Current to Pressure Converter (I-P) Task
Taking Measurements
•
Use the
Up
and
Down
cursor keys to adjust the loop current to the
required value. Alternatively, press
ENTER
and use cursor keys to enter a
finite value. Cursor keys can then be used to nudge the output either up or
down.
If required, change pressure units with INPUT key. A flashing CHECK LOOP
message indicates either an open circuit supply loop (or no external
supply).
Pressure Switch Test (P-SWITCH) Task
•
Contact state will be shown on display. When contacts close, buzzer sounds.
•
To run switch test, close vent valve and press the
RUN
(F1) key.
•
Using the handpump, increase the applied pressure to just below the
switch operating point. Screw the volume adjuster in until the switch
operates (the operating pressure of the switch is then written to the
display).
•
Reduce pressure until the switch releases (indicated by the switch symbol).
The release pressure is then written to the display and the hysteresis
displayed.
PRESSURE INT
bar
F2
F1
External Pressure Source
(Indicator Only)
Pressure
Regulator
RUN
CONTACT STATE
max 30V
TASK : P-SWITCH
SNAPSHOT MODE
DPI 615 IS
Summary of Contents for DPI 610 IS
Page 4: ...Page 4 of 68 K249 Issue No 4 Do Not Print This Page ...
Page 5: ...Druck DPI 610 615 IS Portable Pressure Calibrator User manual K249 GE Infrastructure Sensing ...
Page 6: ......
Page 9: ...iii K249 Issue No 4 English English ...
Page 57: ...English 43 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION ...
Page 58: ...44 K249 Issue No 4 This page is intentionally blank ...
Page 64: ...50 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION Operation Notes ...
Page 65: ...English 51 K249 Issue No 4 Appendix 1 DATALOG FILE EXAMPLE ...
Page 66: ...52 K249 Issue No 4 This page intentionally blank ...