25
K249 Issue No. 4
English
mA Step
To select one of the electrical output programs, press the
OUTPUT
key and
proceed as follows (e.g.),
On selection of (e.g.) Linear, the output display window changes to show the
selected program of output currents (e.g.),
•
Connect an external power source as shown on Page 24.
•
Press
RUN
(F1) to run program. A flashing status display
CHECK LOOP
indicates a fault in the external loop i.e. supply fault or open circuit.
Note:
The dwell time at each step is approximately 10 seconds.
•
Press
STOP
(F1) when running to stop at any point. Press
RUN
(F1) to
resume.
Advanced Task
Select Output