X Temperature simulator
220 104 01
Summary of Contents for Eagle 1000
Page 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 130: ......
Page 135: ......
Page 136: ......
Page 138: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......